Borescope Vs Laryngoscope in Difficult Airway Management by Non-Expert Personnel: A Non-Randomized Pilot Study in a Simulated Environment

Research Article | Open Access | Volume 7 | Issue 1

  • 1. Anesthesiology Resident, CES University, Colombia
  • 2. Medical Student, University of Caldas, Colombia
  • 3. Medical Doctor, University of Caldas, Colombia
  • 4. General Surgery Resident, Military University Nueva Granada, Colombia
  • 5. Anesthesiologist, University of Caldas, Colombia
+ Show More - Show Less
Corresponding Authors
Valentina Franco Uribe, Anesthesiology Program, CES University, Cl. 10a #22 - 04, Medellin, Antioquia, Colombia; Tel: 573206321211
Abstract

Introduction: Though direct laryngoscopy is the method of choice to secure the airway, it has multiple disadvantages. There are different devices to improve the visualization of the airway, but their high cost and need for advanced training limit their use. Borescope would be a cost-effective option with advantages similar to those types of devices.

Objective: The primary objective was to compare successful intubation at first attempt with and without the use of borescope. The secondary objective was to compare the average total intubation time with and without the use of borescope.

Materials and methods: Controlled non-randomized pilot study in a simulated environment. Population frame work of 48 medical students who received training on the basic management of the device. A difficult airway was recreated in a simulation laboratory using the Laerdal Airway Management Trainer model and intubation parameters were compared using a conventional laryngoscope under direct vision and through borescope.

Results: In the group with borescope, 22 out of the 24 participants performed the intubation at the first attempt (91.6%) while in the group without borescope, 15 participants performed the intubation at the first attempt (62.5%) (P: 0.02). No statistically significant difference was found in any other evaluated variable.

Conclusion: Borescope could be an effective and economic alternative compared to other devices, since it has similar or better success rates, even for non-expert personnel.

Citation

Uribe VF, Vélez Ríos DA, Jiménez LC, Pulecio JD, Prado Reina AF, et al. (2019) Borescope Vs Laryngoscope in Difficult Airway Management by Non-Expert Personnel: A Non-Randomized Pilot Study in a Simulated Environment. Int J Clin Anesthesiol 7(1): 1099.

Keywords

•    Airway management
•    Intratracheal intubation
•    Equipment and supplies

INTRODUCTION

Among the main competences that a general practitioner should have, those related to adequately securing the airway stand out. Failure in achieving adequate airway patency in a patient with difficult airway (DA) may compromise his ventilation and oxygenation. The incidence of DA in surgical population depends on the degree of difficulty; it may range from 1 to 18% in patients who require several intubation attempts [1]. Its incidence in general population varies between 1.15 and 3.8% and failed intubation between 0.13 and 0.3%, which may result in complications as serious as bronchoaspiration, upper airway lesions, cerebral hypoxia, and death [2].

Direct laryngoscopy is the method of choice to ensure the airway in both elective and urgent cases. However, multiple disadvantages are observed along this procedure, e.g., difficulty in obtaining optimal glottic visualization or greater cervical spine movement. In the market there are several devices useful to minimize risks, which promise to improve the results and reduce complications. Video laryngoscopes are one of the most used because they obtain higher success rates, but their high cost and the need for a greater learning curve for some of them are the main difficulties that limit their use [3].

In this study, a device known as borescope, an optical fiber that allows the visualization of difficult access paths was used.

At one end, it has a camera with LED light and, at the other end, an USB connector through which the image is transmitted in real time to a cellular device [4], by means of an Android or IOS application.

MATERIALS AND METHODS

This pilot test was conducted from September 2017 to March 2018. A call was sent to the medical students of the University of Caldas. To be included the students must have taken and passed the subject ‘anesthesiology and resuscitation’ in the year immediately prior to the study, to be enrolled in the Medicine program, as established in the official curriculum of the University of Caldas. The students were divided into two groups, and they were assigned an odd or even number according to the order of arrival. All participants received instructions on the incorporation of the borescope into the laryngoscope, its basic function and the procedure to be performed. Data collection was carried out in the University’s clinical simulation laboratory using the Laerdal Airway Management Trainer (Laerdal, Stavanger, Norway) for intubation, where a DA was recreated by means of a cervical collar ensuring a Cormack-Lehane III. The USB / borescope device and the Android V3.1 or higher operating system was used, it allowed the USB host on-the-go (OTG) function, through which an interface was established between the device’s camera and the mobile phone to get the image in real time. Cell phones were not specific, however, they all met the criteria requested by the manufacturer of the device, especially, a resolution greater than 640x480 (p/p). In order to properly use the borescope, the “CameraFi” application (Vault Micro Inc., Seoul, Korea) was downloaded for free to each device from the Google Play store (Figure 1).

