EPAP Devices Optipillows, Bongo Rx, and Theravent: Flow Resistance and The Pressures They Generate

Review | Open Access

  • 1. T.S.Hakim: Sleep Apnea Treatment Unit, Phoenix, AZ,
  • 2. Marco Bonetti: Domus Medica private hospital, Republic of San Marino
  • 3. G Bosco: Dept. of Biomedical Sciences, University of Padova, Padova, ITA.
  • 4. EM Camporesi: University of South Florida, Department of Surgery, Tampa, FL.
+ Show More - Show Less
Corresponding Authors
T.S. Hakim, Ph.D. Sleep Apnea Treatment Unit 10443 N Cave Creek Rd, Suite 110 Phoenix, Arizona 85020 , Phone: 602-944-0847
Abstract

Theravent and Optipillows are expiratory resistance devices indicated for snoring, and Bongo Rx is indicated for mild to moderate obstructive sleep apnea. The goal of this study is to compare side by side the resistance of the three devices and the pressure that they generate. The devices’ resistance was evaluated using steady air flow at 100 ml/sec or simulated breathing. We calculated the device’s resistance either at a flow rate of 100 ml/sec or as the slope of the pressure-flow curve. The pressure generated during expiration with simulated breathing was also measured. The study shows that the resistances of Theravent and Bongo Rx were identical regardless how they were calculated. The resistance of Optipillows was equivalent to resistances of Theravent and Bongo Rx when the side vent was between 50 to 75% closed. Likewise, with simulated breathing, the pressure generated during expiration was identical in Theravent and Bongo Rx, and was comparable to the expiratory pressure with Optipillows when the side vent was closed between 50 to 75%. In conclusion, the three expiratory resistance devices had comparable resistance and generate comparable expiratory pressure. Despite the fact that some devices are cleared for snoring and some for obstructive sleep apnea, any of these three devices can work for either condition. The healthcare provider can recommend whichever is most accessible to the patient. The adjustable resistance on the Optipillows provides an option for the patient to select low resistance while adapting to the device.

Citation

Hakim TS, Bonnetti M, Bosco G, Camporesi EM (2021) EPAP Devices Optipillows, Bongo Rx, and Theravent: Flow Resistance and The Pressures They Generate. J Sleep Med Disord 7(1): 1119.

ABBREVIATIONS

CPAP-continuous positive airway pressure, EPAP- expiratory positive airway pressure, OSA-obstructive sleep apnea

BRIEF SUMMARY

Some EPAP devices are intended to treat snoring while others are intended for obstructive sleep apnea. It is not clear if the intended use is determined by the level of their expiratory resistance or the pressure they generate during expiration. This study shows that despite the fact that some devices are cleared by the FDA for snoring and some for obstructive sleep apnea, the resistance of the three devices are comparable and therefore can be used for either condition. The physician can decide if any of these devices are suitable for their patients.

INTRODUCTION

The pathophysiology of obstructive sleep apnea can be explained by the collapse of upper airways secondary to the slight negative pressure that develops during inspiration. During sleep, the upper airway muscles relax, but the airway passage remains open enough to allow air to flow in most individuals. In some individuals, the upper airway collapses, causing airflow obstruction. Partial obstruction of the upper airway causes snoring noise, and complete block causes total cessation of airflow called obstructive sleep apnea (OSA). The preferred treatment for snoring or obstructive sleep apnea is continuous positive airway pressure (CPAP). The mechanism of preventing OSA is the stenting action by the positive pressure in the upper airway [1]. The positive pressure stabilizes the upper airway, preventing collapse during inspiration. Expiratory positive airway pressure (EPAP) devices have recently been shown to be safe and effective in many patients with snoring and obstructive sleep apnea [2- 6]. They have been considered a welcome addition to the options available for patient care [7-9]. It is unclear why such devices are not always as effective in all patients as CPAP [2, 10]. Previously we discussed the possible reasons why CPAP is superior to EPAP [11]. We emphasized that the primary difference is that with CPAP, the pressure remains positive during the entire breathing cycle eliminating any potential collapse of the upper airway. With EPAP, the pressure is positive only during the expiration and remains near zero or becomes slightly negative during inspiration, allowing upper airway collapse in some users.

