Loading

JSM Dentistry

An Ex vivo Behavior of Siroendo Pocket Eletronic Apex Locator in Primary and Permanent Molars

Research Article | Open Access

  • 1. Department of Pediatric Clinic, Preventive and Community Dentistry, Brazil
+ Show More - Show Less
Corresponding Authors
Raquel Assed Bezerra da Silva, Departamento de Clínica Infantil, Odontologia Preventiva e , Faculdade de Odontologia de Ribeirão Preto, Universidade de São Paulo, Avenida do Café s/n, Monte Alegre, 14040-904, Ribeirão Preto, São Paulo
Abstract

Aim: The aim of the present study was to evaluate ex vivo the accuracy of SIROEndo Pocket electronic apex locator (EAL) for working length determination in permanent molars and primary incisors with and without physiological root resorption.

Methods: One calibrated examiner determined the working length in 21 primary incisors either with or without physiological apical resorption and also in 14 permanent molars. Working length was measured both visually, with the placement of a #15 K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically, using the EAL according to manufacturers’ instructions.

Statistics: Data were analysed statistically using the intraclass correlation test.

Results: Comparison of the real and the electronic measurements revealed high correlation between the methods, regardless tooth type and presence or absence of physiological root resorption. For primary teeth without root resorption, the means were 15.75 mm and 15.66 mm for direct (real) and electronic (SIROEndo Pocket) working length measurements, respectively. Similarly, for primary teeth with root resorption, the means were 12 mm and 11.61 mm for direct and electronic working length measurements, respectively. Permanent teeth evaluation showed means values of 20.75 mm and 21.3 mm for direct and electronic working length measurements, respectively.

Conclusion: SIROEndo Pocket was proven comparable the direct mode for apical foramen location for working length measurement in permanent and primary teeth.

INTRODUCTION

During endodontic treatment, correct establishment of the root canal length is critical to complete removal of infected tissues, thorough canal cleaning, shaping, disinfection, and hermetic sealing of the root canal [1]. Indeed, the correct working length determination is important to avoid over-filling or underfilling, increasing the chances of endodontic treatment failure. It is especially critical in primary teeth, whilst overinstrumentation and overfilling can damage the germ of the permanent tooth. Underfilling, on the other hand, is also a risk factor that accounts for persistence of apical infection [2].

Traditionally, tactile sense and conventional periapical radiographic methods have been used to determine the apical foramen. However, tactile sense is empirical and the radiography can only provide a two-dimensional image of a three-dimensional object [3,4]. Also, radiographic assessment has limitations due to anatomic variations of the canal system, interference of adjacent anatomic structures or technical errors in projection [5,6]. In addition, in most cases, the apical foramen does not coincide with the radiographic apex, which may lead to incorrect working length determination.

Additionally, these methods for the determination of root canal length may yield inaccurate information, particularly in primary teeth, in cases with root resorption [7]. Particularly in pediatric dentistry, children’s acceptance and cooperation for radiographic examination is usually achieved with difficulty, and so a method that could minimize the need for exposing paediatric patients to radiation during this part of root canal therapy would be welcomed. All these factors together have stimulated the development of electronic root canal length measuring devices, the so-called electronic apex locators (EALs), which accurately report the foramen or, more precisely, locate the position of the apical constriction.

Most studies have focused on the investigation into these new technologies for determining the root canal length in permanent teeth [8-13]. The high precision rates achieved in these studies have increased their popularity and motivated their clinical use. However, in primary dentition, a few studies had evaluated these devices for determination of canal working length [7,14-17] evaluated ex vivo the accuracy of the multifrequency electronic apex locator (EAL) Joypex 5 in primary molars, and they concluded that there was no significant difference in between the direct observation and the EAL measurement (p < 0.05). The ICC confirmed the agreement of different methods to measure canal length [17].

It is possible to find in the literature data that supports the potencial use of EALs in primary teeth [Ahmed, 2013], [18] regardless the stage of root resorption however, there are no studies evaluating the SIROEndo Pocket EAL for electronic measurement of working length.

Therefore, the purpose of this study was to evaluate ex vivo the accuracy of SIROEndo Pocket EAL for working length determination in permanent molars and primary incisors with and without physiological root resorption.

