Loading

Annals of Pediatrics and Child Health

Performance of Multiplex Detection Method of IgM Class Antibodies against Toxoplasma gondii, Rubella and Human Cytomegalovirus

Research Article | Open Access Volume 10 | Issue 7 |

  • 1. Instituto de Pesquisa e Desenvolvimento (IP&D), Universidade do Vale do Paraíba (UNIVAP), Brasil.
  • 2. Intercientífica, Parque Tecnológico UNIVAP, Brasil
  • 3. Instituto de Ciência e Tecnologia, Universidade Estadual Paulista (UNESP), Brasil
+ Show More - Show Less
Corresponding Authors
Rainara Moreno Sanches de Almeida, Instituto de Pesquisa e Desenvolvimento (IP&D), Universidade do Vale do Paraíba, Av. Shishima Hifumi 2911, São José dos Campos, São Paulo, Brasil, Tel: 55(12)988036744
Abstract

Serological diagnosis during neonatal screening is crucial in disease prevention. Among the infectious diseases, the most common are toxoplasmosis, rubella, and cytomegalovirus. Traditional diagnostic methods are used to detect a single infectious agent per test. The use of multiplex detection methods increases productivity and reduces the amount of material used, resulting in a more efficient test from a technical, environmental, and economic point of view. The study’s objective was to evaluate the performance of a new diagnostic method aimed at neonatal screening using the multiplex platform of magnetic microspheres from the company Luminex Corporation. For this, tests were carried out for analytical validation of the diagnostic product developed following the rules of the National Health Surveillance Agency (ANVISA) of Brazil. The parameters evaluated were repeatability, reproducibility, linearity, robustness, high dose, minimum detection limit, and analytical specificity. All data obtained met the acceptance criteria of RDC 166/17 of 2017 for the use of the diagnostic product in the national territory. Repeatability and reproducibility tests showed a CV of less than 15% between replicates of the same operator and different operators. The kit showed linearity throughout the operating range with R2 above 0.990, and no effect of high-efficiency dose was observed in the chosen working dilution. In addition, the kit did not show interference from the matrix with the results, and it was observed that small and deliberate changes in the incubation time of each reagent did not have a significant effect on the data obtained.

Keywords

Neonatal screening, Multiplex assay, Magnetic beads, Analytical validation

ABBREVIATIONS

ANVISA: National Health Surveillance Agency; RDC: Resolution of the collegiate board; CV: Variation Constance; TOX: Toxoplasmosis; RUB: Rubella; CMV: Cytomegalovirus.

Citation

de Almeida RMS, Ogawa GM, Siqueira Custódio PR, Socorro Sampaio MA, da Silva NS, et al. Performance of Multiplex Detection Method of IgM Class Antibodies against Toxoplasma gondii, Rubella and Human Cytomegalovirus. Ann Pediatr Child Health 2022; 10(7): 1295.

INTRODUCTION

The detection of antibodies against infectious diseases is an activity widely used in several situations. Among the available laboratory detection techniques, the widely used method is the Enzyme-Linked Immunosorbent Assay (ELISA) which detects a single infectious agent by an assay based on antigen-antibody binding [1].

Although commonly used, traditional methods of serological diagnoses, such as ELISA, have the disadvantage of requiring more significant amounts of samples and time for analysis since each analyte of interest requires a specific assay.

New technologies were developed to optimize diagnostic methods, among which stands out is the xMAP® technology from Luminex Corporation, which allows the simultaneous analysis of several parameters in the same biological sample using microspheres [2]. Neonatal screening is necessary to prevent severe infections from affecting the fetus and neonate.

Several infectious agents could be vertically transmitted from mother to fetus via the transplacental route or direct contact during childbirth. TORCH’s acronym represents the most common infections, including toxoplasmosis, rubella, and cytomegalovirus [3,4].

A multiplex kit was developed for detecting antibodies against infectious diseases using xMAP® technology of magnetic microspheres, aiming to meet the need to develop effective diagnostic methods for neonatal screening. The kit is intended to detect multiple and simultaneous anti-toxoplasmosis, anti-rubella, and anti-human cytomegalovirus IgM antibodies in newborn samples.

