Effects of Three Dry Powder Inhalers on Deposition of Aerosolized Medicine in the Human Oral-PharyngealLaryngeal Region

Research Article | Open Access

  • 1. Department of Industrial Systems and Technology, Jackson State University, USA
+ Show More - Show Less
Corresponding Authors
Mohammed Ali, Biosimulation and Aerosol Research Lab, Department of Industrial Systems and Technology, Jackson State University, JSU Box 18480, Jackson, Mississippi 39217, USA Tel: 1601-979-0327; Fax: 1601-979-4110
Abstract

The dry powder inhaler (DPI) is a popular, effective and convenient drug delivery device for inhalation therapy to treat asthma. However, a large quantity (approximately 54%) of inhaled aerosols deposit in the oropharyngeal region. Deposition in this region is undesirable because it provides minimum therapeutic benefits and has adverse localized or systemic side effects. This study reports a method of examining electrostatic charge effects on deposition of three DPI aerosols (Spiriva™ Handihaler, Advair Diskus™, and Pulmicort™ Turbohaler) in a cadaver-based cast of the human oral-pharyngeal-laryngeal (OPL) regions. Experimental aerosols were generated from the three commercially available DPIs by means of inhalation as boluses, and then characterized with an electronic single particle aerodynamic relaxation time analyzer with or without passing through the OPL regions. The results showed that aerosol particles were not only of different sizes but also carried different positive, negative and zero electrostatic charges. The deposition fraction of total particles (charged and uncharged) in the OPL regions for the Spiriva, Advair and Pulmicort were 22%, 61% and 7%, respectively, whereas the deposition fraction of charged particles in the Spiriva, Advair and Pulmicort generated aerosols were 62%, 67% and 28%, respectively. The inherent net charge to mass ratio were Spiriva 0.76 ± 0.11 µC/g (negative), Advair Diskus 0.49 ± 0.3 µC/g (negative), and Pulmicort 0.46 ± 0.02 (µC/g (negative), respectively. The study results also revealed that inherent charges of smaller (aerodynamic diameter, da < 2.0 ∝m) particles influenced their agglomeration, and therefore, increased their deposition due to inertial impaction and electrostatic space charge forces. In addition, the deposition fraction of these charged particles rapidly increased for Spiriva and Advair but marginally increased for Pulmicort with increasing particle sizes. Electromechanical properties (both aerodynamic size and electrostatic charge) play significant roles in the deposition of dry powder inhaler aerosols in the human oral-pharyngeal-laryngeal regions.

Keywords

Dry powder inhaler , Static charge , Asthma , Aerodynamic size , Oropharyngeal deposition

ABBREVIATIONS

DPI: Dry Powder Inhaler; OPL: Oral-Pharyngeal-Laryngeal; ET: Extra Thoracic; COPD: Chronic Obstructive Pulmonary Disease; ELPI: Electrical Low Pressure Impactor; ESPARTA: Electronic Single Particle Aerodynamic Relaxation Time Analyzer; USP: United States Pharmacopoeia; ETS: Electro-Tech Systems; CMAD: Count Median Aerodynamic Diameter; MMAD: Mass Median Aerodynamic Diameter; GSD: Geometric Standard Deviation; SD: Standard Deviation; DE OPL: Deposition Efficiency In The Oral-Pharyngeal-Laryngeal Region

INTRODUCTION

The oral-pharyngeal-laryngeal (OPL) deposition is the major determinant for lung deposition of an inhaled aerosol [1]. In order to be delivered to the target receptors of the intrathoracic lung, aerosolized pharmaceutical agents must first traverse and penetrate the extra thoracic (ET) airway (i.e., passages of the mouth and throat) [2-4]. That is, to elicit optimum therapeutic responses, medicinal agents such as bronchodilators and steroids employed in the management of asthma should be selectively distributed among lung airways. To accomplish this task, it is imperative that the particle filtering characteristics of the ET region be acknowledged when delivering aerosolized medicines to the lungs [5].