MATERIAL AND METHODS

Tooth selection and preparation

Maxillary and mandibular teeth (primary and permanent) were obtained from the Human Tooth Bank of the Ribeirão Preto School of Dentistry, University of São Paulo (Brazil). Primary teeth either without physiological apical resorption or with resorption up to half of the root, identified by visual evaluation were used. Preliminary radiographs were taken to evaluate root canal anatomy, identify the radiographic apex and exclude teeth with calcification, whose main canal was not visible radiographically. The roots were numbered and stored in sterile saline until use. Primary incisors extraction was done as a result of prolonged retention (no spontaneous exfoliation), orthodontic purposes or no possibility of restoration after caries excavation. Twenty one primary incisors were selected. A total of 14 permanent teeth extracted for periodontal, orthodontic, or prosthetic reasons were also selected for this study (total of 33 root canals).

After endodontic access cavity preparation, a #15 K-file was passively introduced up to the apical foramen to verify canal patency. No root canal preparation was performed.

Direct (real) determination of working length

For direct (real) measurement of working length, a reference point was first marked at the most coronal portion of the tooth crown using a fine paint marker. Then a K-file with a silicone stop was passively introduced into the root canal until its tip was visible at either the apical foramen or the apical resorption level and then withdrawn 1 mm. The K-file was held by a needleholder perpendicular to the edge defined as the reference point and was laid against an endodontic rule in order to measure the working length. The measure was approximate to the nearest half millimetre (Figure 1).

Electronic determination of working length

The electronic working length determination was undertaken using SIROEndo Pocket apex locator (Sirona Dental GmbH, Salzburg, Austria). Primary and permanent teeth were individually fixed in a sponge soaked in saline and the root canals were also filled with saline. Cotton pellets were used to remove excess saline from the pulp chamber. A low-impedance lip-clip electrode was attached to the sponge and the EAL was used according to the manufacturers’ instructions. The file holder was clipped to the metal shaft of the K-file, and the device was adjusted to react when the file advanced into the root canal reached a point 1 mm short of either the root apex or the apical resorption bevel, that is, the point where the display read 1. The choice for using the “1” reading on the apex locator’s display was based on the results of a previous pilot-study (data non showed), in which this reading presented the best correlation with the actual root canal length measurement (direct method) at 1 mm short of the root apex (Figure 2).

Statistical analysis

The measurements (in mm) of direct and electronic determination of the working length were made by an experienced calibrated examiner (kappa = 0.9), and were recorded in specific charts for further comparison of the methods. Data were analysed statistically using the intraclass correlation coefficient (ICC).

RESULTS

Table 1 presents the working length measurements obtained with direct visualization and SIROEndo Pocket EAL in all primary teeth. Additionally, (Table 2) presents results related to all root canals from permanent molars evaluated in the study. In relation to primary teeth without root resorption, the means were 15.75 mm and 15.66 mm for direct (real) and electronic (SIROEndo Pocket) working length measurements, respectively. Similarly, primary teeth with root resorption, the means were 12 mm and 11.61 mm for direct (real) and electronic (SIROEndo Pocket) working length measurements, respectively. Permanent teeth evaluation showed means values of 20.75 mm and 21.3 mm for direct (real) and electronic (SIROEndo Pocket) working length measurements, respectively. Figures 3 and 4 compares graphically the real and electronic working length measurements related to primary and permanent teeth, respectivelly. The figures show a high intraclass correlation between the direct and electronic methods, regardless of the type of tooth (primary or permanent teeth), and the presence or absence of physiological root resorption related to primary dentition.

DISCUSSION

Radiographic method, traditionally the most popular and trusted way for length measurement in the endodontic field, has advantages like direct observation of the anatomy of the root canal system, number and curvature of roots, presence or absence of periapical pathologies and, in addition, acts as an initial guide for working length estimation [19]. There are, however, a number of disadvantages that make this technique not quite suitable in every situation such as, for example, the danger of overestimation of the root canal length even when it seems to be short of the radiographic apex because of normal anatomic variations in the apical region [20]. Other shortcomings of radiographs include technique sensitivity and subjectivity [21,22] the danger of ionizing radiation [23], and errors of superimposition caused by producing a two-dimensional representation from a threedimensional object[5].

The development and production of electronic devices for locating the canal terminus have been major innovations in root canal treatment. The use of electronic working length determination has increased considerably in recent years, not only in the permanent [8-13,24] but also in the primary dentition [14,25-28,2,15,29,30]. The accuracy of the electronic measurements obtained in the present study with SIROEndo Pocket was similar to that reported by other authors using different EALs in primary teeth without physiological root resorption and permanent teeth [9,2,13,15]. In the present study, the results of SIROEndo Pocket showed high correlation with those of the direct method.

