Progressive Myopia Dynamics. Analog Computer Model of Refraction Development

Short Communication | Open Access | Volume 5 | Issue 3

  • 1. Department of Bioengineering, BGKT Consulting Ltd., USA
  • 2. Department of Kinesiology, University of Zagreb, Croatia
+ Show More - Show Less
Corresponding Authors
Greene PR, Department of Bioengineering, BGKT Consulting, Huntington, New York, 11743, Tel: 1-631-935-56-66;
Keywords

Progressive myopia , Analogue circuit , RC time constant , Accommodative demand , Reading glasses , Refraction , Bifocals , Progressive add lenses

ABBREVIATIONS

R(t): Refractive Error [diopters] as a function of time; t: time [years]; R(0): Initial Refraction at t = 0; E(t): environment near-point reading demand [0 to -3 diopters]; τ: System Time Constant [60 to 100 days]; V(t): Voltage as a Function of Time; PAL = Progressive Addition Reading Glasses [+2 to +3 diopters]; D: Diopters: [1 / meters]; C: Capacitance [micro-Farads]; R: resistance [kilo-Ohms]

INTRODUCTION

Although a well-defined and well-documented technical problem, progressive myopia continues to elude the best efforts of our researchers world-wide. In practical clinical terms, -0.75 to -1.0 diopter of myopia corresponds to 3-lines on the Snellen eye chart, typically the difference between 20/20 and 20/10, or the difference between 20/40 and 20/20. Figure 1 and 2 below schematically show the -4 diopter demand on the eye’s focusing system caused by studying, typical of some college students, and the proposed optical solution to this problem, by counter-acting the near point environment with a compensating (+) 3 D lens.

The problem of human progressive myopia is an intriguing, as yet unsolved, multi-parameter mystery. Undoubtedly a part of the explanation is inherited factors, part is environmental, as evidenced by nearwork myopia [1,2]. At present, myopia affects 25% to 40% of the U.S. population, perhaps 50% to 70% of the S.E. Asia area. Fairly harmless at less than - 6 diopters, the more advanced cases, - 7 diopters and beyond, may encounter a variety of medical problems, including macular impairment, detached vitreous or retina, staphyloma in the optic nerve region, loss of best-corrected visual acuity to less than 20/200, glaucoma, and cataract later in life [2,3].

Literature review – Walline [4], Flitcroft [2], Schmid [3], and Viikari [1], present comprehensive reviews of the myopia literature, including various techniques for myopia prevention. Holden et al. [5], review experimental research efforts concerning the myopia problem, showing that the human myopia progression-rate can be reduced by 55% using various types of (+) add lenses, shown here in Figure 3. Goss [6], reviews the effects of lenses on myopia progression in children. Cheng et al. [7], review the use of (+) lenses on myopia progression, citing all of the latest available literature on the subject. Myopia control and prevention is of extreme interest in the military, especially for air force (USAF) and navy pilots (USNA), [8-12].

It should be pointed out that these various myopia prevention techniques do not always work as effectively as we might like. One of the main points of emphasis in this report, is that the eye’s focusing control system seems to require a very special set of optical parameters, in terms of magnitude of correction (-2 to -3 diopters) and duration (equal to the total hours of study), in order to retain refraction stability over a long period of time (i.e. 4 to 8 years, typical of undergraduate and/or graduate work). In particular, it can be shown that the optical visual activity factor on the student’s visual system is quantified as an integral of the number of “diopter-hours”, the accumulated optical “impulse” demand. Evidently, without some sort of intervention procedure, the college student’s visual system will drift further into myopia, typically at a rate of -0.3 to -0.5 diopters/year. It is fairly common to find a typical college myope, graduating 4 years later, with an additional -2 diopters of progressive myopia, often progressing from a nominal -4 diopters to -6, or even -5 to -7 diopters.

