JSM Dentistry

Investigation of Methods for Measuring Mandibular Complete Denture Retention

Research Article | Open Access

  • 1. Department of Geriatric Dentistry, Showa University, Japan
+ Show More - Show Less
Corresponding Authors
Momoe Nakatsu, Department of Geriatric Dentistry, Showa University, Kitasenzoku 2-1-1, Ota-ku, Tokyo, Japan, Tel: 81-3-3787-1151(#236); Fax: 81-3-3787-3971
ABSTRACT

Purpose: We aimed to clarify the effect of different measurement sites on the retention of maxillary complete dentures and to determine the optimal site and loading method for evaluating retention. The factors associated with mandibular complete denture retention are unclear; thus, a retention evaluation method has not been established. We investigated measurement methods and the optimal site by using a model to establish a retention measurement method for mandibular complete dentures.

Method: We manufactured complete resin dentures, based on normal methods used for edentulous mandibular models. We fabricated a covering splint to measure the denture surface. We set four points (A–D) for the measurement sites. We measured retention after filling the space between the denture mucosal surface and the model with four types of intervening saliva.

Results: All four types of intervening saliva could be measured only when measuring site A traction and the downward pressure on the fenestrations. Both conditions had a strong positive correlation (r = 0.94, P<.01). At site A, there was no significant difference in the measurement values with pressure exerted at45°and at 23°, which indicated a positive correlation (r = 0.73, P<.01).

Conclusion: Mandibular complete denture retention can be measured by applying oblique downward pressure on the occlusal plane at the mandibular central incisor midpoint.

CITATION

Ishihara M, Sato Y, Kitagawa N, Nakatsu M, Takeda K, et al. (2017) Investigation of Methods for Measuring Mandibular Complete Denture Retention. JSM Dent 5(1): 1080.

KEYWORDS

•    Mandibular complete denture
•    Retention
•    Incisal edge

INTRODUCTION

With the onset of a super aging society, the number of elderly people in Japan will increase and the number of years denture wearers live will increase. Therefore, the number of intractable cases which has difficulty in denture treatment is predicted to increase, including severe mandible bone resorption and agerelated changes in mandibular position [1-3] and the onset of xerostomia [1,4-6]. Several factors reduce denture retention. High-quality complete denture treatment is needed because complete denture treatment significantly affects the quality of life of elderly people [1,2,7-10].

In the past, much research has focused on the factors that affect denture retention. Östlund [11,12] reported that denture retention conditions differ depending on the intervening saliva layer between the residual ridge and the mucosal surface of the denture base. Furthermore, several reports exist concerning the effect of saliva characteristics and oral moisturizers on denture retention [13-16].In late years the new method of fabricating milled CAD-CAM dentures is established, and a study on retention is accomplished [17].

Therefore, we conducted research focusing on maxillary complete denture retention. We developed a new retention measuring device for objectively assess ingretention [18], and we previously reported chair-side evaluation methods for the optimal sites and loading methods for measuring denture retention [19]. When we investigated the effect of the viscosity of oral moisturizers and the shape of the residual ridge on the retention of maxillary complete dentures, we found that denture retention increased as the viscosity of the oral moisturizer increased, and retention was correlated with the relative position of the anterior alveolar crest rather than the form of the molar residual ridge [20-22].

As indicated previously, information related to maxillary complete denture retention is just beginning to be clarified, but few reports exist on mandibular complete dentures. Furthermore, no gold standard method has been established for measuring the retention of mandibular complete dentures. We therefore investigated measurement methods and the optimal site for measurements by using a model to establish a method for measuring mandibular complete denture retention.

MATERIAL AND METHODS

We assessed normal wearers of complete dentures and manufactured a complete resin denture using an ideal arrangement and floor model, based on normal methods used in edentulous mandibular models for measurements. We fabricated a covering splint to measure the denture surface by hooks, filled the space between the mucosal surface of the denture and the model with intervening saliva, and recorded the measurements. The details of the measurement methods are described later.