The dry powder inhaler (DPI) is a popular drug delivery device used in the treatment of respiratory diseases, such as asthma and chronic obstructive pulmonary diseases (COPDs), because 1) it has been shown to be effective and convenient to use, 2) it does not contain propellant and can remain with required physiochemical properties in a wide change of environmental conditions. As noted, the localized deposition of drug particles at desired sites, such as inflamed tissue, and appropriate receptors has been recognized by clinical investigators as being critical for the effective administration of asthma drugs. Approximately 54% of inhaled aerosols are lost in the oropharyngeal region [6-8]. These deposited drugs are swallowed, enter the gastrointestinal tract, and cause systemic side effects [7]. Furthermore, the low thoracic delivery efficiencies of costly drugs would be an impediment to their use [2,8,9]. It is therefore appropriate to study the particles’ electromechanical properties on deposition behavior in the oral-pharyngeal-laryngeal (OPL) regions.

Several studies have found that electrostatic charge forces influence particle deposition in the ET region along with more commonly recognized forces like inertial impaction, sedimentation, and diffusion [10-17]. It is an accepted practice that most DPI formulations consist of micronized drug blended with larger carrier particles. The electrostatic interaction between drug and carrier is a major determinant of DPI performance like other attributes such as particle size, flow property, formulation, drug-carrier adhesion, respiratory flow rate, and device geometry [18]. This study employed three different DPIs with different formulations, devices, doses and manufacturers. As a result, their electrostatic charge properties were different due to the combined effects of these factors, and the effect of this charge on deposition variation in the OPL was the aim of this study.

The DPI aerosol particles acquire charges via electron and ion transfer during particle-particle and particle-device component contact and separation. More importantly, respiratory airways are conductive but do not prevent particles from charging [19]. The charge distributions (both magnitude and polarity) are dependent on the work functions of the materials coming in contact with each other, the friction and surface area involved in the contact, dielectric properties of the materials, and the ambient relative humidity. The charge magnitude is affected by numerous factors, including drug propellant surfactants, metal surfaces of delivery devices, drug/carrier homogeneities, and excipient particles size distribution [19,20]. Although, it has been standard laboratory procedure to pass an aerosol through a charge neutralizer to attain a Boltzmann equilibrium (i.e., zero net charge), a DPI aerosol has only a slight possibility to become charge neutralized before inhalation [21]. Electrostatic properties of particles consist of: (a) mutual repulsion between particles due to space charge forces, subsequently influencing agglomeration and interactions with other particles; and (b) mutual attraction between particles and neutral inner surfaces of airways due to image charge forces [19,22]. Both forces are dependent on the airborne particle number density [21]. In addition, the adhesive forces among drug particles are due to the electromagnetic forces that act between electrons and protons of the individual molecules [20]. A study with 0.6 µm carnauba wax particles showed that deposition increased from 13% to 22% when the number of elementary charges on particles increased from 2.7 to 7.5 µC [23]. Balachandran [24] reported that a low particle charge (approximately 10 µC per 2 µm particle) produced enhanced pulmonary deposition, while a higher particle charge (approximately 300 µC per particle) favored deposition in the upper airways.

In the literature it has been reported that particle sizes and device geometries affect ET losses [21,25]. However, charge acquisition due to mouth-throat turbulence and its subsequent role on deposition has not been elucidated. The major reason of this lack of information has been the unavailability of a suitable measuring instrument. Drug aerosol characterization studies have used an electrical low pressure impactor (ELPI) to analyze aerosol particle charges [26, 27]. However, the ELPI has a limitation because it provides only the net charge of all particles deposited on its impactor plate rather than for each particle in real time. To resolve this issue, this work used an electronic single particle aerodynamic relaxation time analyzer (ESPARTA) to perform electric charge measurements on a single particle basis and in real time [26]. This investigation had two specific objectives: (1) to characterize and compare the aerodynamic sizes and electrostatic charges of aerosols generated by three DPIs; and (2) to study the effects of these properties on aerosol deposition in a replica cast of the human oral-pharyngeallaryngeal airways (Figure 1). Aerosol researchers usually employ the United States Pharmacopoeia (USP) metal throat or glass throat impinger to sample particles for characterization in an Andersen Cascade Impactor. Studies have found that the metal USP throat and glass throat impinger have significant deposition differences, and the deposition in both were lower than in vivo data [9,28]. A chief motivation for using the OPL replica cast in this study was its intrinsic similarity to in vivo airway anatomy.

