Abstract

Purposes: to statistically find out the achievable and practical pre-treatment verification tolerances for HDR (high dose rate) cylinder treatment using scout AP (anterior and posterior) /LAT (lateral) images and the corresponding organs at risk dosimetry effects.

Materials and Methods: Five previously treated vaginal cuff cancer patients were randomly selected. All of the patients had prescription of 5 mm above the cylinder surface with 500 cGy per fraction for 5 fractions. The treatment plans were developed using the Oncentra treatment planning system with graphical local optimization method. Two scouts (AP/LAT) were taken during CT (computed tomography) scan day and on the treatment day of each fraction. The dummy source was placed in the cylinder and the tip of the dummy source to the top of the pubic bone vertical distance on the AP scout was recorded. The cylinder diameter was also measured. The LAT dummy source angle relative to the horizontal line was also recorded. The treatment for each fraction was then reconstructed using AP/LAT scout images. The source position was registered back to the planning CT. A dummy cumulative plan was then created with a prescription of 2500 cGy for 1 fraction.

Results: The AP shifts were ranging from -5 mm to 10 mm and the LAT shifts were ranging from -13° to 9°. The percentage median dose difference can be ranging up 0.3%-14% for the rectum and -17.8%-8.0% for the bladder. Based on the statistics results, we propose the pre-treatment simulation tolerances: trying to maintain the cylinder in AP direction shift under 5 mm, a warning zone is 5 mm-10 mm. For the lateral angle, the tolerance is trying to maintain the angle difference under 5 degree with 5-10 degree as warning zone.

Conclusion: We have done pre-treatment simulation quality assurance study using AP/LAT scout images. The tolerance for this pre-treatment simulation was proposed. The relative dosimetry effects for the rectum and bladder were reported.

PACS number: 87.55. Gh, 87.55. km, 87.55. Qr, 87.55. tm

Keywords

•    HDR
•    Cylinder
•    Vaginal
•    Tolerance
•    Brachytherapy

Citation

Shi C, Fang D, Iannuzzi C (2013) Pre-treatment 2D Scout Verification Tolerances and Dosimetry Effects on OAR for the Cylinder High Dose Rate Brachytherapy Treatment of Vaginal Cuff Cancer. JSM Clin Oncol Res 1(1): 1001.

INTRODUCTION

The cylinder High Dose Rate (HDR) brachytherapy alone has been considered an effective way to treat early stage vaginal cuff cancer [1]. Although the Organs At Risk (OAR) dose is much lower than the tolerance for complication, [2] however, it is also important to do certain quality assurance to make sure the correct applicator is used and the treatment position is reproducible based on the American Association of Physicists in Medicine (AAPM) task group report 56 and 59 [3,4]. A recent report [5] has found out that the most often happened errors during HDR treatment are the wrong application treatment length and applicator size. “Additionally, not enough patients are being imaged prior to the administration of their brachytherapy” [5]. In current brachytherapy era, it is easy and available to do certain pre-treatment simulation for HDR treatment to prevent such errors happen. However, extensive re-planning CT imaging may not be necessary per recent study [6]. Previous studies have been done about the planning optimization, [7-9] treatment time verification, [10,11] the cylinder insert angle effect on the target [12] and OAR, [13]and the air pocket effects [14]. Few study has been done on the pre-simulation verification tolerances and the corresponding effects on OAR. In this study, our goals: 1. to statistically find out the achievable and practical simulation tolerances for HDR cylinder treatment using scout AP/LAT images; 2. to find out the corresponding dosimetry effects.

Methods and Methods

Five previously treated vaginal cuff cancer patients were randomly selected. All of the patients had a prescription of 5 mm above the cylinder surface with 500 cGy per fraction for 5 fractions. The patients’ general information is listed in table 1.

Table 1: General patient information.

