Patient-Specific Growth-Rod Distraction Frequency may Reduce Chances of Rod Fracture in Juvenile Idiopathic Scoliosis

Research Article | Open Access

  • 1. Department of Bioengineering and Orthoapedic Surgery Engineering Center for Orthopaedic Research Excellence (E-CORE), University of Toledo, USA
  • 2. Department of Orthopaedics, Primus Super Speciality Hospital, India
+ Show More - Show Less
Corresponding Authors
Aakash Agarwal, Engineering Center for Orthopaedic Research Excellence (E-CORE), Department of Bioengineering and Orthoapedic Surgery, University of Toledo, 2801 West Bancroft Street, Toledo, OH 43606, USA
Abstract

Study design: A finite element study to establish the relationship between a patient’s curve flexibility (curve correction under gravity) in juvenile idiopathic scoliosis and the required distraction frequency to avoid rod fracture as a function of time.

Objective: To perform a parametric analysis using a juvenile scoliotic spine model (single mid-thoracic curve with the apex at the eighth thoracic vertebra) and establish the relationship between curve flexibility (curve correction under gravity) and the distraction interval that allows a higher factor of safety for the rods.

Method: The parametric study was performed by varying the material properties of the disc and hence altering the axial stiffness of a scoliotic spine model. The stresses on the rod were found to increase with increased axial stiffness of the spine, and consequently also increased the required optimal frequency to achieve a factor of safety of two for growth rods.

Results and conclusions: The current study demonstrates the possibility of translating fundamental information from finite element modeling to the clinical arena for mitigating the occurrence of growth rod fracture, i.e., establishing a relationship between optimal distraction interval and curve flexibility (curve correction under gravity).

Citation

Agarwal A, Jayaswal A, Agarwal AK, Goel VK (2016) Patient-Specific Growth-Rod Distraction Frequency may Reduce Chances of Rod Fracture in Juvenile Idiopathic Scoliosis. Ann Orthop Rheumatol 4(3): 1073.

Keywords


•    Growth rods
•    Scoliosis
•    Idiopathic scoliosis
•    Rod fracture
•    Distraction frequency
•    Distraction

INTRODUCTION

About 2.5% of children across all age groups have scoliosis [1]. The Scoliosis Research Society defines scoliosis as a lateral deviation of the normal vertical line of the spine which, when measured by X-ray, is greater than 10°. Idiopathic scoliosis is the most common deformity, and it represents 80–85% of all scoliosis cases [2]. Progression of scoliosis in children poses a substantial challenge for spinal surgeons. These young patients are undergoing active growth, and early fusion of any kind would stunt their growth and adversely affect their quality of life. However, if left untreated, a major curve progression becomes imminent, with chances of respiratory insufficiency. This has led to the advent of growth-friendly surgical management of early scoliosis, which aims to avoid, delay, or limit spinal fusion.

Distraction-based dual growth rods are the most commonly used growth-friendly surgical instruments [3,4]. In a typical growth-rod implant surgery, two rods (proximal and distal) are attached along the two lateral sides of the spine using bone anchors. Bone anchors commonly used in posterior spinal surgery can be classified into three categories: (1) hooks (infraor supralaminar hooks, transverse process hooks,andhook-end plate-screw combinations), (2)wires or bands (sublaminar wires, spinous process wires,and transverse process wires), and (3) pedicle screws. Among these, pedicle screws are the strongest and most widely used anchors for the foundation [5].

Initial surgery corrects the curve by about 50% [6-8]. A regular construct lengthening (at six months to a year) is continued for a period of five to ten years after implantation, until the spine’s longitudinal growth stops. During such lengthening surgeries, the proximal and distal rods at each side are distracted apart. The position of the rods after distraction is maintained using a tandem connector [9]. In contrast, magnetically controlled growth rods have a single rod on either side and can be distracted (lengthened) noninvasively.