EPAP devices allow the user to inhale with ease, but during expiration, the device’s resistance causes the upper airway pressure to become positive during exhalation. The device’s effectiveness depends on their resistance and the pressure they generate. Information is sometimes provided about the expiratory resistance. Measurements in clinical trials while using EPAP devices have shown peak nasal pressure during expiration ranges from 5 to 20 cmH2O [12]. It is not clear what causes such a wide range in pressure, considering that the resistance is fixed. The device’s resistance is usually reported as the pressure developed with a steady flow of 100 ml/sec, and the ratio of pressure to flow is reported in cmH2O/L/sec. Provent (Ventus Medical, Belmont, CA) was described to provide resistance values of 50 cmH20/l/sec [2-5]. In contrast, Theravent (Ventus Medical, Belmont, CA) is reported to have lower resistance values of 30 cmH20/L/sec [13]. The resistance of Bongo Rx (Airavant Medical, Deerfield Beach. FL) was only reported in an abstract [14]. Optipillows EPAP mask (Cpapnea Medical, Phoenix, AZ) has an adjustable expiratory side vent that provides resistance values between 2 to 75 cmH2O/L/sec [15]. Some of the EPAP devices are cleared for treatment of snoring (Theravent and Optipillows EPAP Mask), and some are cleared for the treatment of obstructive sleep apnea (Bongo Rx), however it is not clear if the indication for use is related to the pressure generated by the device. This fact can be confusing for clinicians when deciding if the EPAP device is appropriate for their patients. This study aims at characterizing the resistance of three EPAP devices, Optipillows EPAP mask, Bongo Rx, and Theravent, and the pressure that they generate, side by side, using the identical equipment calibrations and procedures.

MATERIAL AND METHODS

Comparison of the EPAP devices using Steady unidirectional flow

Figure 1 shows the experimental setup used to compare the three devices with a steady flow. The design included an air blower with adjustable settings (ResMed S9 CPAP machine). The air from the blower passed through a tube flowmeter with a floating ball (Dwyer Instruments VFB-68, Michigan City, IN) to use for calibration of the flow meter and verify the accuracy of the flow rate measurement. The pressure was measured from a side port utilizing a pressure transducer (Grass Instruments, Warwick, RI). The pressure transducer was calibrated by applying a static pressure from the CPAP machine. A Pneumotach (Biopac Systems, Airflow Transducer TSD 117, Goleta, CA) was included in the setup to allow continuous recording of the flow rate. The pressure transducer and flow rate signals were fed through a data acquisition system (Biopac Systems UM100A, Goleta, CA), displayed on a computer monitor, and stored on a computer. Verification of the pressure and flow rate measurements were done repeatedly during the experiment. The EPAP devices were prepared to be easily connected to the airflow system (Figure 1). The Optipillows EPAP Mask and Bongo Rx were connected to the airflow system via a wide bore Y-connector. Theravent (regular strength) was adhered onto an adaptor’s surface and was connected to the airflow system. The Optipillows EPAP Mask was tested with the side vent 50% and 75% closed. These two settings were selected because they were easy to estimate and identify. The measurements were made using two levels of flow rates, 100 and 200 ml/sec, which are within the range of expiratory flow rate in human adults.

After completing calibration and verification of zero levels, one EPAP device was connected, and the flow was set at 100 ml/sec, then at 200 ml/sec. Adequate time was allowed for the pressure to change and reach a steady level at each flow rate. The EPAP device was switched, and the measurements were repeated for the other EPAP devices in random order. Theravent and Bongo Rx had a fixed-resistance and were measured once with the two flow rates. This measurement was repeated 12 times for each of the EPAP devices in random order.

The resistance of each device was calculated as the ratio of pressure to the recorded flow. Furthermore, the resistance was also calculated using the pressure-flow curve slope between 100 and 200 ml/sec (A more reliable value for assessing resistance). The results are shown as mean±SD, and differences in pressures and resistances were compared using multi-variance analysis with post hoc comparisons of selected pairs using the Bonferroni test. A p-value of less than 0.05 was considered significant.

Comparison of the EPAP devices using simulated breathing

In this series of experiments, we measured the expiratory pressure generated by the EPAP devices using a similar setup to simulate breathing. The setup is described in figure 2 and was designed to mimic normal breathing at rest, generating a known tidal volume of air moving in and out of the EPAP device (200 or 400 ml), with a frequency of breathing (15 per min), and standard flow rate. We utilized a Plexiglas chamber in which a special respiratory breathing balloon was suspended. A Harvard respirator (Harvard Apparatus Model 613 Dual-Phase respirator) was connected to the chamber to create negative pressure swings causing inflation and deflation of the balloon. The swings in the pressure provided a flow of air in and out of the balloon of 200 or 400 ml. The EPAP devices were tested in random order. The data were recorded and stored on the computer.