The retention measuring device

A digital force gauge (Digital Force Gauge RX Series; Aikoh Engineering, Tokyo, Japan) was used as the retention measuring device. Traction and pressure were measured by two methods using the digital force gauge. The pressurizing device was attached by fabricating a hook-shaped attachment (Figure 1).

The mandibular complete denture covering splint

A mandibular edentulous jaw model (G10FE-402K; Nissin Dental Products, Inc., Kyoto, Japan) was used to ensure consistent measurement conditions. The experiment was conducted using resin mandibular complete dentures manufactured based on this model. We measured the retention of mandibular complete dentures fitted in patients. Thus, difficulty in directly attaching a hook for measuring retention was expected. We therefore fabricated a splint that covered the mandibular complete denture. The splint was a 3-mm thermoplastic resin for the denture base (Erkodur; Erkodent, Yokohama, Japan), which was manufactured with a heat-pressuring molding machine (Dual Former; Daiei Dental Product Co., Ltd., Osaka, Japan) and welded to a plaster model of the mandibular complete dentures. The splint configuration was set with the front teeth 2 mm from the cervical line and the molars 5 mm from the marginal side of the cervical line.

Measurement sites

The following four points were set as the measurement sites:

Site A: Mandibular central incisor midpoint

This site was selected because it lifts when retention is measured in clinical practice.

Site B: Central fossa of the first molar on the right side of the mandible

This site reproduces the force added when chewing a highly viscous food.

Site C: Central fossa of the first molar on the left side of the mandible

This site was selected for the same reason as site B.

Site D: Extension of midpoint of first molar

The measurements were obtained at a similar location described in a previous study [23] (Figure 2).

Hook 1 was attached to site A (Figure 3) and hook 2 was attached to site B and site C (Figure 4) with a room temperature curing resin (Unifast III; GC, Tokyo, Japan). Sites A, B, and C were measured with direct traction on the hooks. For site D, dental floss was attached to the hook at sites B and C and traction was applied at the midpoint. Hook 1 and hook 2 were curved 1.2-mm lines (Sun-Cobalt Wire; Dentsply Sankin, Tochigi, Japan).

saliva between the mucosal surface of the mandibular complete dentures and the model. The intervening saliva used included the following: (1) artificial saliva (Saliveht; Teijin Pharma, Tokyo, Japan), (2) oral moisturizer spray (Wet Care; Kissei Pharmaceutical, Nagano, Japan), (3) liquid oral moisturizer (Oral Balance Liquid; T&K, Tokyo, Japan), and (4) geloral moisturizer (Biotene Oral Balance Gel; T&K, Tokyo, Japan). The viscosity was measured using a digital rotational viscometer (Brookfield Rotational Viscometer). The viscosity of two 500-mL samples from each patient was measured at 20°C. The average viscosity measurement was used.

Measurement methods

Each measurement was taken at the four measurement sites described previously. Retention was set as the amount of load when the denture separated from the residual ridge.

Retention measurements were obtained, as follows. First, each measurement was obtained five times with no intervening saliva. Second, each measurement was obtained six times with artificial saliva and with each of the three oral moisturizers. The very first measurement was excluded because the intervening saliva had not become attached to the denture and model. The amount of intervening saliva was set at the volume that covered the entire surface of the model mucosal surface of the denture. After bonding, measurements were obtained by applying approximately 20 N of hand pressure for 10 seconds to the denture.

A measurement was stopped when the splint separated from the denture or when the measurement value exceeded 20 N. These conditions were chosen because past research on maxillary complete dentures demonstrated that an intraoral pressure exceeding 20 N has a high risk of eliciting pain, thereby making it difficult to make measurements. If the aforementioned conditions for stopping measurement occurred twice, then subsequent measurements were stopped. To investigate the effect of angles on measurement conditions, the pressure measurements were obtained 10 times with artificial saliva intervention at site A only at 45° and 23° to the occlusal plane.