 

MATERIALS AND METHODS

(i) Three delivery devices and drugs were studied. (1) DPI Spiriva™ Handihaler, drug tiotropium bromide, 18 mcg per capsule with lactose monohydrate as carrier (Boehringer Ingelheim Pharmaceuticals, Inc., Ridgefield, Connecticut, USA). (2) DPI Advair Diskus™ 500/50, single blister contained the drug fluticasone propionate 500 mcg and salmeterol 50 mcg inhalation powder (GlaxoSmithKline, Research Triangle Park, North Carolina, USA). (3) DPI Pulmicort™ Turbo haler, each metered dose contained 200 mcg budesonide inhalation powder, but the data on excipient amount was not available. (AstraZeneca, Wilmington, Delaware, USA). Three aerosol devices were used for each type of DPIs. Even though all of the formulations contained both are active pharmaceutical ingredient and an excipient, their unequal quantities did not affect the measurement of aerodynamic diameters and electrostatic charges because the ESPARTA instrument characterized each bolus for a period of 5 minutes only (Figure 1).

(ii) The replica cast (Figure 1) was made of polyester resin. It was a life-sized model and included the continuous OPL passages of a healthy 82-year-old male subject. It was made from a post-mortem negative cast and was prepared in the Division of Pulmonary and Critical Care Medicine at the University of California, Irvine, California, USA. The replica OPL cast was placed in a controlled chamber designed to simulate the humid environment inside a human. The relative humidity was maintained above 95% with an automatic humidity controller (Model 514, ETS Electro-Tech Systems, Inc., Glenside, Pennsylvania, USA). The chamber and lab temperature were same, and it was measured 37o C by using an Extech Heavy Duty Hot Wire Thermo-Anemometer (Extech Instruments, Waltham, Massachusetts, USA).

(iii) The Electronic Single Particle Aerodynamic Relaxation Time Analyzer (ESPARTA) was used to measure aerodynamic sizes and electrostatic charges in real time. It was designed and developed in the Aerosol Drug Delivery Research Lab of the University of Arkansas at Little Rock, Little Rock, Arkansas, USA [26]. Its application in the area of aerosol medicine was demonstrated elsewhere [29].

(iv)An aerosol sampling chamber held aerosols for characterization (Figure 2). It had a volume of 28.3 liters and four ports equipped with valves (V2 , V3 , V4 , and V5 ), and one pressure gage. V2 controlled the aerosol flow from the OPL replica cast to the chamber, V3 connected the chamber to a vacuum pump, V4 opened the chamber to ambient air, and V5 controlled aerosol flow from the chamber to the ESPARTA. V1 controlled aerosol inhalation through the cast. The Aneroid pressure gauge monitored the chamber’s vacuum pressure. The chambers inside walls were lined with a grounded wire mesh. Figure 2 depicts the experimental system. To characterize the DPI aerosols, Cheng proposed that bolus inhalations be employed to promote realism [30]. We adopted this methodology. Experiments started with the evacuation of the aerosol sampling chamber to 35.6 cm of mercury (0.5b) to simulate the inhalation of an aerosol bolus once the inlet valves V1 and V2 were opened. The inhaled volume was 4 liters over 8 seconds at a flow rate of 30 L/min. The flow rate was measured using an Extech Heavy Duty Hot Wire ThermoAnemometer (Extech Instruments, Waltham, Massachusetts,USA). To characterize drug aerosols with or without passing through the OPL replica cast, the DPI devices were actuated for two experimental scenarios.

