Staff Organ Dose Assessment and Cancer Risk Estimation in a Cath-Lab Room

Review Article | Open Access | Volume 11 | Issue 1

  • 1. Postgraduate School for Fundamental and Applied Sciences, University of Douala, Cameroon
  • 2. School of Nuclear and Allied Sciences, University of Ghana, Ghana
  • 3. Radation Protection Institute, Ghana Atomic Energy Commission, Ghana
  • 4. National Radiation Protection Agency (NRPA), Cameroon
  • 5. Nuclear Regulatory Authority, Ghana
  • 6. Faculty of Science, University of Maroua, Cameroon
+ Show More - Show Less
Corresponding Authors
Ruth Njantang Nana, Postgraduate School for Fundamental and Applied Sciences, University of Douala, P.O. Box 24157, Douala, Cameroon
ABSTRACT

Staff organ doses were assessed and cancer risk were estimated for varying x-ray tube angulation in a cath-lab room at the emergency department of a hospital. Themoluminescence dosimeters (TLD) were used for the staff’s eye lens, thyroid, chest and gonad dosimetry. A fabricated wood and Perspex phantom were used to represent the staff and the patient respectively. The chest (cardiologist – 0.08 mSv), thyroid (nurse – 0.13 mSv), thyroid (cardiologist – 3.26 mSv) received the highest doses comparatively at AP 0o, LAO 45o and RAO 45o x-ray tube angulation respectively. Generally, the radiation dose to the staff was high for x-ray tube angulation of right anterior oblique (RAO) 45o. The thyroid organ recorded the highest cancer incidence risk comparatively for anterior posterior (AP) 0o, left anterior oblique (LAO) 45o and RAO 45o x-ray tube angulation, and for both sexes. However, generally, the cancer incidence risk estimates for females were higher than that of the males by a maximum factor of 3.2. The highest cancer risk incidence was 342.30 per 105 and 172.78 per 105 for female and male population respectively exposed in the thyroid organ. It is therefore recommended that staff doses should be optimised by using protective equipment (i.e. lead thyroid shield) to enhance staff protection.

KEYWORDS
  • Skin dose
  • Staff dose
  • Fluoroscopy guided procedures
  • Thermoluminescent dosimeters
CITATION

Nana RN, Gyekye PK, Inkoom S, Ndontchueng MM, Motapon O, et al (2023) Staff Organ Dose Assessment and Cancer Risk Estimation in a Cath-Lab Room. J Radiol Radiat Ther 11(1): 1097.

ABBREVIATIONS

PMMA: polymethylmethacrylate; TLD: Themoluminescence dosimeters; Hp: dose equivalent; RAO: Right Anterior Oblique; LAO: Left Anterior Oblique; AP: Anterior Posterior; BEIR: Biological Effect of Ionising Radiation.

INTRODUCTION

The use of fluoroscopy for the purpose of diagnostic and treatment have witnessed substantial improvement in medical image technology. New digital detectors coupled with sophisticated, dedicated software and large advances in computing technique that is used to obtain valuable clinical information within medical images have facilitated the expansion of fluoroscopy guided procedures [1].

Fluoroscopy diagnostic procedures are becoming a routine in many centers around the world. This is because it results in successful clinical outcome and better patient safety. Fluoroscopically-guided procedures lead to a significant number of prolonged procedures, which results in increased radiation exposure to patients and operators [2-4].

Among other categories of professional exposure, occupational exposure in fluoroscopy is one of the areas in which increased exposure is likely to occur, in particular for certain organs as extremities and eye lens [5,6]. In most situations radiation exposed workers in fluoroscopy are monitored using a single dosimeter for whole body dose assessment or using two dosemeters positioned under and above the lead apron (double dosimetry) according to international professional organizations recommendations [7]. Studies [8,9], have highlighted some drawbacks in monitoring occupational exposure in fluoroscopy such as lack of comprehensive personal dose records, in appropriate use of dosimeters etc.

This study aimed at assessing staff doses (cardiologist and nurse) at different angulation of the x-ray tube for cancer risk estimation during fluoroscopy guided procedures in the cath-lab using a wooden fabricated staff.