Despite its advantages, the major failure seen with this instrumentation is growth rod fracture. Growth rod fractures occur in 15% of patients treated with growing rods [4,10,11]. Thompson et al., reported a 29% (two out of seven patients) complication rate of growth rod breakage[12]. Klemme et al., reported 12 rod breakages in 67 patients [13]. In 2005, Akbarnia et al., reported two rod breakages among 23 patients [14]. Yang et al. found that the risk of rod fracture increases with single rods, stainless steel rods, and smaller diameter rods. They also found that rod fracture was more prevalent among patients with preoperative ambulation ability [4]. Nevertheless, rod fracture also occurred in non ambulatory patients. Some researchers also believe that the distraction forces applied are so high that it stimulates growth rather than sustaining it [15]. Sakai et al., presented a case study in which they observe an 8.3-mm gap between T11-T12 immediately after distraction. This space was remarkably wider than the adjacent disc spaces, raising the suspicion of a distraction phenomenon (a severed union between soft and hard tissue), which was confirmed with computed tomography (CT) [16]. Following this, Agarwal et al. [17-19], reported results from a series of studies that shed light on the relationship between distraction (magnitude and frequency) and the residual stresses it generates on the rods. They found that higher distraction forces lead to higher residual stresses and increases the propensity of rod fracture. Additionally, they reported that frequent distraction (utilizing noninvasive procedure) with a lower distraction magnitude could reduce such occurrences. The objective of the current study is to perform a parametric analysis using a juvenile scoliotic spine model (a single mid-thoracic curve with the apex at the eighth thoracic vertebra) to establish the relationship between curve flexibility (curve correction under gravity) and the distraction interval that allows a higher factor of safety for the rods.

METHODOLOGY

This study uses a finite element model of a representative juvenile scoliotic spine (single mid-thoracic curve with the apex at the eighth thoracic vertebra) to run a parametric analysis. As described in previous publications, a normal juvenile spine was used as the foundation to produce multiple representative juvenile scoliotic spines. In this process, a CT was taken of a typical nine-year-old juvenile patient, and the vertebral body and intervertebral height were recorded [17]. Next, a validated T1-S1 normal adult spine model was scaled down to 71% of its original size to represent a juvenile spine, based on the literature data. However, the ratio of vertebral body to intervertebral height in an adult spine differs from that of a juvenile. For this reason, the mesh of this scaled-down finite element model of a normal spine was altered using ABAQUS (Dassault Systèmes, Simulia Inc., Providence, RI) to personalize the vertebral body and intervertebral body height to that of the heights recorded in the juvenile CT data. Thereafter, the desired scoliotic spine was generated with a custom script (MATLAB, Natick, MA UNITED STATES), utilizing polynomial transformations of finite element nodes.

Growth rods were simulated in the FE spine model with eight 4.5-mm titanium alloy (Ti6Al4V) pedicle screws and four 4.5- mm titanium alloy (Ti6Al4V) rods (two distal and two proximal) (Figure 1). Four out of eight pedicle screws were anchored bilaterally at the pedicles of the upper two vertebral foundations; the rest were placed bilaterally at the pedicles of the lower two vertebral foundations. The pedicle screws were kinematically coupled to the pedicles in all three degrees of freedom. The proximal rods were tied bilaterally to their respective ipsilateral proximal pedicle screws, and the distal rods to their ipsilateral distal pedicle screws. The tandem connection was simulated by kinematically coupling the ipsilateral free ends of the rods in all three degrees of freedom. As the model was simulating the postimplantation serial distractions, the initial curve was given a Cobb angle of 35°, kyphosis of 38°, and lordosis of 39°.

After the representative scoliotic juvenile finite element model was developed, all meshed regions were assigned juvenile spine material properties [18,19] (Figure 1). The follower load technique was chosen and simulated to account for the load at different vertebral levels due to upper body mass and muscle contraction [20]. The spine was loaded with 14% body weight at T1, following a 2.6% body weight increase between succeeding vertebrae. The given proportions were used to calculate the follower load for the current nine-year-old patient’s spine with a mean weight of 22 kg. The boundary condition consisted of restraining the inferior surface of the S1 vertebra in all degrees of freedom based on previously approved methodology for growth simulation in finite element modeling [20]. Unlike an adult spine, a juvenile spine has a certain rate of longitudinal growth, which is affected per the Hueter-Volkmann principle. This altered growth rate is captured in the following empirical equation:

G G= ″ + − ″ [ ( )]

where G is the actual growth strain, G″ is the mean baseline growth strain (at a given age), σ is the actual compressive stress on the growth plate (in MPa), σ is the mean baseline stress on the growth plate for the intact spine (in MPa), and b is equal to 1.5MPa-1 for vertebrae. For the intact model, G is equal to G″, and G″ is equal to 0.035 per six months for a nine-year-old child’s spine, according to the published literature [19].