Regarding primary teeth with physiological root resorption, in our study, after using SIROEndo Pocket EAL, ICC value slightly decreased. This result was similar to that reported by NelsonFilho and collaborators which evaluated Digital Signal Processing EAL in primary teeth with root resorption (ICC=0,82) [16].

The primary teeth have some particularities, such as the presence of the phisiological resorption bevel, extended accessories canals, and exposed dentinal tubules due to physiological root resorption, among others. All this factors could interfere on the electrical resistance of the root, promoting changes on the electronic measures. However, many authors agree that the use of EALs in primary teeth is very safe, reliable and recommended [18].

Wankhade and collaborators (2013) also evaluated the root canal length of deciduous teeth with or without resorption. The authors compared the canal length through several methods (EAL, radiographs, tactile sensation and radiovisiography) in relation to the gold standard (actual root canal length). They have found that the mean determined by EAL, both in deciduous teeth without or with root resorption was closest to that of the actual root canal length.

Another study evaluating the use of EALs in primary teeth has found that Root ZX II seems to be a reliable device for obtaining the root canal length in primary maxillary incisor teeth and can be used as an alternative to radiographic technique [29].

Regarding the use of EALs for measurement of working length of permanent teeth, some authors have found that the electronic method was more accurate than the radiographic method, in vivo [24] and in vitro [31].

It is important to mention that the promising results of ex vivo studies do not imply that radiographs can be replaced by the use of the EALs. The use of EALs is a valuable tool for complementing and/or assisting radiographic methods of working length determination, and may reduce the number of radiographs required during endodontic therapy [3]. EALs used alone without the radiographic method cannot give any information about the curvature and direction of the root canal. Radiographic assessment after electronic root canal measurement confirms the root canal pathway because the image of the file within the canal facilitates the observation of a number of anatomical details. In addition, ex vivo studies do not include the errors that may occur whilst measuring working length in the mouth under clinical conditions. Thereby, further studies with this multi-frequency root canal length measuring device under different conditions should be done to confirm its reliability for clinical indication.

CONCLUSION

In conclusion, in this ex vivo study, the SIROEndo Pocket EAL was comparable the direct mode to determine the working length in permanent and primary teeth with and without physiological root resorption.

REFERENCES

1. de Camargo EJ, Zapata RO, Medeiros PL, Bramante CM, Bernardineli N, Garcia RB, de Moraes IG. Influence of preflaring on the accuracy of length determination with four electronic apex locators. See comment in PubMed Commons below J Endod. 2009; 35: 1300-1302.

2. Leonardo MR, da Silva LA, Nelson-Filho P, da Silva RA, Lucisano MP. Ex vivo accuracy of an apex locator using digital signal processing in primary teeth. See comment in PubMed Commons below Pediatr Dent. 2009; 31: 320-322.

3. Gordon MP, Chandler NP. Electronic apex locators. See comment in PubMed Commons below Int Endod J. 2004; 37: 425-437.

4. Kim E, Lee SJ. Electronic apex locator. See comment in PubMed Commons below Dent Clin North Am. 2004; 48: 35-54.

5. ElAyouti A, Weiger R, Löst C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. See comment in PubMed Commons below J Endod. 2002; 28: 116-119.

6. Hoer D, Attin T. The accuracy of electronic working length determination. See comment in PubMed Commons below Int Endod J. 2004; 37: 125-131.

7. Subramaniam P, Konde S, Mandanna DK. An in vitro comparison of root canal measurement in primary teeth. See comment in PubMed Commons below J Indian Soc Pedod Prev Dent. 2005; 23: 124-125.

8. Haffner C, Folwaczny M, Galler K, Hickel R. Accuracy of electronic apex locators in comparison to actual length--an in vivo study. See comment in PubMed Commons below J Dent. 2005; 33: 619-625.

9. Venturi M, Breschi L. A comparison between two electronic apex locators: an in vivo investigation. See comment in PubMed Commons below Int Endod J. 2005; 38: 36-45.

10. Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. See comment in PubMed Commons below Int Endod J. 2006; 39: 408-414.

11. Topuz O, Uzun O, Tinaz AC, Sadik B. Accuracy of the apex locating function of TCM Endo V in simulated conditions: a comparison study. See comment in PubMed Commons below Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007; 103: e73-76.