Statistical analysis

To test normality, Shapiro–Wilk tests were used for all measurement values.The Friedman test was used for the mean retention value under the five conditions (i.e., without intervening saliva and with the four different moisturizers). The Bonferroni’s method was then used for multiple comparisons. Spearman’s rank correlation coefficient was used to assess the relationship between each site. The t-test was used for the angle relationship. Statistical software SPSS ver.19 (SPAW Statistics Base 19; IBM, Tokyo, Japan) was used for statistical processing.

RESULTS

Retention under the five conditions and intervening saliva viscosity

The results of the retention measurements are shown in Figure (6).The splint detached from the site with highly viscous intervening saliva such as liquids and gels, which made measurement impossible. If a measurement was impossible, it was marked with an x on the graph.

When the incisal edge of the central incisor was exposed to a 45° downward pressure, only two measurements were possible with each of the intervening saliva type when the incisal edge of the central incisor was tractioned upwards. When we compared the highly viscous geloral moisturizer with no intervening saliva, artificial saliva, and spray oral moisturizer, we found that retention tended to increase as the viscosity of the intervening saliva increased (Figure 7).

The relationship between traction and pressure

At site A, where measurements were possible for all intervening saliva types, there was a strong positive correlation between retention with traction on the incisal edge of the central incisor and retention with pressure at the incisal edge of the central incisor (r = 0.94, P<.01) (Figure 8).

Angle relationship

There was no significant difference at site A when pressure was exerted at 45° and when it was exerted at 23° (P<.01) (Figure 9). There was a correlation between both angles (r = 0.73 P<.01).

DISCUSSION

The mandibular complete denture covering splint

We fabricated a splint with a hook attached to determine the measurement site extra orally. However, when actually measuring inside a patient’s mouth, a splint would not be fabricated. We instead wanted to measure the dentures currently used by the patient. In this study, we therefore created a fenestration and used a configuration that would enable direct pressure on site A. With this method, measurements were possible with a 45° direct downward pressure on the front teeth in the mouth.

Intervening saliva

The volume of intervening saliva was determined by referencing previous reports. Kawazoe et al. [14], reported that retention was weakened by excess intervening saliva or insufficient intervening saliva between the base and the mucosal surface under the base [14]. They also observed that each person had an optimal volume of intervening saliva. Yamagaki et al., reported that obtaining stable measurements of retention was possible if the test sample covered all surfaces of the model [24]. Therefore, we did not specify a set amount of intervening saliva; the amount was instead set as the amount that covered all mucosal surfaces of the denture base.

We selected four types of intervening saliva: artificial saliva and three types of oral moisturizers with different levels of viscosity. The reasons for this selection are stipulated later. The combination of inorganic electrolyte components in artificial saliva is nearly the same as in normal saliva. In several studies, Östlund et al., first reported that differences in the properties, composition, and outflow of saliva have a significant effect on retention [7,21]. Artificial saliva, which has a viscosity similar to that of human saliva, was used as the intervening saliva to eliminate differences in the salivary conditions of individual patients and to compare the effect on retention.

Yamagaki et al., demonstrated that retention increases as the viscosity of an oral moisturize increases [24]. We therefore selected three types of oral moisturizers with different levels of viscosity to ascertain the effect of moisturizer viscosity on mandibular complete denture retention.

Retention under the five conditions

Yamagaki et al., reported a positive correlation between retention and viscosity: when an oral moisturizer was used, retention increased with increasing viscosity of the oral moisturizer [24]. The current study used moisturizers with low viscosity, moderate viscosity, and high viscosity, based on this previous report. We also found that the regression line showed a positive correlation between retention and viscosity. On the basis of this information, we presumed that it would be sufficiently possible to use a developed retention measuring device for chairside measurements. The reason measurements were possible at site A only where the splint did not detach may be because direct pressure was used and because the coverage on the front teeth was narrower than that in the molar region, which reduced retention.