The borescope camera was fixed 35 mm from the tip on the base of the Macintosh sheet number 3 with clean adhesive tape (strapper) and its stability was verified; this spot was determined considering that the purpose was to achieve the best visualization with as much anatomical information as possible without missing details in the quality of the image. Each participant performed intubation with orotracheal tube 7.5, with balloon and guide. Once the participant was ready, the timer was set to zero and immediately started when the student took the laryngoscope; the stopwatch was stopped once intubation was declared or after a failure. The intubation of the manikin was verified by insufflation with self-inflating bag (Figure 2 and 3).

The primary objective was to compare successful intubation at first attempt with and without the use of borescope. The secondary objective was to compare the average total intubation time until with and without the use of borescope. Evaluated variables: failed intubation, i.e., the impossibility to achieve successful intubation after 120 seconds or after 3 unsuccessful attempts, failed attempt, i.e., when the participant removed the tube from the manikin’s mouth or when esophageal intubation was performed. The number of attempts required by each participant, the total intubation time and whether or not there was esophageal intubation were also evaluated. In addition, 2 subjective variables were recorded: procedure performance difficulty and visualization of vocal cords. The information was collected manually by the research participants and later analyzed with the statistical program Stata version 12. Contingency tables were made, chi test2 and Fisher’s exact tests when there were cells with five cases or less and Student’s t-test for continuous variables with normal distribution.

RESULTS

The inclusion criteria stated in this study were met by 48 students; all the participants were included in the final analysis. There were 2 failed intubation cases in the group without borescope (8.3%) and one case in the group with borescope (4.2%). However, this difference is not statistically significant (P: 0.5). In the group with borescope, 22 out of the 24 participants performed the intubation at the first attempt (91.6%), compared with 15, out of 24 (62.5%), in the group without borescope (P: 0.02). The variable number of attempts is also affected by the use of borescope since most participants who used it required one attempt to achieve successful intubation, while a greater proportion of participants who did not use borescope required more attempts (Table 1).

No statistically significant differences were found in the average total intubation time (40.09 ± 5.85 in the group without borescope and 45.42 ± 6.39 in the group with borescope; Student t-test P: 0.5). Although a trend towards greater number of esophageal intubation was found (2 cases in the group without borescope (8.3%) and none in the group with borescope), it was not statistically significant (P: 0.2). Similarly, no differences were found in the number of students who reported difficulty during the procedure (29.2% without borescope and 37.5% with borescope, P: 0.3). However, a higher proportion of students who used borescope reported having visualized the glottis (87.5%), compared with the group without borescope (62.5%), P: 0.04.

DISCUSSION

This pilot test confirmed that borescope could be an effective alternative to improve the success rate of orotracheal intubation at the first attempt, even in non-expert personnel. A 2016 Cochrane systematic review included 64 studies with 7044 adult participants and compared the use of video laryngoscope (VL) of one or more designs with a Macintosh laryngoscope [5]. The authors concluded that VL may reduce the number of failed intubations, particularly among patients who have a difficult airway, improve glottic vision and decrease laryngeal trauma. There is no difference in patients with airway defined as normal or without predictors of difficult airway; this due to the great heterogeneity in the results dealing with time, attempts and failures in intubation. This is why the borescope model recreating a DA was used; the main finding was that, with the device, the rate of successful intubations in the first attempt (91.6%) increases significantly with respect to the exclusive use of laryngoscope (62.5%).

The first attempt to manage a borescope is not easy for students; it requires a greater learning curve. However, due to the number of successful intubations at the first attempt and the greater visualization of the glottis, it could be easily handled by non-expert personnel. Borescope use may be a part of training and teaching in medical students, by clinical simulation in different scenarios to improve psychomotor skills and abilities in anesthesia practice [6], mainly in airway management.

There is a growing implementation of technology in the practice of anesthesiology and related areas. Sheraton et al. [7], described the use of smart phones for development in the world of anesthesia and the success they will have in clinical practice. The different applications in smart phones are available for free, like the application used in this study, which guarantees easy access for borescope use, thus, success is expected. Low D and Groos [8] used an IOS device to assist nasal fiber optic intubation in a simulation model and improved intubation times in all attempts. These findings are important since they give rise to different management alternatives in different areas of anesthesiology and resuscitation in a world that technologically advances very fast; they also open the way to other works of research about borescope, and not exclusively in airway management.