12. Ding J, Gutmann JL, Fan B, Lu Y, Chen H. Investigation of apex locators and related morphological factors. J Endod. 2010; 36: 1399-1403.

13. Stoll R, Urban-Klein B, Roggendorf MJ, Jablonski-Momeni A, Strauch K, Frankenberger R. Effectiveness of four electronic apex locators to determine distance from the apical foramen. See comment in PubMed Commons below Int Endod J. 2010; 43: 808-817.

14. Leonardo MR, Silva LA, Nelson-Filho P, Silva RA, Raffaini MS. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. See comment in PubMed Commons below Int Endod J. 2008; 41: 317-321.

15. Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. See comment in PubMed Commons below Int Endod J. 2010; 43: 142-147.

16. Nelson-Filho P, Romualdo PC, Bonifácio KC, Leonardo MR, Silva RA, Silva LA. Accuracy of the iPex multi-frequency electronic apex locator in primary molars: an ex vivo study. See comment in PubMed Commons below Int Endod J. 2011; 44: 303-306.

17. Silva EJ, Herrera DR, Souza-Júnior EJ, Rosa TP. Evaluation of the multifrequency electronic apex locator Joypex 5 in primary teeth. See comment in PubMed Commons below Eur Arch Paediatr Dent. 2014; 15: 51-54.

18. Ahmed HM. Anatomical challenges, electronic working length determination and current developments in root canal preparation of primary molar teeth. See comment in PubMed Commons below Int Endod J. 2013; 46: 1011-1022.

19. Ravanshad S, Adl A, Anvar J. Effect of working length measurement by electronic apex locator or radiography on the adequacy of final working length: a randomized clinical trial. See comment in PubMed Commons below J Endod. 2010; 36: 1753-1756.

20. ElAyouti A, Weiger R, Löst C. Frequency of overinstrumentation with an acceptable radiographic working length. See comment in PubMed Commons below J Endod. 2001; 27: 49-52.

21. Radel RT, Goodell GG, McClanahan SB, Cohen ME. In vitro radiographic determination of distances from working length files to root ends comparing Kodak RVG 6000, Schick CDR, and Kodak insight film. See comment in PubMed Commons below J Endod. 2006; 32: 566-568. 

22. Kazzi D, Horner K, Qualtrough AC, Martinez-Beneyto Y, Rushton VE. A comparative study of three periapical radiographic techniques for endodontic working length estimation. See comment in PubMed Commons below Int Endod J. 2007; 40: 526-531.

23. Pendlebury ME, Horner K, Eaton KA. Selection Criteria for Dental Radiography. 1st ed. London: Faculty of General Dental Practitioners, Royal College of Surgeons of England, 2004.

24. Singh SV, Nikhil V, Singh AV, Yadav S. An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators. See comment in PubMed Commons below Indian J Dent Res. 2012; 23: 359-362.

25. Bodur H, OdabaÅŸ M, TulunoÄŸlu O, Tinaz AC. Accuracy of two different apex locators in primary teeth with and without root resorption. See comment in PubMed Commons below Clin Oral Investig. 2008; 12: 137-141.

26. Ghaemmaghami S, Eberle J, Duperon D. Evaluation of the Root ZX apex locator in primary teeth. See comment in PubMed Commons below Pediatr Dent. 2008; 30: 496-498.

27. Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. See comment in PubMed Commons below Int Endod J. 2008; 41: 436-441.

28. Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. See comment in PubMed Commons below Int Endod J. 2009; 42: 115-121.

29. Saritha S, Uloopi KS, Vinay C, Chandra Sekhar R, Rao VV. Clinical evaluation of Root ZX II electronic apex locator in primary teeth. See comment in PubMed Commons below Eur Arch Paediatr Dent. 2012; 13: 32-35.

30. Wankhade AD, Kumar R, Singh RK, Chandra A. Root canal length determination by different methods in primary teeth: an in vivo study. See comment in PubMed Commons below Pediatr Dent. 2013; 35: E38-42.

31. Diwanji A, Rathore A2, Arora R3, Dhar V4, Madhusudan A5, Doshi J6. Working Length Determination of Root Canal of Young Permanent Tooth: An In vitro Study. See comment in PubMed Commons below Ann Med Health Sci Res. 2014; 4: 554-558

Received : 05 Aug 2014
Accepted : 16 Oct 2014
Published : 18 Oct 2014
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
Author Information X