After completing the calibration of the flow and pressure sensors, the respirator was started to simulate breathing. The EPAP devices were prepared as above and connected to the system in random order. The Optipillows EPAP Mask was tested as before with the side vent 50% and 75% closed. The first EPAP was tested and then replaced with another device, and the measurements were repeated. The measurements were repeated 12 times on random EPAP samples.

 

RESULTS

Table 1 shows the devices’ calculated resistance at 100 and 200 ml/sec as well as the slope of the pressure-flow curve between 100 and 200 ml/sec. At 100 ml/sec, Theravent and Bongo Rx’s practically identical (12.76+1.05 and 13.21+1.29 cmH2O/L/sec). Likewise, the calculated resistance at 200 ml/ sec was also identical for Theravent and Bongo Rx but differed from the values calculated at 100 ml/sec. The resistance at 100 ml/sec is usually reported for other EPAP devices. The resistance of Optipillows with the side vent 50 % closed was smaller than the resistance of Theravent and Bongo (p < 0.05) but with 75% closed, the resistance was much larger (p < 0.05). The slope of the pressure-flow curve was also identical for Theravent and Bongo. The slope for Optipillows with the side vent 50% closed was smaller than Theravent and Bongo Rx but was much larger when 75% closed (p<0.05). The slopes are displayed in a graphical format in (Figure 3).

As Figure 3 shows, the slopes were identical for Theravent and Bongo Rx (44.19±1.32 and 45.36±1.15 cmH2O/L/sec). In comparison, the slope with Optipillows was 30.66±2.06 and 76.56±4.43 cmH2O/L/sec with the side vent 50% and 75% closed, respectively.

Figure 4 shows typical tracings of pressure and flow during simulated breathing with 400 ml tidal volume. One breathing cycle (inspiration and expiration) with each of the three EPAP devices is shown. The tracing of each cycle was cut and pasted to prepare the figure. As expected, the pressure during expiration became positive during the entire exhalation period with all three devices figure 4. The flow rate with the different devices was set by the respirator, and was not affected by the device. The pressure wave with Bongo Rx and with Theravent looked identical. Optipillows with the side vent 50 and 75% closed produced pressure waves slightly smaller and larger than the other two devices, respectively. The pressure during inspiration remained near zero in all three devices suggesting that inspiratory resistance is minimal for all three devices. The average peak expiratory pressures during breathing with small (200 ml) and large tidal volume (400 ml) are shown in (Figure 5). Peak expiratory pressure was identical in Theravent and Bongo Rx; 3.86+0.17 and 3.75+0.18 cmH2O during breathing at low volume, respectively (left panel). Optipillows produced peak pressures of 2.32+0.11 and 8.19+0.77 cmH2O during breathing with low volume with the side vent 50 and 75% closed. The peak pressure was larger for all devices at the higher tidal volumes, but the differences were consistent with breathing at low tidal volume: 12.91+0.43 and 12.99+0.43 cmH2O for Theravent and Bongo Rx, respectively. In comparison, the Optipillows produced and 8.03+0.29 and 22.16+1.59 cmH2O with the side vent 50% and 75% closed, respectively.

The results show that peak expiratory pressures with Theravent and Bongo Rx were identical and well within the range that the Optipillows EPAP Mask provided. Peak expiratory pressure with Theravent and Bongo Rx were between the peak pressures generated with Optipillows EPAP mask when the side vent was 50% and 75% closed. The results were equally similar at low and at high tidal volume. The peak pressure at large tidal volume was higher than with the pressure at smaller tidal volume, suggesting that expiratory pressure in humans would be proportional to the tidal volume. The resistance of the side vent on Optipillows EPAP Mask can be easily adjusted to approximately midway between 50 and 75% (approximately 60%) to obtain a resistance comparable to Theravent or Bongo Rx. Furthermore, Optipillows EPAP Mask with the side vent 75% closed provided expiratory resistance equal to another EPAP device, Provent Sleep Apnea Therapy; the calculated resistance at 100 ml/sec, the slope of the pressure-flow curve, and the peak expiratory pressure were comparable (unpublished data).