Developing a new diagnostic product using microspheres demands studies on the performance and other parameters that an in vitro diagnostic kit must present to be used in Brazil. The National Health Surveillance Agency (ANVISA) created and evaluated these performance parameters.

For the validation and registration of the in vitro diagnostic kit and its entry into the national market, different studies are required following the rules of the Resolution of the Collegiate Board (RDC) No. 166 of July 24, 2017, published by ANVISA, which establishes the criteria for validation of analytical methods.

The objective of the present study was to evaluate the performance following the analytical validation criteria of a new diagnostic method for multiple and simultaneous detections of human antibodies of the IgM class against Toxoplasma gondii (TOX), rubella (RUB) and cytomegalovirus (CMV) in the field of newborn screening.

MATERIALS AND METHODS

Performance analysis

For the use of new diagnostic methods developed in Brazil, the product must meet ANVISA’s analytical methods validation criteria. RDC No. 166 of July 24, 2017, applies to analytical methods used in pharmaceutical supplies, medicines, and biological products in all their production stages. Some validation parameters involve precision (repeatability and reproducibility), Analytical Specificity, High Dose, Linearity, Limit of Detection (LOD), and Robustness [5]. All tests carried out in the present study were based on the acceptance criteria for registering a new diagnostic product in Brazil and its use in the national territory.

Characterization and validation of clinical samples used.

Reference samples purchased from Seracare Life Science were used for the three parameters. These samples are Seracare’s antibodies panel: antibody anti-T. gondii antibody panel (AccuSet™ Toxoplasmosis Performance Panel, catalog number 0820- 0321, lot 10344475), antibody IgM carrier sample Rubella (Seracare Rubella IgM positive plasma, catalog number DS-674 -M, lot 9254456), and antibodies carrying IgM anti-CMV (Seracare CMV IgM Positive plasma, catalog number DS-626-M, lot BM217341), all samples were supplied with a certificate of analysis and reference results from other methodologies.

Coupling of antigens to magnetic microspheres

Couplings were performed following the protocol provided by Luminex Corporation. (xMAP® Cookbook 5th Edition). First, the xMAP® magnetic microsphere solutions were added in different low-adhesion USA tubes (catalog 1415-2600). Microspheres 12 (MC10012), 45 (MC10045), and 72 (MC10072) were used for Toxoplasma gondii, Rubella, and Cytomegalovirus antigens, respectively. The microspheres were washed with NaH2PO4 (activation buffer) and incubated with EDC and Sulfo NHS for 20 minutes, protected from light, and on rotation in a tube shaker at 700 rpm. Then the microspheres were washed with PBS-TBN and incubated with the respective antigens for 2h. After incubation, the microspheres were washed and kept in rotation for 20 minutes with block buffer (PBS/1%BSA). Then the microspheres were centrifuged, resuspended in PBS-TBN buffer, and stored under refrigeration.

For reasons of commercial secrecy, the antigens and concentrations used cannot be made available. All couplings make up reagent 2 of the NeoMAP® 3plex IgM kit from Intercientífica, developed in this study, with registration number 80173700020 at ANVISA.

Assay protocol

First, the samples were eluted in 1.5 ml tubes with 200 µL of Elution Reagent (PBS, Tween20, 20% Azide, BSA, and E. coli extract) at a dilution of 1:200 for 60 minutes under agitation at 700rpm. After incubation, the assay plate was prepared by adding 25 µL of the pool of magnetic microspheres coupled to Toxoplasmosis, Rubella, and Cytomegalovirus antigens in each well. Then a washing step was performed by adding 50 µL of Washing Solution (PBS, Tween20, and 20% Azide). Next, 50 µL of the eluate was added to the wells and incubated for 90 minutes under agitation at 700 rpm, protected from light. After incubation, two washing steps were performed, and 50 µL of Anti-human IgM solution labeled with Biotin was added for 30 minutes under agitation on the plate shaker and protected from light. After the incubation, a new washing step was performed, and 50 µL of the Phycoerythrin solution conjugated with Streptavidin was added and maintained under agitation for 30 minutes in the plate shaker at 700 rpm. Then the solution was removed, 50 µL of Washing Solution was added and kept for 1 minute in the plate shaker, taken for reading in the Magpix equipment with analysis in the xPONENT® Software. For reasons of commercial secrecy, the antigens and concentrations used cannot be made available. As well as the Anti-Human IgM and Phycoerythrin conjugated with Streptavidin catalogs. All reagents make up the NeoMAP® 3plex IgM kit developed in this study and with registration 80173700020 at ANVISA.