MATERIALS AND METHODS

This study was conducted in a newly commissioned Cath-Lab room at the emergency department of a hospital. The Cath-Lab room is equipped with an under-couch Philips Azurion 7 M20 Medical Systems (Serial Number: 173965; Total filtration: 2.5 mm of Al; Maximum tube voltage: 125 kV) fluoroscopy tube. All the necessary approvals (i.e. facility and municipal approvals) were obtained in order to conduct this study.

An acrylic polymethylmethacrylate (PMMA) phantom with dimensions 25 (length) x 25 (breath) x 20 (height) cm was used to represent the patient in order to provide the needed scatter radiation to the staff. A fabricated wood of 170 cm height with arms was used to represent the staff. Calibrated four (4) Themoluminescence dosimeters (TLD)’s were placed on the fabricated wood at different heights: 165 cm, 155 cm, 135 cm and 80 cm to measure the dose at the height of the eyes, thyroid, chest and gonad respectively as shown in Figure 1. The TLDs were  calibrated in terms of shallow dose equivalent (Hp) (0.07) and deep dose Hp (10) prior to their use. The TLDs tolerance range of deviation of the element correction factors were ± 30 %. The eye lens and thyroid TLDs were unshielded but the chest and the gonad TLDs were placed under a lead apron of 0.5 mm thickness of lead as practically positioned. The organ doses were estimated from the evaluation of doses measured using the TLDs. However, the eye lens dose was measured in terms of Hp (3). Therefore, Hp (3) was estimated from the Hp (0.07) eye dosimeter reading using equation 1.

Where: μ/ρ represents the mass attenuation coefficient of the eye lens, ρ represent the density of the organ and x represents the depth (distance) considered.

Water has mostly been used as the best representative of the human body. Therefore, equation 2 was used to correct the measured personnel equivalent doses on wood (representative staff) to water [10,11].

Where (μ/ρ)wood and (μ/ρ)water are mass attenuation coefficient for wood and water respectively.

Set up measurement in the Cath-Lab showing staff position for measurements behind the lead shield and lead curtain.

Figure 1 Set up measurement in the Cath-Lab showing staff position or measurements behind the lead shield and lead curtain.

The staff dose was measured at three (3) different angulations of the x-ray tube: Right Anterior Oblique (RAO) 45o, Left Anterior Oblique (LAO) 45o and Anterior Posterior (AP) 0o. The staff standing positions considered were: (1) on the left side of the patient couch behind the lead glass and the lead curtain where the cardiologists are likely to stay during procedure, and (2) on the right of the patient couch at one (1) meter from the patient without any additional shield where the nurses are likely to stay. The phantom was screened for five (5) minutes and five (5) radiographic shots for each of the identified angulation of the x-ray tube. Organ specific cancer risk incidence was estimated from the measured organ doses using the excess relative risk model of Biological Effect of Ionising Radiation (BEIR) committee VII phase 2 report [12]. The cancer risk incidence for all solid cancers except thyroid was estimated using equation 3. The cancer risk incidence for the thyroid was also estimated using equation 4 and 5.

Where ERR is excess relative risk; βs is ERR per unit of dose expressed in Sieverts, which is dependent on sex (s); D is absorbed dose; γ is per-decade increase in age at exposure over the range 0 – 30 years; e is age at exposure (30 years); a is attained age at exposure (60 years); and η is exponent of attained age.

Where ERR is excess relative risk; Gy is dose in gray; and e is age at exposure (30 years).

RESULTS AND DISCUSSION

The radiation dose to the identified organs of the cardiologist and the nurse for different orientation of the x-ray tube has been presented in Table 1. It can be seen that for x-ray tube angulation of AP 0o radiation dose to the chest of the cardiologist was higher than that to the eye lens, thyroid and gonad by a factor of 2.0, 1.1 and 1.3 respectively. It can also be seen that for x-ray tube angulation of LAO 45o , the dose to the eye lens of the cardiologist was more than that to the thyroid, chest and gonads by a factor of 1.6, 1.8 and 1.8 respectively. However, the nurse received the highest dose to the thyroid and it was more than the dose to the eye lens, chest and gonads by a factor of 1.1, 2.6 and 1.6 respectively. The x-ray tube angulation at RAO 45o recorded the highest dose of 3.26 mSv to the thyroid of the cardiologist. The dose to the thyroid was more than that to the eye lens, chest and gonads of the cardiologist by a factor of 54.0, 4.7 and 4.7 respectively. For the same x-ray tube angulation of RAO 45o the nurse received a dose of 0.24 mSv to the thyroid that was more than that to the eye lens, chest and gonad by a factor of 1.1, 3.1 and 4.8 respectively.