Recently, several (+) Add research studies have been published [13-18,7] with encouraging results, showing the progressive myopia-rate can be attenuated by 55% [5], using various (+) Add technologies, i.e. reading glasses, bifocals, or progressive addition lenses (PALs)

There are several reports of myopia developing with Navy submariners, and extensive LASIK use in the US Army [11] to cure the myopia problem, more than 16,000 recruits as of 2003, a total of 26,000 recruits as of 2005, a remarkably large study. The Annapolis Navy pilots at the United State Naval Academy (USNA) are required to be in excellent physical condition, it is a demanding job flying a Mach 2 fighter/bomber. These various (+) Add lenses can optically shift a book or computer at 13 to 20 inches to infinity, thereby reducing the stress on the accommodation system. Various types of reading glasses, i.e. single vision, bifocals, and the new multi-focal progressive lenses (PALs) may be a practical way to stabilize college myopia [19], and pilot myopia [8,9,12]

Nearwork has been proposed as a factor in myopia progression. Near work, such as intense computer work, can increase myopia [12]. Feedback Theory predicts a causative relationship between nearwork and myopia, as nearwork is equivalent to added minus (negative) lenses while looking in the distance. The effective nearwork distance, in diopters, is added as a step input demand-signal to the 1st order control system, which then responds by increasing the myopia.

The purpose of this report is to present a simplified solution technique to solve the differential equations that model progressive myopia. This model for refractive error development of the human eye requires that there is an optical return signal equal to the refractive error, which in turn corrects the refractive error of the eye. The first-order feedback system used here, defined by the transfer function F(s) = 1 / (1+ks), was proposed by Medina & Fariza [20].

MATERIALS AND METHODS

Acquired myopia equations

The human eye has a natural tendency to become myopic with long hours focusing at a near-point environment [21,19,10]. During the course of our college work, exploring reading glasses to stabilize myopia, the most difficult problem has been teaching and explaining the rationale of this optical technique [12]. Students grasp these ideas more readily, using differential equations and electrical engineering concepts, Figure 3 and Figure 4, show an analogue RC circuit simulating myopia progression vs. time R(t). The inset shows an applied negative square-wave representing the daily accommodative nearwork demand signal [20,22]. Equating Volts ~ Diopters, the system exhibits an exponential shift to more myopia [23,24], typically acquiring -0.3 to -0.4 D / year at college, with time constant τ = 60 to 100 days, usually in steps of - 0.5 diopters per year:

Equation 1: R ( t ) = - 5.00 D - 1.50 D x [ 1 – exp (- t / τ ) ] ( DC = 0.5 )

 

 

 

RESULTS

Analog-circuit solution

The electrical circuit, Figure 3, simulates myopia progression vs. time as the voltage at the capacitor, where Volts (v.) represent diopters (D). Initially, as a typical example, the subject’s myopia is specified at - 5.00 D (V) and a negative square-wave varying from 0 to - 3.0 volts is applied, Figure 4, with variable duty-cycle (DC) intermittency factor [3] representing a daily accommodative nearwork demand (DC = 0.5) as the subject progresses from - 5.00 D to - 5.50 D. Medina [25], uses Laplace transform techniques solving 1-st order equations to calculate the drift of refractive status R(t).

Experimental details

Hobbyists with bread-board experience can easily build this analog circuit, using a digital voltmeter and frequency counter (Fluke DVM and Yaesu 500 MHz counter used here), analog voltmeter (Minolta from Radio Shack used here), or dual-beam oscilloscope (Tektronix 100 MHz and Hewlett Packard 60 MHz used here) comparing input and output voltages to monitor accumulated capacitor voltage Vc(t) proportional to the acquired myopia R(t). The square-wave is supplied with a low-frequency function generator (Wavetek 2-MHz and Heathkit 1-MHz used here) oscillating at 2-3 cycles/sec.

The economy oscillators only provide a 50% duty-cycle. One can scratch-build an astable multi-vibrator circuit [25], to provide a variable-intermittency variable duty-cycle square-wave. Some of the advanced “laboratory grade” oscillators include a DCoffset control, so the extra batteries in the Figure 3 circuit are already included. As a practical matter, we use a C = 2,000 μF capacitor, with an R = 30 K Ω resistor, having a time constant τ = RC = 60 s. = 1 min. Each day is accelerated or time-compressed to just 1-second, 2 months takes 1 minute, 1 year takes 6 minutes. Thus, 1-year, accelerated on the laboratory bench, corresponds to 6 time constants (τ = 60 d.) for the visual system. Because the circuit is symmetric with respect to (+) and (-) voltages, the conventional (-) polarity sign convention for concave lenses need not be installed. Conceptually, it is somewhat easier to assemble the circuit using (+) voltages for all components. Then, in order to display the negative going exponential sawtooth, it suffices to reverse the polarity of the oscilloscope probe.

