Positional Consistency of Radiation Isocenter of a Single Energy Medical Linear Accelerator after Use of Fourteen Years for Cancer Treatment

Research Article | Open Access

  • 1. Senior Medical Physicist, Department of Radiation Oncology, HCG Panda Cancer Hospital, Cuttack, India
  • 2. Department of Physics, KIIT University, India
+ Show More - Show Less
Corresponding Authors
Suvendu Kumar Sahoo, Senior Medical Physicist, Department of Radiation Oncology, HCG Panda Cancer Hospital, Cuttack, Odisha, India, Tel: 73-8185-0173
Abstract

The position of radiation isocenter is an important part in quality assurance checks of a medical linear accelerator (Linac). All radiotherapy treatment planning’s and their delivery are done by taking radiation isocenter as a reference point. This isocenter can shift when the gantry, collimator and patient couch do not move symmetrically with each other. In our study, the radiation isocenter verification was done by using a special therapeutic film with digital image processing by using special software. Image acquisition was done using scanning of radiation exposed film by 6 MV photon beam with star shot method. This star shaped beam was designed by the rotation of the collimator, gantry and couch individually. The consistency of position of radiation isocenter was checked five times within last one year randomly by rotation of gantry, collimator to the couch of the medical Linac which is being used since fourteen years for cancer treatment. The mean value throughout the year shows that the size (diameter) for rotational gantry is 0.46 mm, collimator is 0.42 mm and 0.48 mm for the couch. As per AAPM report 40, the size of the Linac radiation isocenter diameter used in this study is still in good agreement because the value of the radiation isocenter is less than 2 mm diameter.

Citation

Sahoo SK, Mishra KC, Kumar VS, Sahoo TK, Mishra SK, et al. (2018) Positional Consistency of Radiation Isocenter of a Single Energy Medical Linear Accelerator after Use of Fourteen Years for Cancer Treatment. J Radiol Radiat Ther 6(1): 1079.

Keywords

•    Medical linear accelerator
•    Radiation isocenter
•    Quality assurance
•    Radiotherapy

INTRODUCTION

Modern radiation therapy is one crucial method for cancer treatment and it is commonly used throughout the world now [1]. One of the modalities of modern radiotherapy used for the treatment of cancer is a Linear Accelerator (Linac). This Linac produces electron beam and X-rays (photons) generated by the generator and have the energy that can be varied [2]. Aggressive Quality Assurance (QA) tests play a very important role in ensuring that the treatment given to patients has been correctly conducted. Modern radiotherapy treatment techniques such as Three-Dimensional Conformal Radiotherapy (3DCRT), Intensity Modulated Radiation Therapy (IMRT), Image Guided Radiotherapy (IGRT) and Stereotactic Radiosurgery (SRS) aim to deliver maximum radiation dose to the cancer cells and the minimum radiation dose to the surrounding critical structures and normal tissues. Therefore, various QA tests should be conducted on Linear Accelerator for ensuring accurate dose delivery [3]. The response of radiotherapy depends on various factors, one of them being the accuracy of the location of irradiation on the planning to the implementation of performed irradiation. Verification of radiation isocenter has a very important role on the quality of radiation therapy delivered [4]. Radiation isocenter is a very fundamental part of commissioning and QA testing because isocenter of linac locates the area of irradiation in radiotherapy [5]. Basically isocenter is an imaginary point which is at 100 cm distance from the target of medical linac placed along central axis of beam. The axes of rotations of gantry, collimator and couch meet at this point which is called as isocenter. Verification of isocenter radiation aims to determine Linac beam diameter and it should be checked periodically.

In the radiotherapy treatment, equipment used should be very accurate. The proper position of isocenter mostly uses positional mechanism of collimator, gantry, and couch that aim to optimize the target range and minimize exposure of surrounding healthy tissues [6]. As per Task Group 40 American Association of Physicists in Medicine (AAPM), verification of radiation isocenter is conducted periodically [7]. Also recommendation provides tolerance limit of radiation isocenter to be 2 mm, because the radiation isocenter can be shifted during the gantry, collimator and couch rotation [8]. Therefore it is necessary to verify the radiation isocenter on the X-ray beam plane of 6 MV linear accelerator using therapeutic film with digital imaging process.

MATERIALS AND METHODS

In our study we used a Primus Linac (Siemens) having single energy which is available in our Centre and it is used since fourteen years for cancer treatment. Now-days around 40 patients are treated daily by this linac. Kodak EDR2 therapeutic films were used to expose the radiation with star patterns by rotation of gantry, collimator and couch. In star patterns we used angles 00 , 300 , 600 , 900 , 1200 and 1500 for collimator, gantry and couch. Then the exposed films were processed manually in dark room. For scanning of processing films we used EPSON perfection V700 scanner for digital images. A software (PTW, ISOCHECK 1.1) was used for calculation and positional. Isocenter is very important in radiotherapy treatment using a Linac. Isocenter QA in radiotherapy Linac is a star shot method [5]. For the radiation isocenter star shot method is shown in Figure 1.