Table 1: Cardiologist and nurse radiation dose at the level of the eye, thyroid, chest and gonad

 

X-ray

Tube Angulation

Dose Equivalent, Hp (0.07) (mSv)

Eye lens

Thyroid

Chest

Gonad

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

AP 00

0.04

-

0.07

-

0.08

-

0.06

-

LAO 450

0.11

0.12

0.07

0.13

0.06

0.05

0.06

0.08

RAO 450

0.06

0.22

3.26

0.24

0.70

0.07

0.07

0.05

(-) means not available

Abbreviations: RAO: Right Anterior Oblique; LAO: Left Anterior Oblique; AP: Anterior Posterior

It can be seen that the organs receiving the high doses varied with varying tube angulation for both the cardiologist and the nurse. This could be attributed to the proximity of these organs to the x-ray tube. The varying dose quantity to the cardiologist and nurse for the same x-ray tube angulation could be attributed to the x-ray tube orientation with respect to the staff positioning to the patient table and the protective equipment provided. Generally, tube angulation at RAO 45o recorded high doses to the cardiologist and the nurse. It is recommended that the cardiologist and the nurse should effectively use the protective equipment (i.e. protective goggles, thyroid shield and aprons) at all times during the procedures.

Excess relative risk of cancer incidence of the eye lens, thyroid, chest and gonads of female and male staff in the cath lab room has been presented in Table 2 and 3 respectively. It must be indicated that the models used for the cancer incidence are highly speculative because of various random and systematic uncertainties embedded in them [13].

Table 2: Excess relative risk cancer incidence for the eye lens, thyroid, chest and gonad of female staff

 

X-ray Tube Angulation

Excess Relative Risk (per 105)

Eye lens

Thyroid

Chest

Gonad

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

AP 00

1.80

-

7.35

-

3.60

-

2.28

-

LAO 450

4.95

5.4

7.35

13.65

2.70

2.25

2.28

3.04

RAO 450

2.70

9.90

342.30

25.20

31.50

3.15

2.66

1.90

(-) means not available

Abbreviations: RAO: Right Anterior Oblique; LAO: Left Anterior Oblique; AP: Anterior Posterior

Table 3: Excess relative risk cancer incidence for the eye lens, thyroid, chest and gonad of male staff

 

X-ray Tube Angulation

Excess Relative Cancer Risk (per 105)

Eye lens

Thyroid

Chest

Gonad

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

Cardiologist

Nurse

AP 00

1.08

-

3.71

-

2.16

-

0.72

-

LAO 450

2.97

3.24

3.71

6.89

1.62

1.35

0.72

0.96

RAO 450

1.62

5.94

172.78

12.72

18.90

1.87

0.84

0.60

(-) means not available

Abbreviations: RAO: Right Anterior Oblique; LAO: Left Anterior Oblique; AP: Anterior Posterior

At x-ray tube angulation of AP 0o, thyroid recoded the highest cancer incidence risk for both male and female staff. Also, for LAO 45o and RAO 45o x-ray tube angulation, the thyroid recorded the highest cancer incidence risk comparatively for both the cardiologist and the nurse, and the sexes. This suggest that irrespective the x-ray tube angulation in the cath-lab room, protection of the thyroid by using lead thyroid shield is very paramount. Generally, the cancer incidence risk estimates for females were higher than that of the males by a maximum factor of 3.2. The highest cancer risk incidence was 342.0 per 105 female population when exposed in the thyroid to a radiation dose of 3.26 mSv with age at exposure and attained age of 30 and 60 years respectively.