Scenario 1: At the inlet of the cast the instant before opening V1 , while V2 was open and V3 , V4 , and V5 were closed (i.e., inhaling through the cast)

through the cast) Scenario 2: At the inlet of the sampling chamber the instant before V2 was open while V3 , V4 , and V5 were closed (i.e., bypassing the cast)

Prior to each run in the series of experiments, the aerosol sampling chamber was vacuum cleaner and the cast was washed with distilled water. To ensure that the inner walls of the sampling chamber had the same effects on particle motion, the DPI devices were actuated at the same inlet of the aerosol sampling chamber. This supported an assumption of equal particle losses in the scenarios described above. Once the sampling chamber was filled, V1 was shut to implement Scenario 1, or V2 was shut to implement Scenario 2, and V4 was opened for sampling by the ESPARTA instrument for a period of 5 minutes for both scenarios 1 and 2. It was unlikely; therefore, that a variation of drug quantity in each dose from an individual inhaler affected the measurements and comparisons of charged particles in each bolus. The procedure was repeated for 5 consecutive runs. The sizes and charge distributions were measured in each case. Raw data was acquired through Lab View (National Instruments, Austin, Texas, USA) and mined by Aerosol Particle Data Analyzer software (developed at the Aerosol Drug Delivery Research Lab of the University of Arkansas at Little Rock, Little Rock, Arkansas, USA).

 

RESULTS AND DISCUSSION

In order to be able to compare the electromechanical properties’ effects on the deposition fraction of the generated aerosols from three DPIs, the particle aerodynamic size and charge distributions are plotted in (Figures 3-5). Figure 3 shows the comparison of the particle aerodynamic size distributions of three DPIs. The 5-minute counts without passing through the OPL cast were, Spiriva: 5923 ± 77, Advair: 5889 ± 20, and Pulmicort: 5110 ± 25 (mean ± SD). Aerosols from the Spiriva and Advair were widely distributed. The Pulmicort had a narrower distribution than the other two. Figure 4 shows the comparisons of the count median aerodynamic diameter (CMAD), and the mass median aerodynamic diameter (MMAD) of three DPIs. The CMAD and MMAD of Spiriva were 3.61 ± 0.07 µm, and 4.99 ± 0.03 µm, respectively; Advair were 3.61 ± 0.19 µm, and 5.29 ± 0.15 µm, respectively; and Pulmicort were 2.86 ± 0.02 µm, and 3.65 ± 0.1 µm, respectively. Compared to the other two, Pulmicort showed the best reproducibility result (geometric standard deviation, GSD< 0.02). Figure 5 shows the electrostatic charge distributions of three DPI aerosols. As can be seen, the Pulmicort had less number of charged particles than the other two though the share of negatively charged particles was higher for each DPI. The inherent net charge to mass ratio was also negative for all of them, e.g., Spiriva: -0.76 ± 0.11 µC/g, Advair: -0.49 ± 0.3 µC/g, andPulmicort: -0.01 ± 0.02 µC/g. All three DPI aerosols contained both positively and negatively charged particles. Spiriva and Advair aerosol particles carried a large number of elementary charges. As a result, deposition efficiency were greatly affected (Figure 6) by their charge distributions. The deposition fraction will be defined as the ratio of the number of particles removed from the aerosol (i.e., deposited) while traveling through the replica OPL cast to the number of particles originally entering it. Table 1 shows the normalized data from 5 runs of charged particle counts and both charged and uncharged particle deposition efficiencies for each DPI. It also summarizes the count median aerodynamic diameter (CMAD), mass median aerodynamic diameter (MMAD), and electrostatic net charge to mass ratio (specific charge) for all DPI aerosols before passing through the OPL replica cast.

As can be seen, oral-pharyngeal-laryngeal deposition of all (charged and uncharged combined) particles of three DPIs were, Spiriva: 22%, Advair: 61%, and Pulmicort: 7%. Comparing the deposition patterns of charged and uncharged particles, it was observed that the uncharged particles deposited much less efficiently than the charged particles. Uncharged particle depositions were, Spiriva: 34%, Advair: 51%, and Pulmicort: 8%, whereas, the charged particles depositions were, Spiriva: 68%, Advair: 67%, and Pulmicort: 28%. Some significant differences in the deposition efficiencies between different polarity particles were also observed. More negatively charged

Table 1: Summary of the laboratory data.