We found very few studies about the use of this device in humans. Ahmad Sabry et al. [9], used borescope for nasal intubation and concluded that it would be an economical alternative to a fibrolaryngoscope since they used it with 8 patients without any complications. Karippacheril et al. [10], reported their initial experience with the device in 24 patients older than 16 years taken to elective surgery; these researchers intubated 22 patients at the first attempt, and the other 2 required 2 attempts, none of them had associated complications. The authors concluded that this is a personal and economic device and that it is safe and reliable for clinical use.

Additionally, Chaparro et al. [2], conducted a non-systematic review to show the current state of videolaryngoscopes and their impact on airway management. The authors argue that the visualization of the larynx through devices such as VL does not guarantee the success of intubation, that the effectiveness of VL in AD has limited evidence and that the success of intubation with VL is related to experience in its management, with a learning curve that does not exceed 10 patients. Finally, the authors conclude that there are still controversies in the management of AD and the safety it represents for patients. Considering these aspects, it is important to conduct future research regarding the limitations that borescope could have both in airway management and in its safety, which will require different studies despite the fact that it has been shown to be safe in humans.

It would be necessary to conduct controlled clinical trials in simulation scenarios and later with actual patients in order to compare the use of borescope with other devices, such as videolaryngoscopy and standard laryngoscope.

Table 1: Number of intubation attempts according to group.

Group Attempts Total
1 2 3
n % n % n %
With borescope 22 91.6 1 4.2 1 4.2 24
Without borescope 15 62.5 7 29.2 2 8.3 24
P: 0.02 (Fisher's exact test)

 

CONCLUSION

Although borescope is a new device and it has not been proven as a clinical tool in airway management, it could work as an effective and cheaper alternative to other devices with similar or even better successful rates.

ACKNOWLEDGEMENTS

Dr. Fernando Arango, for his collaboration in the design and analysis of this study

Financing

The funds used for this pilot study were provided by the researchers. No financial assistance was received from any other institution and / or person outside of it. The University of Caldas provided the clinical simulation laboratory where the model Laerdal Airway Management Trainer (Laerdal, Stavanger, Norway) was used for intubation, and other equipment. There were 4 borescopes which were acquired with the researchers’ own resources for an approximate price of 10 US dollars each.

Limitations

Very few studies that could serve as a reference have used borescope as a clinical tool in patients as well as in simulation scenarios. This pilot test is limited because the students were not assigned randomly to the use of the borescope and also due to small sample size given the difficulty of finding students that met the inclusion criteria.

REFERENCES

1. Benumof JL. Management of the difficult adult airway. With special emphasis on awake tracheal intubation. Anesthesiology. 1991; 75: 1087-1110.

2. Chaparro-Mendoza A, Luna-Montúfar CA, Gómez JM. Videolaryngoscopes: the solution for difficult airway management or just another strategy? Systematic review. Rev Colomb Anestesiol. 2015; 43: 225-233.

3. Karippacheril JG, Le Cong M. Videolaryngoscopy using an Android smartphone: a direct digital technique. Indian J Anaesth. 2016; 60: 143-145.

4. Trivedi JN. An economical model for mastering the art of intubation with different video laryngoscopes. Indian J Anaesth. 2014; 58: 394- 6396.

5. Lewis S, Nicholson A, Cook T, Smith A. Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation for general anesthesia. Cochrane Database Syst Rev. 2016; 11: CD011136.

6. Gómez LM, Calderón M, Sáenz X, Reyes G, Moreno MI, et al. Impact and benefit of clinical simulation in the development of psychomotor skills in anesthesia: a double blind randomized clinical trial. Rev Col Anestesiol. 2008; 36: 93-107.

7. Sheraton TE, Wilkes AR, Hall JE. Mobile phones and the developing world. Anaesthesia. 2012; 67: 945-950.

8. Low D, Goos M. Using an Apple iPad to assist fibreoptic intubation. Anaesthesia. 2012; 68: 110-111.

9. Ahmad Sabry MHI, Aboughazy A. The use of Usb endoscope (borescope) to guide nasal intubation - a pilot study. Abstract PR528. Anesth Analg. 2016; 123: 237.

10. Karippacheril JG, Umesh G, Ramkumar VJ. Inexpensive videolaryngoscopy guided intubation using a personal computer: initial experience of a novel technique. Clin MonitComput. 2014; 28: 261.

Received : 07 Feb 2019
Accepted : 28 Feb 2019
Published : 06 Mar 2019
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X