In addition to growth, another clinically observed phenomenon, autofusion, was also incorporated into the model. As previously shown, the forces required to achieve distraction increase with subsequent distraction [21]. This is attributed to autofusion at the spinal segments and is an important aspect included in the current models. In brief, the phenomenon was incorporated into the models by increasing the stiffness of the spine as a function of time, using the available data on diminished lengthening on subsequent distractions [17].

Parametric analysis

The simulation was performed by varying the stiffness of the spine and applying distraction up to a maximum stress of 255 MPa. Thereafter, optimal distraction force was applied to each model with different stiffness values: 8.2 N/mm, 12.5 N/mm, 13.8 N/mm, 14.7 N/mm, 16.2 N/mm, 18.2 N/mm, 20.2 N/mm, 21.9 N/mm, 23.1 N/mm, and 25.5 N/mm. A stress of 255 MPa was chosen to keep the factor of safety equal to two and to account for possible stress concentrations and notching in implanted titanium rods due to surgical steps, as the fatigue strength of titanium alloy (Ti6Al4V) is 510 MPa. The stiffness was varied by increasing the modulus of elasticity of nucleus pulposus and annulus fibrosus. As described in previous publications [17,18], the slopes of increase in the spine’s longitudinal stiffness (in tension) with respect to Young’s modulus (N/mm2 ) of nucleus pulposus and shear modulus (N/mm2 ) of annulus fibrosus were 0.55 mm-1 and 7.2 mm-1, respectively. Data recorded during the simulation included the percentage of Cobb’s angle correction by gravity (representing unconstrained traction), axial stiffness, and the distraction interval required to achieve stresses limited to 255 MPa.

RESULTS AND DISCUSSION

Figure (2) shows the relationship between the axial stiffness and the percentage correction in Cobb’s angle due to gravity alone. As expected, it shows that the unconstrained correction under gravity is higher for a flexible curve and lower for a rigid curve. Figure (3) shows the relationship between the percentage correction in Cobb’s angle due to gravity alone and the required distraction interval for limiting the maximum von Mises stress to 255 MPa on the growth rods. The distraction interval required to limit the stresses to the selected nominal value reduces with increased stiffness of the spine. Please note that these values are for only the first distraction interval following the initial surgery.

Figure (4) shows how the distraction interval requirement changes with autofusion to keep the factor of safety at two; that is, the maximum von Mises stresses are still equal to 255 MPa. It can be seen that the distraction interval reduces for each model as the spine becomes stiffer with time (autofusion). Hence, an optimal distraction frequency is a time-dependent variable. Figure (5) shows the frequency of distraction that must be achieved to keep the maximum von Mises stress under the specified factor of safety. This represents the correlation that exists between the initial Cobb’s angle correction and the dynamic frequency that is required to keep the stresses on the rods under limit. In both figure (4) and figure (5), the model simulated with 0.2% Cobb’s angle correction always resulted in stresses higher than 255 MPa. A patient with such high stiffness would require a technology that can sense the stress and lengthen automatically.

Rod fracture is a common complication of growing-rod treatments and has been highlighted in several studies [4,12-14]. Therefore, lowering the stresses on the rods will help to reduce the occurrence of failure. Spinal stiffness varies among patients and changes after surgery (autofusion). Based on the extent of the patient’s spinal stiffness, even the lowest stresses (the optimal distraction force) generated may result in growth rod fracture. A proposed hypothesis for this situation would be to change the distraction frequency, therefore lowering the magnitude of the required optimal distraction force, according to the patient’s spinal stiffness.

Optimal distraction force refers to distraction force required to produce height gain equal to normal height as observed in the non instrumented spine [17-19]. The distraction frequency and distraction forces are integrated; i.e., for every distraction frequency there is an optimal distraction force. This optimal value is equivalent to the force required to distract the spine for the specific distraction interval. It is because of this reduction in optimal distraction force that the stresses on the rod reduce. Therefore, the distraction force of approximately zero magnitude would ideally mean a growth rod technology that can sense the change in compressive stresses and lengthen automatically as a negative feedback mechanism.