Drug

Delivery

Device

1DE OPL

5All particles

Uncharged

Charged

% of Charged

+ counts

- counts

2Net q/m

+ ve q/m

- ve q/m

(µC/g)

3CMAD

(µm)

5All particles

St. Dev.

4MMAD

(µm)

5All particles

St. Dev.

SpirivaTM

22%

34%

68%

62 %

1,783

1,899

- 0.76

+ 5.30

- 6.61

3.61

0.07

4.99

0.03

AdvairTM

61%

51%

67%

67 %

1,626

2,313

- 0.49

+ 4.77

- 6.30

 

3.61

0.19

5.29

0.15

PulmicortTM

7%

8%

28%

28 %

475

618

- 0.46

+ 5.47

- 6.63

2.86

0.02

3.65

0.10

Abbreviations:

1 DE OPL = Deposition efficiency in the Oral-Pharyngeal-Laryngeal Regions

2 Net q/m = Net charge/mass ratio

3 CMAD = count median aerodynamic diameter

4 MMAD = mass median aerodynamic diameter

5 All particles = both uncharged and charged particles

particles were deposited than were their positive counterparts. For example, deposition of positive and negative particles of Spiriva were 45% and 87%, respectively; Advair were 57% and 74%, respectively; and Pulmicort were 0.2% and 54%, respectively. Figure 6 represents the OPL cast deposition fraction as a function of particle aerodynamic size, which shows how the charged properties affected the deposition patterns. Although the deposition fractions were increased linearly with increasing particle size for all DPIs, the rate of increase was highest for the one which contained the largest percentage of charged particles. The generation of aerosols from a DPI depends upon 1) the integration of inhaler design, 2) the drug and carrier powder mixture formation, and 3) patient’s inspiratory effort. In this study variable 1 and variable 2 are differed among the tested DPIs. Therefore the difference in deposition observed in the present study could be due to differences in design of inhaler devices, flow properties, excipients, or active ingredients in addition to electrostatic charge and particle size.

The study results of this work showed the Spiriva had the highest number of charged (67%) particles, of which 41% were positive and 59% were negative. The Adavair had the second highest number of charged (62%) particles, of which 48% were positive and 52% were negative. Hence, there were two electrostatic force situations. First of all, the uni-polar charged particles induced greater space charge forces and mutual repulsions. As a result, particles came closer to cast walls and were captured. Secondly, bi-polar charged particles agglomerated due to Coulombic attraction forces, which allowed smaller particles to become larger and deposit more efficiently within the cast due to inertial impaction. The agglomeration due to electromagnetic forces is consistent with Finlay’s theory [20], and the increased impaction (due to formations of larger sizes) is consistent with the findings of Yu and Diu [31]. In contrast, the Pulmicort had a relatively small amount (28%) of charged particles, of which 43% were positive and 57% were negative. As a result, very little agglomeration took place, and the aerosol losses on cast walls were comparatively low (7%). During the inhalation of an aerosol bolus, a complicated inlet velocity profile and flow conditions in the OPL replica cast may have affected the abilities of particles to gain electrostatic charges through tribo-electric charging processes [16,32]. Electrostatic charge forces had major influences on the deposition properties of all DPI aerosols. These influences were an indication of enhanced charged particle deposition in the cast. Therefore, it can be summarized that overall deposition efficiencies of charged particles were higher than uncharged particles. This observation agrees with Yu’s [22] mathematical model and the experimental studies done by Hashish et al. [13] and Cohen et al. [14]. One study reported that surface modification of carrier/excipient particle can be reduced by using force control agents such as Plurionic F-68, Cremophor RH40, soya lecithin, glyceryl monostearate, and magnesium stearate [33]. Such attempts may reduce the gaining of high charges of the aerosolized particles.