 

Table 1: Calculated resistance of the 3 EPAP devices at the flow of 100 ml/sec, 200 ml/sec, or from the slope of the pressure-flow curve between 100 to 200 ml/sec (i.e., Δ Pressure/ Δ Flow). Optipillows EPAP Mask was either ½ closed or ¾ closed. Values are mean ± SD.
  Calculated Resistance of the Devices with steady unidirectional flow (cmH2O/L/sec)
Flow Rate Optipillows EPAP Mask 1/2 Optipillows EPAP Mask 3/4 Theravent Bongo Rx
100 ml/sec 8.78 ± 0.91 28.51 ± 2.10 12.76 ± 1.05 13.21 ± 1.29
200 ml/sec 20.05 ± 1.40 54.90 ± 3.10 28.44 ± 1.23 28.89 ± 1.08
Slope 30.66 ± 2.06 76.56 ± 4.43 44.19 ± 1.32 45.36 ± 1.15

 

DISCUSSION

The preferred treatment for obstructive sleep apnea is continuous positive airway pressure (CPAP). EPAP devices have become a viable option for the treatment of snoring and obstructive sleep apnea because of the abundance of data showing their safety and effectiveness [2, 3, 5, 16]. With EPAP, the pressure in the upper airway remains close to zero during inspiration and therefore upper airway collapse remains possible [12]. Nevertheless, the upper airway is thought to remain more dilated because of increased pressure during expiration, diminishing the potential for upper airway obstruction. To understand the effectiveness of EPAP devices, it is useful to know the EPAP device’s resistance and the pressure they potentially generate during expiration. The measurements that we made are accurate and reflect the devices’ performance. The measurement that we made using a steady flow at 100 ml/sec is typically reported about the EPAP devices’ resistance. The simulated breathing measurements were done to assess how much pressure is generated during expiration in adult humans. The use of simulated breathing is useful because the breathing pattern can be controlled. There are some differences compared to breathing in humans. The flow wave with the respirator was a sinusoidal wave with a peak pressure or flow in the middle of expiration. During normal breathing, the pressure and flow during expiration while using an EPAP device tend to exhibit a flattened plateau instead of a rise to a peak point [12]. Thus the peak pressure in human may be slightly less than predicted by simulated breathing. Otherwise, simulated breathing patterns represented normal breathing in tidal volume, frequency, and flow rate. The results show that Theravent and Bongo Rx have almost identical expiratory resistance and generate equal pressure during expiration. Yet, one EPAP device is cleared for the treatment of snoring (Theravent), while, the other is indicated for the treatment of mild to moderate obstructive sleep apnea (Bongo Rx). The Optipillows EPAP Mask can provide resistance equivalent to Theravent or Bongo Rx when the side vent is closed midway between 50 and 75%. The Optipillows EPAP Mask has been cleared for snoring treatment but generates expiratory pressure equal to Bongo Rx. Bongo Rx is a prescription-only EPAP device, whereas Theravent and Optipillows are over-thecounter devices.

The Optipillows EPAP Mask with side vent 75% closed provided expiratory resistance and peak pressure comparable to a fourth EPAP device: Provent Sleep Apnea Therapy, which is indicated for all severities of obstructive sleep apnea (unpublished data). The side vent in the Optipillows EPAP mask is 12 mm long, and would be easy to estimate if it is approximately 50% or 75% closed, and somewhere in between. The patient can estimate 60% resistance on the Optipillows EPAP mask (approximately equal to Theravent and Bongo), and make adjustments thereafter based on their comfort. It is not necessary to be exact, because each patient generates pressure proportional to the size of their tidal volume, as shown in Figs 5. More importantly, the patient can select a resistance to get the most comfort and benefit.

We have taken much care to make sure we calibrate the pressure and flow before each experiment, and therefore, our measurements are accurate. The resistances of other devices are slightly different from the values that we calculated. For example, the resistance of Theravent calculated at 100 ml/sec was reported to be 30 cmH2O/L/sec [13]. Our measurements suggest that it is closer to 13 cmH2O/L/sec. It is possible that the reported values of resistance for other EPAP devices were obtained using different measuring tools to calibrate the pressure and flow and thus may account for some of the differences. Bongo Rx’s resistance was reported in an abstract [14].The values that were reported in the abstract were done at different flow rates, but are consistent with the values in the present study. For example at a flow of 166 ml/ sec (10 L/min), they report a resistance value of 17.4 cmH2O/L/ sec for Bongo, which is within the values reported in this study at 100 and 200 ml/sec (Table 1). The resistance is usually reported as the pressure generated at a flow of 100 ml/sec, but such value may be unreliable because the resistance at another flow rate would yield a different value, such as at 200 ml/sec. The slope of the pressure-flow curve is a much more reliable value to describe the device’s resistance. We estimated the slope from the data in the abstract between 166 and 333 ml/sec, to be 50 cmH2O/L/sec, which is consistent with the values in Table 1. More importantly, in this study, we used one calibration method to test all three EPAP devices side by side. Thus the differences between the EPAP devices that we report in this study are accurate. We estimated that when the side vent is closed midway between 50 and 75% on Optipillows (approximately 60%), the resistance and expiratory pressure were equal to Theravent or Bongo Rx. The Optipillows EPAP Mask has a resistance range of 2 to 75 cmH2O/L/sec and can be easily adjusted in small steps. It can be adjusted to 75% to make it equivalent to Provent. As importantly, a new user of EPAP devices may not be able to adapt to the high resistance of the EPAP device, but Optipillwos EPAP Mask allows the patient to use minimal resistance while trying to adapt to the EPAP device.

