Angled Field-in-Field as a New Advanced Radiotherapy reatment Planning Form of the Field-in-Field Technique. A Dosimertic and Treatment Planning Study
- 1. The Islamia University of Bahawalpur, Pakistan
- 2. Shaukat Khanum Memorial Cancer Hospital and Research Center, Pakistan
Abstract
Introduction: Field-In-Field (FIF) is a manually based forward Intensity-Modulated Radiation Therapy (IMRT) plan for which the calculated dose is modified in certain dose distribution areas by creating multiple lower-weighted reduction fields based on the primary field. The most common used form of FIF is the non-angled FIF, but if the inner subfield(s) is/are set at a different gantry angles from the major one, we will get a different design of the FIF technique, which is the angled FIF technique.
Aim: The purpose of this study was to study the Angled Field-In-Field technique - as a new advanced form of FIF technique - and its related isodose lines shapes.
Method: Using TPS and on the 2D-Array, a simple form of Non-Angled FIF (NAFIF) plan consists of 7×7cm2 field entire another larger 15×15cm2 field as a main field is designed on TPS. The two fields were set at the zero gantry position. Other FIF plan forms are designed in which the smaller inner field is set to (5°, 10°, 15°, 20° and 25°) gantry angles while the main field is fixed at zero gantry angle. The all FIF plans were verified using the 2-Dimintional Ionization Chamber Array (2D-Array) and its related Verisoft software. The Angled FIF, wedged and open fields were compared for studying their related isodose lines different shapes.
Results: The results showed that the Angled Field-In-Field technique can be used as anew advanced technique for the radiotherapy and it has a new shape of isodose lines which is a stair shape.
Future work: We predict that a Multiple-Angled Field-In-Field technique in which, the inner fields can be set at different angles in the same or reverse direction of the main angled field. This design has unknown isodose shape, but we predict that it can result in a semi-sloped stair shape, so it needs more study.
Citation
Hammoury SI (2015) Angled Field-in-Field as a New Advanced Radiotherapy Treatment Planning Form of the Field-in-Field Technique. A Dosimertic and Treatment Planning Study. J Radiol Radiat Ther 3(1): 1055.
Keywords
• Radiotherapy
• Angled Field-in-Field (AFIF)
• Linear accelerator
• Treatment planning system TPS
• IMRT
ABBREVIATIONS
FIF: Field-In-Field; IMRT: Intensity-Modulated Radiation Therapy; NAFIF: Non-Angled Field-In-Field; 2D-Array: 2-Dimintional Ionization Chamber Array; CRT: Conformal Radiation Therapy; 3-D CRT: Three-Dimensional Conformal Radiation Therapy; CT: Computed Tomography; RTPS: Radiation; TPS: Treatment Planning Systems; DRRs: Digitally Reconstructed Radiographs; DVHs: Dose Volume Histograms; MLCs: Multi-Leaf Collimators; PTV: Planning Target Volume; AFIF: Angled FieldIn-Field technique; Linac: Linear Accelerator; DICOM: Digital Imaging and Communications in Medicine System; SSD: Source to Surface Distance; CMS: Computerized Medical Systems; XiO : Name of Three Dimensions Treatment Planning System; MAFIF: Multiple Angled FIF Technique
INTRODUCTION
Three-Dimensional Conformal Radiation Therapy (3D-CRT)
Three-dimensional (3D) conformal radiation therapy has been demonstrated to improve tumor targeting and to reduce normal tissue volume exposed in several malignancies [1]. The ideas of three-dimensionality, beam shaping, and irradiation of tumours through multiple fields from different beam angles to reduce the dose to normal tissues have always been present in radiotherapy practice. When the appropriate technology to deliver 3-D CRT, such as Computed Tomography (CT) simulators, radiation treatment planning systems (RTPS) capable of performing three dimensional dose calculations, producing digitally reconstructed radiographs (DRRs) and DVHs, and beam shaping devices such as multi-leaf collimators (MLCs) became available, this way of planning and delivering radiotherapy soon gained popularity [2,3].