In this study, it is logical to recognize certain limitations. It purposely avoided particles smaller than 0.5 µm because the ESPARTA was not able to detect sizes below this limit; the OPL replica cast was made of polyester resin; and it also assumed a uniform state of temperature and humidity within the replica cast; and no growth or evaporation of particles in transit. Within this framework, this study did identify and quantitate the interrelated effects of electrostatic charges and particle sizes on the deposition of various DPI aerosols. It is believed that the study findings have practical values and clinical implications and suggest that the developers of aerosol drug delivery devices consider electrostatic charge effects while designing future improved products.

CONCLUSION

In conclusion, this study demonstrates the electromechanical properties (both aerodynamic size and electrostatic charge) significant roles in the deposition of aerosols in the human oralpharyngeal-laryngeal regions. The DPI aerosol deposition in the OPL region (often refers as extra thoracic region) is much less for the drug delivery device that generates fewer charged aerosols than for the others, a fact which clinicians may choose to consider in inhalation therapy protocols.

ACKNOWLEDGEMENT

The author would like to thank Emeritus Professor Malay K. Mazumder of the University of Arkansas at Little Rock, Arkansas, USA for enabling to use the ESPARTA instrument for conducting this study.

REFERENCES

1. Borgström L, Olsson B, Thorsson L. Degree of throat deposition can explain the variability in lung deposition of inhaled drugs. J Aerosol Med. 2006; 19: 473-483.

2. Swift DL, Strong JC. Nasal Deposition of Ultrafine 218Po Aerosols in Human Subjects. J. Aerosol Sci. 1996; 27: 1125–1132.

3. Patel P, Mukai D, Wilson AF. Dose-response effects of two sizes of monodisperse isoproterenol in mild asthma. Am Rev Respir Dis. 1990; 141: 357-360.

4. Martonen TB, Smyth HD, Isaacs KK, Burton RT. Issues in drug delivery: concepts and practice. Respir Care. 2005; 50: 1228-1252.

5. Cheng YS. Aerosol Deposition in the Extrathoracic Region. Aerosol Sci Technol. 2003; 37: 659-671.

6. Kim CS, Eldridge MA, Sackner MA. Oropharyngeal deposition and delivery aspects of metered-dose inhaler aerosols. Am Rev Respir Dis. 1987; 135: 157-164.

7. Fink JB. Metered-dose inhalers, dry powder inhalers, and transitions. Respir Care. 2000; 45: 623-635.

8. Newman SP, Clarke AR, Talaee N, Clarke W. Lung deposition of 5 mg Intal from a pressurized metered dose inhaler assessed by radiotracer technique. Int J Pharm. 1991; 74: 203-208.

9. Biddiscombe MF, Melchor R, Mark VHF, Marriott RJ, Taylor AJ, Short MD, et al. The lung deposition of salbutamol, directly labeled with technetium –99m, delivered by pressurized metered dose and dry powder inhalers. Int J Pharm. 1993; 91: 111-121.

10.Ali M, Reddy RN, Mazumder MK. Electrostatic charging effect on workplace aerosol particle deposition in a hollow throat cast. J. Electrostatics. 2008; 66: 401-406.

11.Ali M. Pulmonary drug delivery. In: Handbook of Non-Invasive Drug Delivery Systems, 1st ed. Kulkarni V, editor. The Netherlands: Elsevier; 2009: 209-246.

12.Hinds WC. Aerosol technology: properties, behavior and measurement of airborne particles. 2nd ed. New York: John Wiley & Sons Inc: 1999.

13.Hashish AH, Bailey AG, Williams TJ. Selective deposition of pulsed charged aerosols in the human lung. J Aerosol Med. 1994; 7: 167-171.

14.Cohen BS, Xiong JQ, Fang CP, Li W. Deposition of charged particles on lung airways. Health Phys. 1998; 74: 554-560.

15.Staniforth JN. The importance of electrostatic measurements in aerosol formulation and preformulation. In: Respiratory drug delivery IV. Byron PR, Dalby RN, Farr SJ, editors. Deerfield (IL): Interperm Press; 1994.

