JSM Ophthalmology

Visual Simulation of Retinal Images with Various Designs of Pinhole Contact Lenses

Research Article | Open Access | Volume 3 | Issue 1

  • 1. Department of Ophthalmology, Keio University School of Medicine, Japan
+ Show More - Show Less
Corresponding Authors
Kazuno Negishi, Department of Ophthalmology, Keio University School of Medicine, 35, Shinanomachi, Shinjuku-ku, Tokyo 160-8582, Japan, Tel: 81-3-3353-1211;
ABBREVIATIONS

PCLs: Pinhole Contact Lenses; VA: Visual Acuity; CCD: ChargeCoupled Device

INTRODUCTION

Presbyopia, which manifests from loss of accommodative function of the crystalline lens, is a ubiquitous visual disability of the aging eye. The demand for clear, comfortable near vision is increasing with the widespread use of devices such as mobile phones, tablets and computers. It was recently reported that presbyopia is related to difficulties with activities of daily living and resulting social impediments (social impediments should be described) [1] and that both distance and near visual impairment are independently associated with a poorer quality of life [2].

There have been many optical solutions to improve near vision for presbyopic patients. These solutions include not only the classic approaches of bifocals, trifocals, progressivepowered spectacles [3,4] or contact lenses (multifocal or mono vision) [5,6] but also a variety of surgical procedures, including corneal refractive surgeries [7,8] and intraocular lens [9-11] based surgeries. However, no procedures have gained general acceptance because of associated disadvantages, i.e., some have poor predictability, regression, limited effectiveness, and irreversibility, [12-14] while others are new and unproved over time

Recently, a small aperture corneal inlay has been reported to be useful for treating Presbyopia [15-21]. The inlay is designed to increase the depth of field using the principle of small aperture optics. However, it is not a perfect method, because patients may have adverse symptoms postoperatively including dry eye, halo, glare, or night vision disturbances. The explanation of the small aperture inlay was reported in several cases previously [22]. There also may be a less than optimum effect [23]. Moreover, irreversibility without surgical procedures is a major concern.

Non-surgical correction using a small aperture contact lens may have an advantage. We could only find one study on a contact lens based pinhole system [16]; however, the aim of that study was to assess the effects of different artificial pupillary designs on visual performance using a pinhole-based system as a surrogate for corneal inlays, and the lens designs considered were small with nearly pupillary size, 3.5-mm, 2.5-mm, and 1.6- mm diameter, central apertures for an 8.0-mm diameter opaque zone and a 1.6-mm central aperture for a 4.0-mm diameter opaque zone. The current study assessed the optimal design of pinhole contact lenses (PCLs) without refractive power as a noninvasive treatment, to obtain a full range of vision from far to near using a visual simulation system and to evaluate the optical performance.

MATERIALS AND METHODS

Visual Simulation System

Our system simulates the retinal image in an eye with a contact lens. The schematic of this system is shown in (Figure 1). The fundamental approach in this system involves simulating retinal images in eyes with a contact lens under conditions that closely resemble the actual clinical environment. Our system consists of a model eye combining lenses with an artificial pupil and a charge-coupled device (CCD) camera (EO-1312CCD, Edmund Optics Inc., Barrington, NJ) with a pixel size of 4.65 by 4.65 µm. The dimensions and refractive indexes of each surface in the model eye are provided in (Table 1). The total spherical aberration of this model eye was +0.32 µm for a 6-mm aperture diameter. The contact lens can be held in front of the model eye and t is easily exchangeable. The centering of the contact lens can be almost perfectly secured using the special contact lens holder designed for this study. The size of the artificial pupil between the combined lenses is also exchangeable. The images formed by the system can be detected by the CCD camera and observed on the monitor of a personal computer. The different PCLs were placed in front of the model eye, and the images detected by the CCD camera were evaluated.

Examination Settings

Visual simulations were performed using this system at 5, 1, and 0.3 meters through a 3-mm aperture using a Landolt visual acuity (VA) chart. Among the various sizes of Landolt charts, the 0.50 log MAR chart, which is equivalent to the telephone directory print size2 [4], was picked up to evaluate the contrast levels at each distance. The refraction of the model eye was set to -1.0 diopter [D], because it has been reported that the best compromise of depth of focus is obtained when the eyes have some residual myopic defocus (range -0.75 to -1.00 D) [17].