 

REFERENCES

1. Abbey NC, Cooper KR, Kwentus JA. Benefits of nasal CPAP in obstructive sleep apnea is due to positive pharyngeal pressure. Sleep. 1989; 12: 420-422.
2. Berry RB, Kryger MH, Massie CA. A novel nasal expiratory positive airway pressure device for the treatment of obstructive sleep apnea: a randomized controlled trial. Sleep. 2011; 34:479-485.
3. Kryger MH, Berry RB, Massie CA. Long term use of a nasal expiratory positive airway pressure (EPAP) device as a treatment for obstructive sleep apnea. J Clin Sleep Med. 2011; 7: 449-453.
4. Rosenthal L, Massie CA, Dolan DC, et al. A multicenter, prospective study of a novel nasal EPAP device in the treatment of obstructive sleep apnea. Efficacy and 30-day adherence. J Clin Sleep Med. 2009; 5: 532-537.
5. Walsh JK, Griffin KS, Forst EH, Ahmed HH, Eisenstein RD, Curry DT, et al. A convenient expiratory positive airway pressure nasal device for the treatment of sleep apnea in patients non-adherent with continuous positive airway pressure. Sleep Med. 201; 12: 147-152.
6. Mahadevia Ak, Onal E, Lopata M. Effects of expiratory positive airway pressure on sleep-induced respiratory abnormalities in patients with hypersomnia-sleep apnea syndrome. Am Rev Respir Dis. 1983; 128: 708-711.
7. Wu H, Yuan X, Zhan X, Li L. A review of EPAP nasal device therapy for obstructive sleep apnea syndrome. Sleep Breath. 2015; 19: 769-774.
8. Riaz M, Certal V, Nigam G, Abdulatif J, Zaghi S,et al. Nasal Expiratory positive airway pressure devices (Provent) for OSA: A systematic review and meta-analysis. Sleep Disord. 2015: 734798.
9. Schiza S, Mermigkis C, Bouloukaki I. Expiratory positive airway pressure (EPAP) nasal device therapy: a welcome addition to obstructive sleep apnea syndrome therapy. Sleep Breath. 2015; 9: 775-776.
10.Rossi VA, Winter B, Rahman NM, et al. The effects of Provent on moderate to severe OSA during continuous positive airway pressure therapy withdrawal: a randomized controlled trial. Thorax. 2013; 68: 854-859.
11.Hakim TS, Bosco G and Camporesi EM. Adaptation to EPAP Devices is Initially More Difficult than to CPAP: What is the Reason?. J Sleep MedDisord. 2020; 6: 1101.
12.Patel AV, Hwang D, Masdeu MJ, et al. Predictors of response to a nasal expiratory resistance device and its potential mechanisms of action for treatment of obstructive sleep apnea. J Clin Sleep Med. 2011: 7: 13-22.
13.Doshi R, Westbrook P. A novel non-prescription nasal device (Theravent) to treat snoring. Sleep Diagnosis Therapy. 2012: 7: 1-5.
14.Gay P. A comparison of two nasal expiratory PAP devices for treatment of OSA. Chest Annual Meeting, New Orleans. 2019
15.Hakim TS, Camporesi EM . A new expiratory resistance device fortreatment of snoring and mild obstructive sleep apnea. J Sleep Disor Treat Care. 2018: 7.
16.Colrain IM, Brooks S, Black JA. A pilot evaluation of a nasal expiratory resistance device for the treatment of obstructive sleep apnea. J Clin Sleep Med. 2008; 4: 426-433.

Hakim TS, Bonnetti M, Bosco G, Camporesi EM (2021) EPAP Devices Optipillows, Bongo Rx, and Theravent: Flow Resistance and The Pressures They Generate. J Sleep Med Disord 7(1): 1119.

Received : 13 Apr 2021
Accepted : 24 Apr 2021
Published : 26 Apr 2021
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X