At present, the growth rod technique is not a single-surgery technique: several invasive distractions must follow the main surgery, and the child suffers extreme morbidity and discomfort as a result. For example, a five-year-old child implanted with growth rods would undergo 10 to 14 consecutive distraction surgeries. Moreover, there have been many instances of failure. With traditional growth rods, changing the frequency of distraction wasn’t an option because of its invasive nature. A higher frequency of distraction (1 year) lowers the growth potential. However, new growth rod technology in the spine industry now means that serial distractions could be achieved noninvasively. The results from the sensitivity study help translate this information into clinical practice. We developed and simulated a representative scoliotic model varying in material properties (to account for stiffness variation among scoliotic patients) to establish a relationship between axial stiffness of the spine, percentage correction in Cobb’s angle due to gravity, and required distraction interval, for a factor of safety equal to two. The study results show that by measuring the percentage correction before the surgery, a specific distraction interval could be chosen based on the required factor of safety for the growth rods. The idea behind using a graphical representation instead of a single number for the ideal distraction frequency is based on the existence of variance in spinal stiffness among the patients.

As with all surgical procedures, patient selection is an important factor that affects the efficacy of any technique. The ideal frequency for one patient may not be suitable for the other. Therefore, this study refrains from selecting a particular distraction frequency as a cautionary measure. Furthermore, ideal distraction forces change for a given patient with time (function of autofusion).We do recommend higher frequencies of distraction whenever possible with the optimal distraction force as an ideal choice to reduce von Mises stresses on the rods, but properly understanding assumptions is a prerequisite before using these numbers for the patients. These specific values and recommendations are limited to the nine-year-old age group based on typical anatomical considerations (intervertebral disc height). Additionally, the growth rods used in this study were 4.5 mm in diameter and used Ti6Al4V as their material type [17].

There area few limitations to the current study. First, the surgical procedure of initial rod attachment was not simulated; and the scoliotic curves are considerably larger before the surgery (more than 45–50°). However, given their nonstructural nature, they exert only 5–8% of total von Mises stress for about 40–50% of correction [9]. Second, due to lack of awareness about how suboptimal distraction forces and distraction frequencies itself results in rod fracture, there isn’t a clinically available direct method to apply the findings of this study (i.e., the ability to measure loads vs. distraction during surgery). Some indirect methods include presurgical planning using finite element modeling with the patient’s standing and side-bending radiographs. Nevertheless, new technologies are being developed that could either provide or calculate the optimal distraction based on sensors incorporated in the device [22,23].

 

CONCLUSION

The current study describes a method to translate fundamental information from finite element modeling to the clinical arena for mitigating the occurrence of growth rod fracture, and it establishes a relationship between optimal distraction interval and curve rigidity. The study results show that to reduce the stresses on the rods, the distraction interval can be shortened as a function of time-consecutive ones per the patient’s requirements (measured from initial flexibility measurement).This is not feasible with traditional growth rods, but other noninvasive distraction methods, such as magnetically controlled growth rods, present great potential.

REFERENCES

1. Lonstein JE, Carlson JM. The prediction of curve progression in untreated idiopathic scoliosis during growth. J Bone Joint Surg Am. 1984; 66: 1061-1071.

2. Gillingham BL, Fan RA, Akbarnia BA. Early onset idiopathic scoliosis. J Am Acad Orthop Surg. 2006; 14: 101-102.

3. Akbarnia BA, Mundis GM Jr, Salari P, Yaszay B, Pawelek JB. Innovation in growing rod technique: a study of safety and efficacy of a magnetically controlled growing rod in a porcine model. Spine (Phila Pa 1976). 2012; 37: 1109-1014.

4. Yang JS, Sponseller PD, Thompson GH, Akbarnia BA, Emans JB, Yazici M, et al. Growing rod fractures: risk factors and opportunities for prevention. Spine (Phila Pa 1976). 2011; 36: 1639-1644.

5. Mahar AT, Bagheri R, Oka R, Kostial P, Akbarnia BA. Biomechanical comparison of different anchors (foundations) for the pediatric dual growing rod technique. Spine J. 2008; 8: 933-939.

6. Akbarnia BA. Selection of methodology in surgical treatment of adolescent idiopathic scoliosis. Orthop Clin North Am. 1988; 19: 319- 329.