Analysis of results

In all tests, the constant of variation (CV) between replicates of each sample was verified. For this, the formula below was used (Figure 1). In addition, GraphPad Prisma and Action Stat software were used for the statistical analysis of the data.

Figure 1 Formula for constant of variation (CV).

Figure 1: Formula for constant of variation (CV).

Measurement accuracy

The repeatability tests were carried out in the Research and Development laboratory of the company Intercientifica, and the reproducibility tests were carried out in the Quality Control laboratory in another sector of the same company, where all the equipment and the operator were different. The assays were carried out with nine high, medium, and low reactivity samples in triplicate and on other days.

Minimum Limit of Detection (LOD)

The LOD Limit of Detection must be demonstrated by obtaining the lowest amount of the analyte present in a sample that can be detected but not necessarily quantified under established experimental conditions. For this, 73 measurements of the blank were performed, with the blank consisting of all reagents, except for the sample. Statistical analysis was performed using the Action Stat software, which presented the minimum detection values for each parameter in the multiplex assay.

Analytical Specificity

Specificity was assessed by the method’s ability to unambiguously identify or quantify the analyte of interest in the presence of components that may be present in the sample, such as matrix components. The test protocol was carried out using the blood components after separating the plasma in triplicate to verify if matrix interference occurs because it is in the plasma that the antibodies are found. Data analysis was performed using the formula below (figure 2), which considers the results in Mean Fluorescence Intensity (MFI) and the blank.

Figure 2 Formula used for analytical sensitivity analysis.

Figure 2: Formula used for analytical sensitivity analysis.

Linearity

The linearity of the method was verified through tests with highly positive samples for each parameter in serial dilution 1:2 with 16 drops. After the trial, linear regression analysis was performed on each sample, evaluating the regression coefficient (R2). The defined performance target was R2 greater than 0.99 according to ANVISA’s acceptance criteria.

High Dose

The high dose prozone effect was verified using highly reactive samples for the three antigens. The test was carried out with samples in serial dilution 1:2 with eight drops starting with the dilution of the samples in 1:10.

Robustness

To verify whether small, deliberate changes in some steps cause changes in the reactivity of the samples. Tests were carried out with 10 minutes more and 10 minutes less than the standard incubation time of each step of the test protocol: sample elution, incubation with the microspheres, incubation with Antihuman IgM, and incubation with phycoerythrin. The acceptance criterion was a CV of less than 20% between samples with different incubation times compared to the reference (original time of each stage).

RESULTS

Measurement Accuracy

The repeatability test (Table 1), showed a CV of less than 15% in all samples, demonstrating no significant difference between triplicates for all analyzed parameters.

Table 1: Repeatability analysis.
TOX
Sample Well 1 Well 2 Well 3 M SD CV%
1 25737 27455 27450 27450 809 3
2 22508 25824 26491 25824 1742 7
3 8076 8626 8785 8626 304 3
4 3331 3929 4326 3929 409 10
5 2737 3514 3644 3513 400 11
6 3476 2967 3163 3163 210 7
7 370 376 319 370 26 7
8 346 336 324 336 9 3
9 287 203 254 254 34 13
RUB
Sample Well 1 Well 2 Well 3 M SD CV%
1 28094 30167 30621 30167 1100 4
2 26494 26115 25560 26115 380 1
3 20084 22046 21360 21360 813 5
4 15956 14871 15745 15745 470 3
5 10074 10350 9518 10074 346 3
6 6011 5815 5762 5815 107 2
7 3171 3113 2856 3113 137 4
8 1584 1542 1122 1542 209 13
9 773 732 668 732 43 6
CMV
Sample Well 1 Well 2 Well 3 M SD CV%
1 14934 15109 14604 14934 210 1
2 12386 12066 12469 12386 174 1
3 11111 12947 12580 12580 793 6
4 7422 7575 7471 7471 64 1
5 7546 8394 8094 8094 351 4
6 4402 4812 4066 4402 305 7
7 1236 1156 1151 1156 39 3
8 896 892 869 891 11 1
9 579 569 580 579 5 1
Abbreviations: M: medium; SD: Standard deviation; CV: Constance of variation.