Field-In-Field technique
Field-in-field planning, another technique used to generate the effect of intensity-modulated fields but based on forward treatment planning (Figure 1). It is used extensively at M. D. Anderson Cancer Center for planning radiation treatments of the breast [4,5].
3D-CRT planning software helps in displaying the 3D dose distribution at different levels in the planned target volume (PTV). Physical or dynamic wedges are commonly applied to obtain homogeneous dose distribution in the PTV. Despite all these planning efforts, there are about 10% increased dose hot spots encountered in final plans. To overcome the effect of formation of hot spots, a manual forward planning method has been used [6,7]. In this method, one or two more beams with multi-leaf collimator (MLC) of different weights are added in addition to the main used beams in the major plan and sometimes; when we use the FIF technique, we can dispense the physical and dynamic wedges.
Non-Angled Field-In-Field technique (NAFIF)
It consists of a main field and a smaller field inside the main one. The smaller field is set at the same gantry angle of the main field. This can give the standard definition of the known and usually used FIF technique.
Angled Field-In-Field technique (AFIF)
It has the same design as the NAFIF technique but the smaller field will be set at a different gantry angle from the main field.
MATERIALS AND METHODS
Phantom setup, CT scanning and preparation within TPS
For the field-related verification process, no special phantom was necessary; the 2-Diminsional Ionization Chamber Array (2D-Array) was impeded between two blocks from solid slab phantom slices. One block of 5cm thickness was put under the 2D-Array for backscattering and the second one of 4.5cm thickness was put above the 2D-Array surface where the chambers of 2D-Array arranged in the device at one plane under the surface by 0.5 cm. Therefore, the depth above the chambers was 5 cm. The 2D-Array device with the blocks was put on the Linac-couch and adjusted where the chamber at the center of the device at isocenter (at 100 cm from the radiation source) [8-10] (see Figure 2). The phantom arrangement was CT scanned then in exactly the same way as it was later used for the verification measurements. To achieve an adequate spatial resolution during the following verification dose calculations, it was essential to scan the phantom with a sufficiently small slice thickness. We have scanned the phantom with a slice thickness of 2 mm. The scanned phantom was imported via a Digital Imaging and Communications in Medicine system (DICOM) to TPS. Directly after import, it was convenient to define a user origin within TPS exactly at the effective measuring point of the central ion chamber of the array.
Treatment Planning Procedures
Open Field procedures (on the TPS):
- The following steps were done respictively;
1) On the 2D-Array phantom, we designed a field with size of 15x15cm2 at zero gantry and Source to Surface Distance (SSD) =94.8 position as it is shown at the next Figure (3).
2) Dose distribution was calculated by the TPS.
3) The total plan was sent to the VeriSoft to be verified.
Field-In-Field procedures
We can use different arrangements of FIF to deal with the patient treatment planning like; (15,7) field, which means that beam a 15x15 cm2 square field will be opened as a major one and another smaller (or minor) field will be set inside that the major one. Other different arrangements like: (20,5) , (20,10), ….., (40,20), .…. . In this study we chose the (15,7) as an example of FIF (Figure 4).
Non-Angled Field-In-Field (NAFIF) procedures (on the TPS):
- The following steps were done respictively;
1) On 2D-Array phantom, we designed a field of size 15x15cm2 at zero gantry position, another smaller field of size 7x7cm2 was designed intire the larger one at the same gantry position as it is shown at the next Figure 5.
2) The dose distribution was calculated by the TPS.
3) The total plan was sent to the VeriSoft to verify the application of this beam on the treatment machine (the Linac).
Angled Field-In-Field (NAFIF) procedures (on the TPS):
- All previous procedures of section 2.2.4 were repeated at (5 ? , 10 ? , 15 ? , 20 ? and 25 ? ) gantry angles as follows in the next Figures 6-10.
Wedged Field procedures (on the TPS):
- The following steps were done respictively;
1) Using the TPS, a 15x15cm2 treatment field is designed on the 2D-Array phantom.