Patient ID

Age

Staging

Diameter (mm)

Active length (mm)

P1

67

T1 NX MX

35

40

P2

37

T1a N0 M0

30

40

P3

69

T1a N0 M0

35

45

P4

92

T2 NX MX

35

40

P5

64

T1a N0 M0

30

40

The patients were scanned using GE lightspeed CT (computed tomography) simulator with a slice thickness 2.5 mm. The cylinder, rectum, bladder were contoured by a physician. The isodose 100% was normalized to the catheter points which were created with active catheter dwell positions with a distance of half of the cylinder diameter size plus 5 mm. The isodose 100% distribution was further optimization using Oncentra treatment planning system (Nucletron, an Elekta company, Atlanta, GA, version 4.1) graphical local optimization method. The 100% isodose line was dragged to conform to the catheter points and follow the cylinder tip curvature. Two scouts (anterior-posterior-AP and lateral-LAT direction) were taken during CT scan day. The dummy source was placed in the cylinder and the tip of the dummy source to the top of the pubic bone vertical distance on the AP scout was measured as baseline. The cylinder diameter was also measured for 2nd check of the applicator. The LAT dummy source angle relative to the horizontal line was also measured as baseline. A typical AP and LAT scout images were shown as Figure 1.

Figure 1 A) 3D view of the catheter reconstruction for treatment in AP  direction. B) 3D view of the catheter reconstruction for treatment in LAT  direction.

Figure 1:  A) 3D view of the catheter reconstruction for treatment in AP direction. B) 3D view of the catheter reconstruction for treatment in LAT direction.

For each fraction treatment, two similar scout images were taken to verify the insert angle, distance, and application diameter. The patient was then transferred using stretcher to the treatment room once the simulation was done and the physician approved for treatment.

In order to reconstruct the treatment for each fraction, the scout images including the baseline scouts were imported into Oncentra treatment planning system and the source were reconstructed. The reconstruction origin was the right femur bone center. The dwell position time and relative weight were the same as the plan. The reconstructed source coordinates were recorded. The planning source coordinates and the baseline source coordinates were then co-registered using three-dimensional rigid registration method as shown in equation (1).

X' = RX + T                (1)

Where X is the reconstructed baseline source coordinates and X’ is the planning CT coordinates. The rotation matrix R and transformation matrix T were then determined by using Singular Value Decomposition (SVD) method. Once the rotation and transformation were known, other fractions source coordinates were then able to be transferred into the planning CT coordinates. A full set treatment sources reconstruction in CT coordinates were illustrated in Figure 2.   

Figure 2 A) 3D view of the catheter reconstruction for treatment in PA  direction. B) 3D view of the catheter reconstruction for treatment in Lateral  direction. C) 3D view of the catheter reconstruction for treatment in Inferiorsuperior direction.

Figure 2: A) 3D view of the catheter reconstruction for treatment in PA direction. B) 3D view of the catheter reconstruction for treatment in Lateral direction. C) 3D view of the catheter reconstruction for treatment in Inferiorsuperior direction.

A dummy plan was then created with a prescription of 2500 cGy for 1 fraction. Each of the five reconstructed catheters was given the same dwell time as the plan. The cumulative dose was then computed. Since the reconstructed catheter 100% isodose may invade into either the bladder or rectum, only the median dose for the bladder and rectum were considered for relatively dosimetry effect.

RESULTS

Figure 3

Figure 3 A) AP scout shift from the baseline for all five fractions and patients.  B) LAT scout angle shift from the baseline for all five fractions and patients.

Figure 3 : A) AP scout shift from the baseline for all five fractions and patients. B) LAT scout angle shift from the baseline for all five fractions and patients.

shows the AP and LAT scout image shifts for all five patients. The AP shifts were ranging from -5 mm. Table 2

Table 2: The median rectum and bladder dose difference between the plan and the dummy reconstructed deliver plan.

 

Rectum –

Plan %

Rectum -Deliver %

Difference %

Bladder-

Plan %

Bladder-Deliver %

Difference %

             

P1

55.8

62.7

6.9

51.0

54.9

3.9

P2

66.4

75.4

9.0

47.9

47.1

-0.8

P3

66.3

66.6

0.3

62.8

65.9

3.1

P4

52.1

55.4

3.3

59.9

67.9

8.0

P5

48.0

62.0

14.0

35.7

17.9

-17.8

shows the rectum and bladder median percentage dose difference between the plan and the dummy reconstructed delivered plan. The percentage difference can be ranging up 0.3%-14% for the rectum and -17.8%-8.0% for the bladder depending on the cylinder insert reproducibility

Based on the statistics results from Figure 3 and table 2 and considering the convenience for the patient and treatment, we propose the pre-treatment simulation tolerances as: trying to maintain the cylinder in AP direction shift under 5 mm, and a warning zone is 5 mm-10 mm. Anything more than 10 mm shift may be adjusted or with both physician and physicist agreement. For the lateral angle, trying to maintain the angle difference under 5 degree and keep the 5-10 degree as warning zone. Anything more than 10 degree shifts may be adjusted for quality assurance purpose. With this tolerance setting, the bladder and rectum median dose can be controlled within tens of percentage, which may not cause serious complication for vaginal cuff HDR alone treatment. If the HDR treatment were followed with external beam treatment, a tighter tolerance may be necessary to avoid the warning zone. Due to the study limitation, the rectum and bladder maximum dose and DVHs were not available. However, the median doses were less affected by the source position errors and could also represent the dosimetry effects for the OAR.