16.Murtomma M, Strengell S, Laine E, Bailey AG. Measurement of electrostatic charge of an aerosol using a grid-probe. J Electrostatics. 2003; 58: 197-207.

17.DeHaan WH, Finlay WH. Predicting extrathoracic deposition from dry powder inhalers. J Aerosol Sci. 2004; 35:309-331.

18.Islam N, Gladki E. Dry powder inhalers (DPIs)--a review of device reliability and innovation. Int J Pharm. 2008; 360: 1-11.

19.Bailey AG, Hashish AH, Williams TJ. Drug delivery by inhalation of charged particles. J Electrostatics. 1998; 44: 3-10.

20.Finlay WH. The mechanics of inhaled pharmaceutical aerosols: an introduction. San Diego: Academic Press; 2001Balachandran W. Control of drug aerosol in human airways using electrostatic forces. J Electrostatics. 1997; 40&41: 579-584.

21.Martonen TB, Rosati JA, Isaacs KK. Modeling deposition of inhaled particles. In Aerosols handbook: measurements, dosimetry, and health effects. Ruzer LS, Harley NH, editors. Boca Raton (FL): CRC Press; 2005.

22.Yu CP. Precipitation of unipolarly charged particles in cylindrical and spherical vessels. J Aerosol Sci. 1977; 8: 237-241.

23.Melandri C, Tarroni G, Prodi V, Zaiacomo TD, Formignani M, Lombardi C. Deposition of charged particles in the human airways. J Aerosol Sci. 1983; 14: 657-669.

24.Balachandran W. Control of drug aerosol in human airways using electrostatic forces. J Electrostatics. 1997; 40&41: 579-584.

25.Geller DE, Eigen H, Fiel SB, Clark A, Lamarre AP, Johnson CA, et al. Effect of smaller droplet size of dornase alfa on lung function in mild cystic fibrosis. Dornase Alfa Nebulizer Group. Pediatr Pulmonol. 1998; 25: 83-87.

26.Ali M, Mazumder MK, Martonen TB. Measurements of electrodynamic effects on the deposition of MDI and DPI aerosols in a replica cast of human oral-pharyngeal-laryngeal airways. J Aerosol Med Pulm Drug Deliv. 2009; 22: 35-44.

27.Glover W, Chan HK. Electrostatic charge characterization of pharmaceutical aerosols using electrical low-power impaction (ELPI). J Aerosol Sci. 2003; 35: 755-764.

28.Zhang Y, Finlay WH, Matida EA. Particle deposition measurements and numerical simulation in a highly idealized mouth-throat. J Aerosol Sci. 2004; 35: 789-803.

29.Ali M. Operating performance comparisons between laser doppler velocimetry and time of flight techniques. J Management Engg Integ. 2010; 2: 1-13.

30.Cheng YS. Modeling aerosol drug delivery. In: Optimization of aerosol drug delivery. Gradon L, Marijnissen J, editors. Boston (MA): Kluwer Academic Publishers; 2003b.

31.Yu CP, Diu CK.A comparative study of aerosol deposition in different lung models. Am Ind Hyg Assoc J. 1982; 43: 54-65.

32.Cross JA. Electrostatics: principles, problems and applications. Bristol, England: Institute of Physics Publishing; 1987.

33.Singh DJ, Jain RR, Soni PS, Abdul S, Darshana H, Gaikwad RV, et al. Preparation and Evaluation of Surface Modified Lactose Particles for Improved Performance of Fluticasone Propionate Dry Powder Inhaler. J Aerosol Med Pulm Drug Deliv. 2014; 17.

Ali M (2015) Effects of Three Dry Powder Inhalers on Deposition of Aerosolized Medicine in the Human Oral-Pharyngeal-Laryngeal Regions. J Drug Des Res 2(1): 1009.

Received : 23 Feb 2015
Accepted : 19 Mar 2015
Published : 23 Mar 2015
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X