2) A wedge of angle 15 ? is applied to this field as it was shown at the next Figure (11).
3) Then the dose distribution was calculated by the TPS.
4) The total plan was sent to the VeriSoft to be verified.
5) All the previous four steps were repeated but at (30 ? , 45 ? , 60 ? ) wedge angles as follows in the next figures 12-14.
Dosimetric Verification of FIF technique using the two-Dimensional Ionization Array (2D-Array) and The Analysis
Software “VeriSoft”: To be ensured that the FIF fields are valid on the treatment machine during the treatment of the real patient, we made a dosimetric verification of it using the 2D-Array seven29 solid phantom. We performed a treatment planning with FIF for three different cases with three different tumor sites (Breast, Prostate and Brain). The 2D-Array in combination with the VeriSoft analysis software was used as a dosimetric verification tool of clinical FIF fields. For this purpose, the CMS (Xio 4.6.2) was used, which has the ability of performing a FIF plan and SIEMENS ARTISTE Clinac accelerator, which is equipped with dynamic multileaf collimator.
Verification of Angled Field-In-Field planning technique: Using TPS, five angled FIF plans were performed. Each plan contained a large field 15x15cm2 and a smaller field 7x7cm2 where the smaller field was inside the larger one and the samller field was set at a different gantry angle for every plan. The five gantry angles used are; (5 ?, 10 ?, 15 ?, 20 ? and 25 ? ). The dose profiles of plans were sent separately to the VeriSoft to be compared with that measured by the 2D-Array.
RESULTS AND DISCUSSION
Results
Field-in-field is a new technique enables us to generate the same or better effect of intensity-modulated fields. Additionally, FIF is carrying on forward treatment planning.
Verification of Field-In-Field planning technique: The next Figure 15 showed the matching percentage between the plan received from TPS and that measured by 2D-Array. The result was 100% (excellent) matching, where the total dose point were 729, the evaluated dose points were 371, the passed points were 371 and the failed points were 0 points. Although of being the result of matching was 100% but we noticed some semi yellow colored regions within the gamma distribution (in the right lower window of Figure 15) which means that there were some failed points which couldn’t be counted in the matching percentage. This was because of being the used gamma index criteria was; 3 mm Distance-To-Agreement, 3% Dose Difference with reference to maximum dose of measured data set and suppress doses below 5% of maximum dose of measured data set (see Figure 16). This criteria was the standard one according to the VeriSoft system guide but if we used a different criteria, the matching percentage might be changed.
Figure 15: Shows a print screen of the VeriSoft software that shows a comparison result between the (15,7) FIF plan (at gantry angle=0 ? ) received from the TPS and the measured one by the 2D-Array.
Comparison between Angled Field-In-Field and Wedged Field planning techniques: We used the (15,7) FIF plan (where the 7x7cm2 field was at gantry angle=25 ? and the larger filed 15x15cm2 was at gantry angle=0 ? ) versus a wedged plan of a single 15x15cm2 field with a 60 ? virtual wedge. The angled (15,7) FIF plan at Figure 27a already had the same effect of the wedged field regarding to the dose distribution, where it decreased the dose weight to the volume against the gantry angle which achieved the same effect of the thick side of the wedge and increased the dose weight to the volume toward the gantry angle which achieved the same effect of the thin side of the wedge. Also we noticed that the isodose lines of the AFIF plan had a shape differs from the shape of the wedged filed plan isodose lines, where in FIF plan, the decreased dose weight at the volume against gantry angle made the isodose lines also decreased toward the phantom surface at the same volume but the increased dose weight at the volume toward gantry angle made the isodose lines increased far from the phantom surface at the same volume. The resulted shape of Angled FIF isodose lines was like stairs shape. So we called that shape of the AFIF isodose lines (the Stair Shape) (new addition). And the wedged and open fields gave a slope and straight shapes respectively (see Figure 27 and 28).
The next table 1, showed a comparison between the local doses received by the 2D-Array 15 ionization chambers marked in figure 24c for angled field-in-field, wedged and open fields.