7. Akbarnia BA, Cheung K, Noordeen H, Elsebaie H, Yazici M, Dannawi Z, et al. Next generation of growth-sparing techniques: preliminary clinical results of a magnetically controlled growing rod in 14 patients with early-onset scoliosis. Spine (Phila Pa 1976). 2013; 38: 665-670.

8. Akbarnia BA, Emans JB. Complications of growth-sparing surgery in early onset scoliosis. Spine (Phila Pa 1976). 2010; 35: 2193-2204.

9. Flynn JM, Tomlinson LA, Pawelek J, Thompson GH, McCarthy R, Akbarnia BA, et al. Growing-rod graduates: lessons learned from ninety-nine patients who completed lengthening. J Bone Joint Surg Am. 2013; 95: 1745-1750.

10. Bess S, Akbarnia BA, Thompson GH, Sponseller PD, Shah SA, El Sebaie H, et al. Complications of growing-rod treatment for early-onset scoliosis: analysis of one hundred and forty patients. J Bone Joint Surg Am. 2010; 92: 2533-2543.

11. Skaggs D, Akbarnia B, Flynn J, Myung K, Sponseller P, Vitale M. Classification of treatment of early-onset scoliosis. In second international congress of early onset scoliosis and growing spine, Montreal. 2008.

12. Thompson GH, Akbarnia BA, Campbell RM Jr. Growing rod techniques in early-onset scoliosis. J Pediatr Orthop. 2007; 27: 354-361.

13. Klemme WR, Denis F, Winter RB, Lonstein JW, Koop SE. Spinal instrumentation without fusion for progressive scoliosis in young children. J Pediatr Orthop. 1997; 17: 734-742.

14. Akbarnia BA, Marks DS, Boachie-Adjei O, Thompson AG, Asher MA. Dual growing rod technique for the treatment of progressive earlyonset scoliosis: a multicenter study. Spine (Phila Pa 1976). 2005; 30: S46-S57.

15. Olgun ZD, Ahmadiadli H, Alanay A, Yazici M. Vertebral body growth during growing rod instrumentation: growth preservation or stimulation? J Pediatr Orthop. 2012; 184-189.

16. Sakai DS, LA Tomlinson, JP Dormans. Distraction Phenomenon After Lengthening of Spinal Growing Rods. JBJS Case Connector. 2012; 2: e45.

17. Agarwal A, AK Agarwal, A Jayaswal, V Goel. Smaller Interval Distractions May Reduce Chances of Growth Rod Breakage Without Impeding Desired Spinal Growth: A Finite Element Study. Spine Deformity. 2014; 2: 430-436.

18. Agarwal A, AK Agarwal, A Jayaswal, and VK Goel, Effect of distraction force on growth and biomechanics of the spine: a finite element study on normal juvenile spine with dual growth rod instrumentation. Spine Deformity. 2014; 2: 260-269.

19. Agarwal A, Zakeri A, Agarwal AK, Jayaswal A, Goel V. Distraction magnitude and frequency affects the outcome in Juvenile Idiopathic Patients with Growth Rods: Finite Element Study Using a Representative Scoliotic Spine Model. The Spine Journal. 2015; 15: 1848-1855.

20. Shi L, Wang D, Driscoll M, Villemure I, Chu WC, Cheng JC, et al, Biomechanical analysis and modeling of different vertebral growth patterns in adolescent idiopathic scoliosis and healthy subjects. Scoliosis. 2011; 6: 11.

21. Noordeen HM, Shah SA, Elsebaie HB, Garrido E, Farooq N, Al-Mukhtar M. In vivo distraction force and length measurements of growing rods: which factors influence the ability to lengthen? Spine (Phila Pa 1976). 2011; 36: 2299-2303.

22. Abolaeha OA, Smart Growing Rod Device for the Treatment of Early Onset Scoliosis. University of Dayton. 2013.

23. Hoven L, Oude Weernink A, Verboom S, Wel H. an innovative distraction method for patients with spondylocarpotarsal synostosis syndrome: calculation of the continuous distraction force and the design of a continuous distraction device. 2015

Received : 06 Jul 2016
Accepted : 01 Aug 2016
Published : 02 Aug 2016
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X