 

DISCUSSION

Practical applications

The refraction response remains stable, R(t)=const., when using (+) Add lenses of strength +3.00 diopters, Figures 1, 2, 3 thus stabilizing the progressive myopia. This is because the battery switch, simulating reading glasses of one type or another, is synchronized with the square-wave input cycle, producing a null voltage, Figure 3. A practical clinical result is that different types of (+) reading glasses [5], have the capability to optically shift a book or computer screen from a typical reading distance of 1/3 meter (13 inches) to infinity, thereby reducing the focusing demand on the visual system.

 

CONCLUSIONS

This type of series electrical circuit is termed an accumulator or “integrator” circuit. Viewing myopia development in terms of this equivalent analog-circuit, the system responds to the applied square-wave cyclic load with a decaying sawtooth exponential, Figure 4, as can be demonstrated on an oscilloscope. This RC circuit is a minimum complexity single-parameter model, requiring only the subject’s time constant τ. Circuits of this type demonstrate the required magnitude and duration of (+) lens use, necessary to counter-act a -3.0 diopter student work load.

Emmetropization

Medina et al. [19,20], find that the individual’s emmetropization-rate and emmetropization time-constant determine the ensuing progressive myopia-rate. This means that an individual showing rapid emmetropization early in life, and then later becoming myopic, may also demonstrate rapid myopiarates, indicating that an advanced case of progressive myopia can be predicted decades ahead of time. This “rate-equivalence” between emmetropization and myopia is consistent with the results presented here, indicating that progressive myopia is a matter of the recently corrected eye continually re-calibrating its parameters to achieve best possible focus, averaged throughout the day.

 

SUMMARY

Herein, a simple R-C series analog circuit is presented to solve the basic 1-st order myopia equations, showing how nearwork causes a myopizing progression, additional to that created by the constant value negative lens used to correct myopia. This analogcircuit technique is a general result, which as a practical matter, can account for many different variations of system parameters and initial conditions, including the patient’s initial myopia (-5.00 D in this example), the nearwork load factor (50% intermittency in this example), the strength of reading glasses (+3D “Add” in this example) and the individual’s emmetropization timeconstant (60 days in this example). Each individual is different, so it is important to build this degree of flexibility into the circuit. In Figure 3, a -5 v battery is used to simulate -5 diopters of myopia, and a -3 v alternating square wave simulates the nearwork demand. The system response to reading glasses, a type of undercorrection [26-28], is solved simply by replacing the -5 v battery with -2 v.

ACKNOWLEDGEMENTS

This work was supported in part by NIH NEI Grant EY 005013. Special thanks to Bin He, Otis Brown, Antonio Medina, Frank Young, Bill Kranz, and Tom McMahon for many helpful discussions.

REFERENCES

1. Viikari K. Learn to understand & prevent myopia. BoD-Books on Demand. 2011.

2. Flitcroft DI. The complex interactions of retinal, optical and environmental factors in myopia aetiology. Prog Retin Eye Res. 2012; 31: 622-660.

3. Cheng D, Woo GC, Drobe B, Schmid KL. Effect of bifocal and prismatic bifocal spectacles on myopia progression in children: three-year results of a randomized clinical trial. JAMA Ophthalmol. 2014; 132: 258-264.

4. Walline JJ. Myopia Control: A Review. Eye Contact Lens. 2016; 42: 3-8.

5. Holden B, Sankaridurg P, Smith E, Aller T, Jong M, He M. Myopia, an underrated global challenge to vision: where the current data takes us on myopia control. Eye (Lond). 2014; 28: 142-146.

6. Goss DA. Effect of spectacle correction on the progression of myopia in children--a literature review. J Am Optom Assoc. 1994; 65: 117-128.

7. Cheng D, Woo GC, Schmid KL. Bifocal lens control of myopic progression in children. Clin Exp Optom. 2011; 94: 24-32.

8. Hoogerheide J, Rempt F, Hoogenboom WP. Acquired myopia in young pilots. Ophthalmologica. 1971; 163: 209-215.

9. Miller RE, Woessner WM, Dennis RJ, O’Neal MR, Green RP. Survey of spectacle wear and refractive error prevalence in USAF pilots and navigators. Optom Vis Sci. 1990; 67: 833-839.

10. Gmelin RT. Myopia at West Point: past and present. Mil Med. 1976; 141: 542-543.

11. Hammond MD, Madigan WP, Bower KS. Refractive surgery in the United States Army, 2000-2003. Ophthalmology. 2005; 112: 184-190.