Although the acceptance criterion used was 15%, it is common for low reactivity samples to present a relative standard deviation above the recommended one due to the intrinsic variability of the method resulting from the high dynamic range (0-100,000MFI).

In Table 2, it is possible to observe CV less than 15% in all parameters when comparing the tests performed by two operators in two different laboratories.

Table 2: Reproducibility Analysis.
TOX
Sample Well 1 Well 2 Well 3 M SD CV% M lab 1 M Lab 1 vs Lab2 SD CV(%)
1 22495 23908 22571 22571 649 3 27450 25010 2440 10
2 22222 23608 24019 23608 769 3 25824 24716 1108 4
3 8021 7750 7951 7951 115 1 8626 8288 338 4
4 4158 4277 4107 4158 71 2 3929 4043 114 3
5 2761 2775 2446 2761 152 5 3514 3139 377 12
6 2616 2536 2326 2536 122 5 3163 2849 314 11
7 639 646 672 647 14 2 724 685 39 6
8 323 302 325 323 11 3 370 347 24 7
9 352 380 308 352 30 8 336 344 8 2
RUB
Sample Well 1 Well 2 Well 3 M SD CV% M lab 1 M Lab 1 vs Lab2 SD CV(%)
1 33277 36123 32433 33277 1579 5 30167 31722 1555 5
2 28718 30542 29543 29543 746 3 26115 27829 1714 6
3 24138 24113 23756 24113 174 1 21360 22736 1376 6
4 18332 16409 18236 18236 885 5 15745 16990 1245 7
5 12423 12187 12073 12187 146 1 10074 11130 1056 9
6 7573 6998 7270 7270 235 3 5815 6543 727 11
7 3844 3891 3921 3891 32 1 3113 3502 389 11
8 1868 1852 1545 1852 148 8 1542 1697 155 9
9 934 868 828 868 44 5 732 800 68 8
CMV
Sample Well 1 Well 2 Well 3 M SD CV% M lab 1 M Lab 1 vs Lab2 SD CV(%)
1 15154 14815 14775 14815 170 1 12386 13601 1214 9
2 13316 16048 16051 16048 1289 8 12580 14314 1734 12
3 12101 12565 12182 12182 203 2 11259 11720 462 4
4 6373 6591 6306 6373 122 2 7471 6922 549 8
5 9202 8711 8484 8711 300 3 8094 8402 308 4
6 5041 4784 4372 4784 275 6 4402 4593 191 4
7 3385 3243 3306 3306 58 2 2771 3038 268 9
8 1188 1071 1115 1115 48 4 1156 1135 20 2
9 510 526 479 510 19 4 579 544 34 6
Abbreviations: Results from labs 1 and 2 with median, standard deviation, and CV of nine samples analyzed with the NeoMAP® 3plex IgM kit (MMedian, SD- standard CV – Constant Variation, Op- operator, Lab- laboratory).

Limit of Detection (LOD)

In all parameters, there was low reactivity in the blank, indicating no undesired reactivity of the elution and washing buffer with the microspheres coupled to the different antigens (Table 3).

Table 3: Limit of Detection in MFI of the NeoMAP® 3plex IgM Kit.
Limit of Detection (LOD)
  TOX RUB CMV
Average 36 49 41
Standard deviation 5 9 4
Degrees of liberty 71 71 71
Detection limit 44 65 47

Analytical specificity

Table 4 demonstrates that the reactivity of blood components in the absence of antibodies is below the detection limits of all parameters, indicating no significant reactivity of these components.