A stellar beam shot is obtained by placing the film in the field of radiation by rotational motion of a gantry, collimator and a couch. Film will be exposed to radiation from different angles of rotation so that the beams intersect to form stars. Radiation isocenter verification is done by determining the smallest circle of the intersection of all the fields, with the center of the circle called radiation isocenter and the radius of the circle called the size of the radiation isocenter. To determine shifts radiation isocenter can be seen from the result of the diameter of the beam intersecting circle of irradiation [6]. The positional consistency of radiation isocenter was checked five times within last one year with the gap every 2-3 months by rotation of Gantry, collimator and the couch of the medical Linac in our centre. As per AAPM report 40, the size of the Linac radiation isocenter diameter should not be more than 2 mm diameter.

RESULTS AND DISCUSSION

Positional consistency of radiation isocenter has been performed by using star shot image where the radiation beam shaped like a star due to the rotation of one of the Linac component, the gantry, the collimator and the couch individually. To determine the diameter of the radiation isocenter, image processing was done by using PTW isocheck 1.1 software. The physical visualization of radiation isocenter of collimator, gantry and couch are shown in Figure 2 after processing by the software.

Positional determination of radiation isocenter diameter on Linac is very important and will have an impact on the quality of radiation therapy given to patients. For accurate radiotherapy treatment delivery, gantry, collimator, and couch positions combined are checked properly during QA of Linac. If the positional accuracy of radiation isocenter is within the specified limit then it is clear that the positional accuracies of gantry, collimator, and couch are also within permissible limit. The results of verification of radiation isocenter are shown in Table 1 for all five experiments. As per result the maximum diameter length of isocenter for collimator rotation angles obtained is 0.6mm and mean value is 0.42mm with standard deviation of 0.11. Similarly for gantry rotation angles obtained is 0.6mm and mean value is 0.46mm with stand deviation of 0.10. It has also been found that for couch rotation angles the maximum diameter of isocenter is 0.70mm and mean is 0.48 mm with standard deviation of 0.14. Our results are very much similar with the results which were reported by Muhammad Irsal et al, [10]. The shape of radiation isocenter diameter causes enlargement of penumbra or increases radiation dose on irradiation field shape, especially for small field treatments. This can happen because of asymmetry and instability of the tools used. Based on the use of Linac for radiotherapy irradiation, collimator often used for irradiation method on every patient is different. The positional consistency of radiation isocenter is in very good agreement as the standard deviation values of all three cases are very small. This means the positional inaccuracy of isocenter is very very less and it is well within the limit prescribed by AAPM report 40. The size of the Linac radiation isocenter diameter used in this study is still in good agreement because the value of the radiation isocenter is less than 2 mm diameter.

CONCLUSION

In view of our experimental data the verification of radiation isocenter performed on collimator, gantry and the couch, it shows that the size of radiation isocenter diameter for collimator is 0.42 mm, for gantry 0.46mm and 0.48 mm for the couch. On the basis of AAPM report 40, our result is extremely good and it is well within the permissible limit as our Linac is very old and we are using this since fourteen years for cancer treatment in our centre.

ACKNOWLEDGEMENTS

The authors would like to acknowledge all staff of the hospital for their assistance and continuous encouragement in making this work possible.

REFERENCES

1. Choirul A. Monte Carlo Simulation for Electron Contamination On Linac Beam 6 MV Aircraft Production Elekta SL15 Thesis Mathematic and Natural Science Medical Physics of Post Graduate Programme (Jakarta: University of Indonesia). 2011.

2. Deinar F. Dose Distribution to Asymmetry Field 6 MV X-rays with Depth Variation Thesis Mathematic and Natural Science Medical Physics of Post Graduate Programme (Jakarta: University of Indonesia). 2010.

3. Peace T, Subramanian B, and Ravindran P. An Experimental Study on Using a Diagnostic Computed Radiography System as a Quality Assurance Tool in Radiotherapy Australasian Physical and Engineerig. Sci Med. 2008; 313: 216-234.

4. Yin Z, Kai D, Garth C, Erik T, Elwood A, Ken KW. Alignment of Multiradiation Isocenter for Megavoltage Photon Beam. J Applied Clin Med Physics. 2015; 166: 314-324.

5. Gonzalez A, Castro I, and Martinez AJ. A Procedure to Determine the Radiation Size in a Linear Accelerator. Med Phys. 2004; 316: 1489- 1493.

6. Depuydt T, Penne R, Verelle D, Hrbacek J, Lang S, Leysen K, et al. Computer-Aided Analysis of Star Shot Films for High-Accuracy Radiation Therapy Treatment Unit IOB Publishing. Phys Med Biol. 2012; 57: 2997-3011.

7. Gerald J K, Lawrence C, Michael G, William F H, Steven L, Robert J M, et al. Comprehensive QA For Radioation Oncology: Report of Taskgroup 40. Med Phys. 1994; 21: 581-618.

8. Weiliang D, Jennifer LJ, Wei J, dan Rajat JK. On the Selection of Gantry and Collimator Angles for Isocenter Localization Using Winston-Lutz Tests. J Appl Clin Med Phys. 2016; 17: 167-178.

9. Muhammad Irsal. Verification of Radiation Isocenter on Linac Beam 6 MV using Computed Radiography. J Phys. Conf Ser. 2017.

Received : 20 Mar 2018
Accepted : 15 May 2018
Published : 17 May 2018
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X