Correlation between traction and pressure

Based on statements in the denture examination and test methods advocated by the Japan Prosthodontic Society (Tokyo, Japan), an evaluator should attempt to remove artificial teeth with the fingers when evaluating whether denture retention is appropriate. This approach is consistent with traction at site A. However, pressure may be exerted in the direction of site A when upward traction is applied to the molar region. This is similar to the direction of force that pulls upward in the molar region and causes the dentures to detach when viscous food is chewed in the mouth. We therefore preferred exerting pressure to site A because measurements can be obtained as is and without using a splint when obtaining actual retention measurements in the mouth.

Angle relationship

No significant difference was found when a 45° and 23° downward pressure was applied to site A. Because we found a correlation between both angles, we presumed that measurements obtained with either angle would produce virtually the same values. This finding therefore suggested that a detailed specification of an angle is unnecessary when measuring the actual retention force.

CONCLUSION

This study suggests that mandibular complete dentures retention measurement may be possible with oblique downward pressure on the occlusal plane at the mandibular central incisor midpoint.

FUTURE PROSPECTS

In the future, we want to obtain direct measurements of retention in the mouths of patients actually fitted with mandibular complete dentures, and to clarify the efficacy of these measurements. On conducting this measurement, it will be necessary to specify a patient’s mouth opening capacity and to eliminate the effect of the buccal mucosa, lip pressure, and so forth. If it were clarified that the measurements in the patient’s mouth are efficacious, it would be possible to investigate the relationship between retention and the shape of the residual ridge and the relationship between retention and the position of the anterior alveolar ridge. This will make it possible to determine intractable cases and to determine the therapeutic effect, based on pre- and post-treatment measurements. It will then be possible to provide better complete denture treatment and to contribute to improving the quality of life of elderly edentulous patients.

ACKNOWLEDGEMENTS

The authors would like to thank the staff of the Department of Geriatric Dentistry of Showa University (Tokyo, Japan).This work was supported by a grant from the Japan Society for the Promotion of Science (Tokyo, Japan; KAKENHI grant number JP26462938).

REFERENCES

1. Polzer I, Schimmel M, Müller F, Biffar R. Edentulism as part of the general health problems of elderly adults. Int Dent J. 2010; 60: 143- 155.

2. Jacobson TE, Krol AJ. A contemporary review of the factors involved in complete denture retention, stability, and support. Part I: retention. J Prosthet Dent. 1983; 49: 5-15.

3. de Oliveira Junior NM, Rodriguez LS, Mendoza Marin DO, Paleari AG, Pero AC, Compagnoni MA. Masticatory performance of complete denture wearers after using two adhesives: a crossover randomized clinical trial. J Prosthet Dent. 2014; 5: 1182-1187.

4. Närhi TO. Prevalence of subjective feelings of dry mouth in the elderly. J Dent Res. 1994; 73: 20-25.

5. Navazesh M, Brightman VJ, Pogoda JM. Relationship of medical status, medications, and salivary flow rates in adults of different ages. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1996; 81: 172-176.

6. Ikebe K, Morii K, Kashiwagi J, Nokubi T, Ettinger RL. Impact of dry mouth on oral symptoms and function in removable denture wearers in Japan. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005; 99: 704-710.

7. Yoshida M, Matsuo K, Wadamoto M, Sato Y, Akagawa Y, Turu H. The influence of complete denture therapy on quality of life of edentulous patients, Hiroshima Daigaku Shigaku Zasshi. 1993; 25: 257~26.

8. Sato Y, Hamada S, Akagawa Y, Tsuga K. A method for quantifying overall satisfaction of complete denture patients. J Oral Rehabil. 2000; 27: 952-957.