CONCLUSION

In this study, we have done pre-treatment simulation quality assurance study using AP/LAT scout images. The tolerance for this pre-treatment simulation was proposed. The relative dosimetry effects for the rectum and bladder were reported.

REFERENCES

1. Solhjem MC, Petersen IA, Haddock MG. Vaginal brachytherapy alone is sufficient adjuvant treatment of surgical stage I endometrial cancer. Int J Radiat Oncol Biol Phys. 2005; 62: 1379-1384.

2. Barney BM, MacDonald OK, Lee CM, Rankin J, Gaffney DK. An analysis of simulation for adjuvant intracavitary high-dose-rate brachytherapy in early-stage endometrial cancer. Brachytherapy. 2007; 6: 201-206.

3. Nath R, Anderson LL, Meli JA, Olch AJ, Stitt JA, Williamson JF. Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine. Med Phys. 1997; 24: 1557-1598.

4. Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR, Williamson JF. High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys. 1998; 25: 375-403.

5. Richardson S. A 2-year review of recent Nuclear Regulatory Commission events: what errors occur in the modern brachytherapy era?. Pract Radiat Oncol.. 2012; 2:157-163.

6. Malik S, Banu PA, Rukhsana N, Ahmed M, Yasmin Z. A proposed protocol on HDR cylinder treatments: proof of avoidance of re-planning of CT based fractionated treatment, using a critical, statistical and graphical analysis of clinical data. J Contemp Brachytherapy. 2012; 4:146-51.

7. Li Z, Liu C, Palta JR. Optimized dose distribution of a high dose rate vaginal cylinder. Int J Radiat Oncol Biol Phys. 1998; 41: 239-244.

8. Li S, Aref I, Walker E, Movsas B. Effects of prescription depth, cylinder size, treatment length, tip space, and curved end on doses in highdose-rate vaginal brachytherapy. Int J Radiat Oncol Biol Phys. 2007; 67: 1268-1277.

9. Sivakumar SS, Solomon JGR, Supe SS, Vadhiraja BM, Rao KK, Vidyasagar MS. Optimization in high dose rate vaginal cylinder for vaginal cuff irradiation. Reports of Practical Oncology & Radiotherapy. 2008; 13: 35-48.

10. Mayo CS, Ulin K. A method for checking high dose rate treatment times for vaginal applicators. J Appl Clin Med Phys. 2001; 2: 184-190.

11. Buzdar SA, Gadhi MA, Rao MA, Laghari NA, Anees M. Verification of computerized treatment planning for HDR 192Ir brachytherapy for gynaecological cancer. J Pak Med Assoc. 2009; 59: 113-115.

12. Hoskin PJ, Bownes P, Summers A. The influence of applicator angle on dosimetry in vaginal vault brachytherapy. Br J Radiol. 2002; 75: 234-237.

13. Patel S, Mehta KJ, Kuo HG, Yaparpalvi R, Kalnicki S. Organ at Risk Interfraction Dose Variance in Relationship to Insertion Angle of a Vaginal Cylinder for HDR Brachytherapy. Int. J. Radiation Oncology Biol. Phys. 2012; 84: S452-S453.

14. Richardson S, Palaniswaamy G, Grigsby PW. Dosimetric effects of air pockets around high-dose rate brachytherapy vaginal cylinders. Int J Radiat Oncol Biol Phys. 2010; 78: 276-279.

Shi C, Fang D, Iannuzzi C (2013) Pre-treatment 2D Scout Verification Tolerances and Dosimetry Effects on OAR for the Cylinder High Dose Rate Brachytherapy Treatment of Vaginal Cuff Cancer. JSM Clin Oncol Res 1(1): 1001.

Received : 23 Oct 2013
Accepted : 16 Nov 2013
Published : 19 Nov 2013
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X