Figure 28: showed a line chart for the local doses shows in table 1. For the open field the line chart nearly took a straight shape, but for the wedged field seemed as a sloped shape and for the angled field-in-field nearly took a stair shape. The stair isodose is a completely new shape of isodose lines which did not be mentioned before so it is a new addition.
Radiation Therapy Planning: As we mentioned in the previous chapter, we used three types of cancer tumors, Breast, Brain and Prostate tumors. We used two radiotherapy planning techniques, Field-In-Filed (FIF) and Intensity Modulated Radiation Therapy (IMRT) techniques for each tumor type.
Discussion
This study was designed to study the Angled Field-In-Field as a new advanced radiotherapy treatment planning form of the Field-In-Field technique.
Nearly, all the previous publications that used the FIF technique, used it in its simple design, which is the Non-Angled FIF [3,11-14]. For example but not limited to; [15], quantified the cold spots under geometrical uncertainties in field-in-field techniques for whole breast radiotherapy, and [16], evaluated a simplified “field-in-field” technique (SFF) that was implemented in their department of Radiation Oncology for breast treatment. In both the two examples, FIF was used in the NAFIF form. But when we used it in the AFIF form, we got the result of being it can be used as an alternative plan of the wedged plan and also we got a new isodose lines shape (the Stair shape).
Table 1: The local doses received by the 15 ionization chambers included in the field 15x15 cm2 for angled field-in-field, wedged and open fields.
Points | Angled Field-In-Field | Wedged Field | Open Filed |
1 | 79.3 | 172.6 | 87.1 |
2 | 91 | 178.2 | 100.5 |
3 | 92.3 | 163 | 101.8 |
4 | 96.5 | 148.9 | 102.1 |
5 | 100.5 | 134.6 | 101.3 |
6 | 100.1 | 121.1 | 100.3 |
7 | 99.9 | 109.8 | 99.9 |
8 | 100 | 100 | 100 |
9 | 100 | 90.9 | 99.9 |
10 | 100.5 | 83.1 | 100.4 |
11 | 100.8 | 76.4 | 101.3 |
12 | 95.9 | 70.1 | 102.1 |
13 | 92.3 | 63.7 | 101.8 |
14 | 90.9 | 57.3 | 100.5 |
15 | 77.5 | 45.4 | 87.1 |
FUTURE WORK
1- Prediction of Advanced Field-In-Field (FIF) forms: We predict two types of AFIF that may be applied separately or mixed. They are:
- Multiple non-angled FIF technique: It can be consisted of a main field and multiple smaller fields inside the main one. All fields are at the same gantry angle of the main field. Each one of the smaller fields will have a portion of the main field’s dose weight aiming to achieve a good dose distribution, a good avoidance to the organs at risk, and reduction of both the hot and/or the cold areas within the treatment field. In the next Figure 29, we used, for example, a main field of 20x20 cm2 size and multiple three smaller fields of sizes 15x15 cm2 , 10x10 cm2 and 5x5 cm2 and they are arranged from outside to inside respectively where the filed 15x15 cm2 will be inside the 20x20 cm2 field, the field 10x10 cm2 will be inside the 15x15 cm2 field and so on. There will be a very important note because it can be used as a 3D compensator rather than the resulted stair isodose shape.
- Multiple angled FIF technique (MAFIF): It has the same design as the multiple non-angled FIF technique but all/ some of the smaller fields will be at a different gantry angle from the main field. As it is shown in the next Figure 30, we designed an example of the multiple angled FIF technique which consists of a main 20x20 cm2 field, and three smaller 15x15 cm2 , 10x10 cm2 and 5x5 cm2 fields. The smaller three fields are arranged from outside to inside respectively. Each one of these three fields will be given a dose weight from the main weight. The isodose lines will take the semi-sloped stair shape as it is shown in the figure 18.
- The inner fields can be set in different angles in the same or reverse direction of the main angled field. This design has unknown isodose shape, so it needs more study.