9. Fukai K, Takiguchi T, Ando Y, Aoyama H, Miyakawa Y, Ito G, et al. Mortality rates of community-residing adults with and without dentures. Geriatr Gerontol Int. 2008; 8: 152-159.

10. Regis RR, Cunha TR, Della Vecchia MP, Ribeiro AB, Silva-Lovato CH, de Souza RF. A randomized trial of a simplified method for complete denture fabrication: patient perception and quality. J Oral Rehabil. 2013; 40: 535-545.

11. Östlund SG. Some physical principles in the retention of dentures. Northwest Univ Bull. 1948; 49: 11-20.

12. Östlund SG. Saliva and denture retention. J Prosthet Dent.1960; 10: 658-663.

13. Sekine H, Tajima T, Yamagata H. Studies on the retention of dentures (1st Report). J Prosthet Dent. 1963; 8: 67.

14. KawazoeY, Hamada T, Yamada S. Studies on the retention of dentures (Part 4). Changes of intervening salivary volume during tapping movement. J Prosthet Dent. 1976; 20: 148-152.

15. Komatsu S, Bandai M, Takizawa H, Ishikawa T, Maruoka M, Tamura F, et al. [Studies on retention of denture base 3. Wettability of denture base]. Showa Shigakkai Zasshi. 1991; 11: 89-99.

16. Bandai M, Sekiguchi Y, Takayanagi Y, Toyoshima Y, Komatsu S, Migou S, et al. [Studies on retention of denture base. 1. Viscosity of mediating fluid and palatine shape]. Showa Shigakkai Zasshi. 1989; 9: 288-296.

17. AlHelal A, AlRumaih HS, Kattadiyil MT, Baba NZ, Goodacre CJ. Comparison of retention between maxillary milled and conventional denture bases: A clinical study. J Prosthet Dent. 2017; 117: 233-238.

18. Aoyagi K, Sato Y, Kitagawa N, Okane M, Kakuda T, Takayama M. Development of a simple chair-side evaluation method for complete denture retention forces and its reproducibility. Jpn J Gerodont. 2014; 29: 21-28.

19. Kakuda T, Sato Y, Kitagawa N, Nakatsu M, Aoyagi K, Takayama M, et al. Examination of optimal sites and loading methods for measuring maxillary complete denture retention. Jpn J Gerodont. 2015; 30: 25- 36.

20. Ishihara H, Kitagawa N, Sato Y, Hara S, Hosono Y, Ishibashi S. A study on the utility of a scale for examination in residual ridge height based on an objective assessment. Nihon Hotetsu Shika Gakkai Zasshi. 2007; 51: 751-759.

21. Ishibashi S, Sato Y, Kitagawa N, Hara S, Hosono Y, Ishihara H. A study on the clinical utility of a scale for examination in residual ridge. Ann Jpn Prosthodont Soc. 2009; 1: 157-165.

22. Takayama M, Sato Y, Kitagawa N, Nakatsu M, Aoyagi K, Kakuda T, et al. The effects of viscosity of oral moisturizers and residual ridge form on the retention force of maxillary complete dentures. JSM Dent. 2016; 4: 1077.

23. Horita H, Shiraishi K, Yamabe Y, Goto H, Yoshimatsu T, Fujii H. Clinical evaluations on removability of removable partial dentures. Nihon Hotetsu Shika Gakkai Zasshi. 1989; 33: 517-521.

24. Yamagaki K, Kitagawa N, Sato Y, Okane M, Mashimo J. The relation between the physical properties of oral moisturizer and denture retention force. Jpn J Gerodont. 2011; 26: 402-411.

Ishihara M, Sato Y, Kitagawa N, Nakatsu M, Takeda K, et al. (2017) Investigation of Methods for Measuring Mandibular Complete Denture Retention. JSM Dent 5(1): 1080.

Received : 24 Jan 2017
Accepted : 13 Mar 2017
Published : 15 Mar 2017
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
Author Information X