12. Greene PR, Grill ZW, Medina A. Mathematical Models of College Myopia. Optik (Stuttg.). 2016; 127: 896-899.

13. Gwiazda J, Hyman L, Hussein M, Everett D, Norton TT, Kurtz D, et al. A randomized clinical trial of progressive addition lenses versus single vision lenses on the progression of myopia in children. Invest Ophthalmol Vis Sci. 2003; 44: 1492-1500.

14. Leung JT, Brown B. Progression of myopia in Hong Kong Chinese schoolchildren is slowed by wearing progressive lenses. Optom Vis Sci. 1999; 76: 346-354.

15. Fulk GW, Cyert LA, Parker DE. A randomized trial of the effect of single-vision vs. bifocal lenses on myopia progression in children with esophoria. Optom Vis Sci. 2000; 77: 395-401.

16. Oakley KH, Young FA. Bifocal control of myopia. Am J Optom Physiol Opt. 1975; 52: 758-764.

17. Yang Z, Lan W, Ge J, Liu W, Chen X, Chen L, et al. The effectiveness of progressive addition lenses on the progression of myopia in Chinese children. Ophthalmic Physiol Opt. 2009; 29: 41–48.

18. COMET Group. Myopia stabilization and associated factors among participants in the Correction of Myopia Evaluation Trial (COMET). Invest Ophthalmol Vis Sci. 2013; 54: 7871-7884.

19. Lin LL, Shih YF, Lee YC, Hung PT, Hou PK. Changes in ocular refraction and its components among medical students--a 5-year longitudinal study. Optom Vis Sci. 1996; 73: 495-498.

20. Medina A, Fariza E. Emmetropization as a first-order feedback system. Vision Res. 1993; 33: 21-26.

21. Fledelius HC. Myopia profile in Copenhagen medical students 1996- 98. Refractive stability over a century is suggested. Acta Ophthalmol Scand. 2000; 78: 501-505.

22. Greene PR, Medina A. Analogue Computer Model of Progressive Myopia–Refraction Stability Response to Reading Glasses. J Comput Sci Syst Biol. 2016; 9: 104.

23. Greene PR, Brown OS, Medina AP, Graupner HB. Emmetropia approach dynamics with diurnal dual-phase cycling. Vision Res. 1996; 36: 2249-2251.

24. Greene PR, Vigneau ES, Greene J. Exponential prevalence and incidence equations for myopia. Clin Exp Optom. 2015; 98: 210-213.

25. Ku DN, Greene PR. Scleral creep in vitro resulting from cyclic pressure pulses: applications to myopia. Am J Optom Physiol Opt. 1981; 58: 528-535.

26. Medina A. The progression of corrected myopia. Graefes Arch Clin Exp Ophthalmol. 2015; 253: 1273-1277.

27. Medina A. Detecting the effect of under-correcting myopia. Graefes Arch Clin Exp Ophthalmol. 2015; 254: 409-410.

28. Medina A, Greene PR. Progressive Myopia and Lid Suture Myopia are Explained by the Same Feedback Process: a Mathematical Model of Myopia. J Nat Sci. 2015; 1: e121

Citation

Greene PR, Medved V (2017) Progressive Myopia Dynamics. Analog Computer Model of Refraction Development. JSM Ophthalmol 5(3): 1058.

Abstract

Purpose: An AC – DC series electrical circuit is developed to simulate myopia progression R = R(t) [diopters], as the accumulated negative voltage at the series capacitor, with voltage Vc = V(t) [volts] representing the refractive error R(t) [diopters].

Methods: This circuit is an analogue computer which automatically solves 1st order system response equations for refractive status progression as a function of time, R(t). The new system voltage level is approached exponentially, typically with a time constant τ = RC = 60 to 100 days.

Results: This analogue computer technique is a general result, and can incorporate any combination of initial refraction R(0), refractive correction R1, and daily nearpoint demand - dR(t) to the visual system. As an example, the initial conditions for the subject’s myopia are specified at Vc = - 5.0 volts DC [R(0) = - 5.00D], with an applied negative square-wave varying from 0 to -3.0 volts AC, representing a typical daily accommodative near work demand.

Conclusions: In terms of practical applications, positive +Add reading glasses, bifocals, and progressive addition lenses (PALs) have the effect of optically neutralizing the near-point environmental demands, stabilizing the negative system drift towards progressive myopia.

Received : 04 Feb 2017
Accepted : 29 Aug 2017
Published : 31 Aug 2017
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X