Table 4: Analytical specificity.
  TOX RUB CMV
Replica 1 2 0.5 0
Replica 2 0 -4.5 -5
Replica 3 1 1.5 -4

High Dose

In the figure below (Figure 3), it is possible to see that in the TOX parameter, a high dose effect occurs from the dilution of 1/40, causing a decrease in reactivity even with a higher concentration of antibodies. The RUB parameter occurs from the 1/160 dilution, and the CMV parameter from the 1/20 dilution.

Figure 3 Reactivity curve of the samples in relation to the dilution factor in the parameters of TOX, RUB, and CMV. (A) TOX, (B) RUB and (C) CMV.

Figure 3: Reactivity curve of the samples in relation to the dilution factor in the parameters of TOX, RUB, and CMV. (A) TOX, (B) RUB and (C) CMV.

Linearity

The response range is linear and occurs between 15 - 10,000 MFI with R2 and 0.99 correlation for TOX, 170 - 3,000 MFI with R2, 0.99 correlation for RUB, and 10 - 6,000 MFI with R2 and 0 correlation, 99 for CMV. At MFI values above or below these ranges, the test cannot correlate MFI with the concentration unit of the samples.

Table 5: Reactivity of samples incubated 80, 90 and 100 minutes in the elution step.
TOX 80min 90min 100min M. SD. CV (%)
1 615 623 615 615 4 1
2 21279 22669 21397 21396 629 3
3 1610 1629 2080 1629 217 13
4 507 518 572 518 28 5
RUB 80min 90min 100min M. SD. CV (%)
1 27770 25990 26267 26267 782 3
2 181 249 222 222 28 13
3 329.5 389 423 389 38 10
4 53 65 60 60 5 8
CMV 80min 90min 100min M. SD. CV (%)
1 453 552 457 457 46 10
2 3033 4540 3222 3222 670 21
3 9208 9778 12205 9778 1299 13
4 300 411 382 382 47 12
Abbreviations: M: Medium; SD: Standard deviation; CV: Constant of variation.

Robustness

Robustness studies indicate that variations of ten minutes plus or minus in incubations do not result in significant differences in the reactivity of samples with an overall CV of less than 15%. The results are in the tables below (Tables 5, 6, and 7).

Table 6: Reactivity of samples incubated 20, 30 and 40 m minutes in the secondary antibody step.
TOX 80min 90min 100min M. SD. CV (%)
1 640 623 698 640 32 5
2 18606 22669 22793 22669 1945 9
3 1717 1629 1907 1717 116 7
4 536 518 599 536 35 6
RUB 80min 90min 100min M. SD. CV (%)
1 25301 25990 27931 25989 1113 4
2 188 249 230 230 25 11
3 393 389 423 393 15 4
4 60 68 68 65 3 5
CMV 80min 90min 100min M. SD. CV (%)
1 493 552 508 507 25 5
2 2840 4540 4099 4098 720 18
3 10352 9778 10966 10351 485 5
4 386 411 419 411 14 3
Abbreviations: M: Medium; SD: Standard deviation; CV: Constant of variation.

 

Table 7: Reactivity of samples with incubation of 20, 30 and 40 minutes in the phycoerythrin step.
TOX 20min 30min 40min M. SD. CV (%)
1 727 623 566 623 67 11
2 19032 22669 22974 22669 1791 8
3 1803 1629 1816 1803 85 5
4 567 518 623 567 43 8
RUB 20min 30min 40min M. SD. CV (%)
1 25175 25990 28076 25990 1222 5
2 203 249 221 221 19 9
3 362 389 376 376 11 3
4 68 65 73 68 3 5
CMV 20min 30min 40min M. SD. CV (%)
1 519 552 456 518 40 8
2 3392 4540 4200 4200 482 12
3 9879 9778 12273 9879 1153 12
4 379 411 415 411 16 4
Abbreviations: M: Medium; SD: Standard deviation; CV: Constant of variation.

 

DISCUSSION

Since the introduction of neonatal screening, several technological advances have been used for the early detection of congenital conditions. Implementing new methodologies allows for a significant increase in the number of diseases screened, with the capacity to include up to fifty disorders [5].