CONCLUSION

Staff doses have been assessed and cancer risk estimated for a cath lab room. It was observed that varied organs of the cardiologist and nurse received high doses comparatively for varying tube angulation. The chest (cardiologist – 0.08 mSv), thyroid (nurse – 0.13 mSv), thyroid (cardiologist – 3.26 mSv) received the highest doses at AP 0o, LAO 45o and RAO 45o x-ray tube angulation respectively. Additionally, the quantity of radiation dose to the cardiologist and nurse varied for the same x-ray tube angulation. Generally, the radiation dose to the staff was high for x-ray tube angulation at RAO 45o. The thyroid organ recorded the highest cancer incidence risk comparatively for AP 0o, LAO 45o and RAO 45o x-ray tube angulation, and for both sexes. However, generally, the cancer incidence risk estimates for females were higher than that of the males by a maximum factor of 3.2. The highest cancer risk incidence was 342.30 per 105 and 172.78 per 105 for female and male population respectively exposed in the thyroid organ. It is therefore recommended that staff doses should be optimised by using protective equipment (i.e. lead thyroid shield) to enhance staff protection.

ACKNOWLEDGEMENTS

The authors would like to acknowledge the radiographer and the medical physicist of the hospital and the staff of school of nuclear and allied sciences for their support during the data collection.

REFERENCES
  1. Sechopoulos I, Trianni A, Peck D. The DICOM Radiation Dose Structured Report: What It Is and What It Is Not. J Am Coll Radiol. 2015 ; 12: 712-713.
  2. Hassan AE, Amelot S. Radiation exposure during Neurointerventional procedures in modern biplane angiographic systems: a single-site experience. Interv Neurol. 2017; 6: 105-116.
  3. Kirkwood ML, Guild JB, Arbique GM, Anderson JA, Valentine RJ, Timaran C. Surgeon radiation dose during complex endovascular procedures. J Vasc Surg. 2015; 62: 457-463.
  4. Stewart FA, Akleyev AV, Hauer-Jensen M, Hendry JH, Kleiman NJ, Macvittie TJ, et al. ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs–threshold doses for tissue reactions in a radiation protection context. Ann ICRP. 2012; 41: 1-322.
  5. Vano E, Kleiman NJ, Duran A, Rehani MM, Echeverri D, Cabrera M. Radiation Cataract Risk in Interventional Cardiology Personnel. Radiat. Res. 2010; 174: 490-495.
  6. Vano E, Kleiman NJ, Duran A, Romano-Miller M, Rehani MM. Radiation- Associated Lens Opacities in Catheterization Personnel: Results of a Survey and Direct Assessments. J Vasc Interv Radiol. 2013; 24: 197- 204.
  7. International Commission on Radiological Protection. Publication 103: The 2007 Recommendations of the International Commission on Radiological Protection. Ann ICRP. 2007; 37: 2-4.
  8. Vano E, Fernandez JM, Sanchez R, Dauer LT. Realistic Approach to Estimate Lens Doses and Cataract Radiation Risk in Cardiology When Personal Dosimeters Have Not Been Regularly Used. Health Phys. 2013; 105: 330-339.
  9. Ciraj-Bjelac O, Rehani MM. Eye Dosimetry in Interventional Radiology and Cardiology: Current Challenges and Practical Considerations. Rad Prot Dosim. 2014; 162: 329-337.
  10. Chu-Ki K, Jung-Kwon O, Jung-Pyo H, Jun-Jae L. Density calculation of wood by portable X – ray tube with consideration of penetrating depth. J wood Sci. 2014; 60: 105-110.
  11. Hubbell JH, Seltzer SM. Tables of x – ray mass attenuation coefficients (version 1.4). National Institute of Standards and Technology, Gaithersburg, MD. 2004.
  12. National Research Council. Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII Phase 2. The National Academies Press. Washington, DC. 2006.
  13. Pradhan, AS. On the risk to low doses (<100 mSv) of ionizing radiation during medical imaging procedures - IOMP policy statement. J Med Phys. 2013; 38: 57-58.

Nana RN, Gyekye PK, Inkoom S, Ndontchueng MM, Motapon O, et al (2023) Staff Organ Dose Assessment and Cancer Risk Estimation in a Cath-Lab Room. J Radiol Radiat Ther 11(1): 1097.

Received : 23 Jan 2023
Accepted : 24 Feb 2023
Published : 26 Feb 2023
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
JSM Ophthalmology
ISSN : 2333-6447
Launched : 2013
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X