Recent technological advances have developed platforms capable of multiplexing multiple molecular and immunological assays for high-throughput screening. It is a modern platform whose main advantages are saving time and reagents [5].

Among the new technologies, the multiplex platform of magnetic microspheres from the company Luminex Corporation allows the development of assay methods with up to 100 different analytes in the same well of the assay plate [6].

Analytes must be coupled to these microspheres to perform target detection. The coupling is carried out by the chemical interaction of the carboxyl groups on the microspheres’ surfaces with the proteins’ primary amines. These bonds are covalent and most often performed with antibodies or specific antigens [8,9].

In general, these microspheres are used in scientific research, and there needs to be more information about them in neonatal screening. Screening has the potential to prevent serious health problems, including death. Worldwide, neonatal screening programs have evolved from simple tests to comprehensive and complex systems capable of detecting more than 50 different conditions in different countries [9].

The screening program is used to track diseases and mainly to seek treatment and provide the family with resources so that those affected become healthy. In Brazil, more and more diseases are being implemented in the Unified Health System (SUS) screening programs, including the addition of infectious diseases such as Toxoplasmosis, Rubella, and Cytomegalovirus correspond to a high number of severe congenital infections [10].

With the expansion of screened diseases, developing new technologies that optimize and implement analyses in all regions, including the country’s most remote regions, is essential. For the use of these new diagnostic methods developed in Brazil, the product must meet ANVISA’s acceptance criteria for validation of analytical methods.

The precision of an analytical procedure expresses the proximity of agreement (degree of dispersion) between a series of measurements obtained from the same sample [11]. In this study, the repeatability and reproducibility assays performed for precision analysis showed a CV of less than 15% in all analyzes, demonstrating agreement between different tests with the same sample. The kit under development is considered semiquantitative. This definition occurs because the kit qualitatively delivers the results, that is, positive or negative. However, occasionally quantitative values can be used, mainly in the case of recollection of the same patient.

The MFI values of a result can be compared with results from a later collection (from the same individual) to verify whether there has been a significant increase or decrease in reactivity.

Thus, it is necessary to demonstrate that the MFI response of the product is proportional to the concentration of antibodies in the samples.

The linearity of the method is the ability to obtain results directly proportional to the concentration of the specific target antibody against the analyte of interest in the sample [12].

Our study verified linearity using samples with data on reactivity units. As these samples have unit values, it is possible to verify how the response occurs about the dose: MFI and AU/ mL or s/co ratio. The tests indicate that all tested parameters can produce a response (in MFI) corresponding to the dosage of IgM antibodies in the reference samples.

The high dose prozone effect was verified using highly reactive samples for the three antigens. In the assay, it was possible to observe an effect of a high dose at the highest concentrations with a decrease in reactivity. However, this decrease did not occur significantly with CV less than 15% when comparing the MFI values. Furthermore, although there is inhibition of antigenantibody binding, it is far from the dilution used in the study.

All parameters showed low LOD values, with 44, 65, and 47 MFI, respectively, for TOXO, RUB, and CMV. LOD corresponds to the lowest concentration of the analyte that can be detected, but not necessarily quantified, under established experimental conditions; this is because, usually, an assay cannot measure analyte concentrations down to zero [12].

Analytical specificity is the ability to unambiguously evaluate the analyte in the presence of other components that may be present in the samples, such as the matrix. The method under development must be able to produce a response only for the specific target [13].

In the study, the NeoMAP® 3plex IgM kit proved specific in all parameters, with no matrix reactivity in the assay performed only with the red blood cell concentrate. That is, the kit could generate interference-free signals on the multiplex platform. Validation is fundamental for the efficient operation of new analytical methods and must be carried out at all stages, from the raw materials used to the finished product. The phrase validation implies a feasibility demonstration activity and aims to demonstrate that the developed product meets the recommended acceptance criteria and is suitable for the proposed objectives [10,11]. In the present study, the kit developed followed all the criteria proposed for product registration in Brazil and its use in the national territory.

CONCLUSION

The repeatability and reproducibility tests showed a CV of less than 15% in all samples, demonstrating the reliability of the data obtained in all tests. High-dose assays showed that at higher concentrations, there is a tendency to inhibit antigen-antibody binding, requiring a defined dilution of 1:200. The obtained data presented performance compatible with the validation criteria of analytical methods of ANVISA, being approved for use in the country. The data generated in this study are contained in the product registration (Registration number 80173700020).

ACKNOWLEDGMENTS

Acknowledgments to the Coordination for the Improvement of Higher Education Personnel (CAPES) for the master’s scholarship that enabled the development of the study. Acknowledgments to the company Intercientifica for sharing the infrastructure of the Research and Development laboratory and for making the xMAP technology of magnetic microspheres available.

REFERENCES

1. Gan SD, Patel KR. Enzyme immunoassay and enzyme-linked immunosorbent assay. J Investigative Dermatol. 2013; 133: 1-3.

2. Stephen L. Multiplex immunoassay profiling. In: Methods in Molecular Biology. Humana Press Inc.; 2017; 169-176.

3. Chen L, Liu J, Shi L, Song Y, Song Y, Gao Y, et al. Seasonal influence on TORCH infection and analysis of multi-positive samples with indirect immunofluorescence assay. J Clin Lab Anal. 2019; 33.

4. García-Bermejo I, de Ory-Manchón F. Serological diagnosis of congenital infections and algorithms to improve diagnostic efficacy. Enferm Infecc Microbiol Clin. 2015; 33: 20-26.

5. ANVISA. Diário oficial da união órgão: ministério da saúde/agência nacional de vigilância sanitária/diretoria colegiada. 2017.

6. Sahai I, Marsden D. Newborn screening. Critical Reviews in Clinical Laboratory Sciences. 2009; 46: 55-82.

7. Khalifian S, Raimondi G, Brandacher G. The Use of Luminex Assays to Measure Cytokines. J Investigative Dermatol. 2015; 135: 1.

8. Dunbar SA. Applications of Luminex® xMAPTM technology for rapid, high throughput multiplexed nucleic acid detection. Vol. 363, Clín Chimica Acta. 2006; 71-82.

9. Mazhari R, Brewster J, Fong R, Bourke C, Liu ZSJ, Takashima E, et al. A comparison of non-magnetic and magnetic beads for measuring IgG antibodies against Plasmodium vivax antigens in a multiplexed beadbased assay using Luminex technology (Bio-Plex 200 or MAGPIX). PLoS One. 2020; 15.

10.Therrell BL, Padilla CD. Newborn screening in developing countries. Curr Opin Pediatr. Lippincott Williams and Wilkins. 2018; 30: 734- 749.

11. Mayuri D, Ravindranath S, Doltade M. Analytical Method Development and Validation: A Review. J Drug Delivery and Therapeutics [Internet]. 2019; 9: 563-570.

12. Rajam G, Carlone G, Kim E, Choi J, Paulos S, Park SH, et al. Development and validation of a robust multiplex serological assay to quantify antibodies specific to pertussis antigens. Biologicals. 2019; 57: 9-20.

13.Taleuzzaman M. Limit of Blank (LOB), Limit of Detection (LOD), and Limit of Quantification (LOQ) Organic and Medicinal Chemistry. Organic & Medicinal Chem IJ. 2018; 7.

14.Sankar R, Ravisankar P, Naga Navya C, Pravallika D, Navya D. A Review on Step- by-Step Analytical Method Validation Controlling hypertension: A brief review View project A Review on Step-by-Step Analytical Method Validation View project A Review on Step-by-Step Analytical Method Validation. IOSR J Pharmacy. 2015; 5.

de Almeida RMS, Ogawa GM, Siqueira Custódio PR, Socorro Sampaio MA, da Silva NS, et al. Performance of Multiplex Detection Method of IgM Class Antibodies against Toxoplasma gondii, Rubella and Human Cytomegalovirus. Ann Pediatr Child Health 2022; 10(7): 1295.

Received : 24 Nov 2022
Accepted : 19 Dec 2022
Published : 23 Dec 2022
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
TEST Journal of Dentistry
ISSN : 1234-5678
Launched : 2014
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X