PCLs

This PCL was designed to increase the depth of field and to decrease blur due to refractive error simultaneously without refractive correction using the principle of small aperture optics. The PCL had a clear central zone in a 6.0-mm diameter opaque zone, which was designed to achieve a pinhole effect to obtain the depth of field. The PCL itself had no refractive power (Figure 2). The contact lenses were made of silicone (water content, 0%). The overall diameter and the base curve of the PCL were 14.0 mm and 8.5 mm, respectively.

Investigation of the Optimal Diameter of the Clear Central Zone

Visual simulations at near (0.3 m), intermediate (1 m), and distance (5m) were performed using PCLs with different clear central zones (2.0, 1.8, 1.6, 1.4, and 1.2 mm) to investigate the optimal diameter of the clear central zone to obtain the depth of field. As a control, visual simulation was also performed using a conventional contact lens with the same material, size, and curvature

Evaluation of Contrast Levels from Simulated Retinal Images

The contrast levels of the gaps of the Landolt VA charts in the simulated images were analyzed using Photoshop software (Adobe, San Jose, CA) to determine the optimal PCL design (Figure 3).

 

RESULTS AND DISCUSSION

Investigation of the Optimal Diameter of the Clear Central Zone

Figure 4 shows the simulated retinal images of the 0.5 log MAR chart with the PCLs and the control. Figure 5 shows the contrast levels of the gap of the Landolt’s ring in the simulated retinal image of the 0.5 log MAR chart with the PCLs and the control. The results showed that the PCLs could improve resolution for distance compared to the control, and PCLs with the 1.2- and 1.4- mm clear central zones could obtain the depth of field from far to near although the resolution at near was poor.

DISCUSSION

In this study, the retinal images for eyes with variously designed PCLs without refractive power to obtain a full range of vision from far to near were simulated objectively using a newly designed visual simulation system in our lab. Only one previous report was found on visual function that used a contact lens based pinhole system [16]. In that report, visual performances with four different types of PCL designs were evaluated on the same patient. However, in that study, PCLs were worn only on the non-dominant eye, and the measurements were performed binocularly and subjectively. Therefore, the results did not directly reflect the optical performance of the PCLs themselves.

In addition, the diameters of the central aperture were 1.6 mm and over notably different from our study

In our study we first assessed the optical performance of the PCLs themselves with various central zone sizes. Our results showed that the PCLs with only a 1.2- or a 1.4-mm clear central zone could keep the contrast level at all distances tested, although the contrast level at 0.3 m was relatively low.

Recently, a small aperture corneal inlay has been reported to be useful for treating Presbyopia [7,8,15,17-23]. Non-surgical correction of presbyopia has some advantages compared with surgical correction including a corneal inlay. First, the most important advantage is reversibility. Because PCLs are easily exchangeable, patients can try different designs and refractive powers of PCLs to determine the optimal prescription. Corneal inlays also are reversible; however, lifting a flap to remove the inlay or to modify the refraction may increase the risk of postoperative complications including for example, infection and epithelial in growth [25]. Thus, reoperation should be avoided. Another advantage is that PCLs do not require special equipment (e.g., surgical microscope, femtosecond laser, or other surgical instruments), because it is a non surgical treatment. Therefore, the technique can be performed easily in any setting, and it may be useful even in countries that do not have appropriate medical equipment

In contrast, the PCLs have some disadvantages. First, they require daily contact lens care. Therefore, patients who cannot care for their contact lens or are contact lens intolerant cannot use them. Another disadvantage is the difficulty of obtaining stable positioning of the PCL, because it usually moves after blinking. Tabernero and Artal [17] reported that decentration from the optimal position of about 0.5 mm could significantly reduce the retinal image quality and overall vision with a small Figure 4 Simulated retinal images of the 0.5 log MAR chart with the PCLs and the control. Figure 5 Calculated contrast levels of the gap of the 0.5 the logarithm of the minimum angle of resolution (logMAR) chart in the simulated images with the pinhole contact lenses with clear central zone of 2.0, 1.8, 1.6, 1.4, and 1.2 mm. M: Meters. aperture corneal inlay. Gatinel et al [18] reported that two patients implanted with a small aperture corneal inlay that were recentered 2 weeks and 3 weeks postoperatively had significant improvements in the VA and quality of vision. Those reports suggested the importance of centration in the small aperture system. Upward movement of the typical soft contact lens at the vertical meridian was reported to be 342 ± 155 μm after blinking, [26] which might negatively affect the results

The current study had some limitations. First, we did not assess the effect of decentration. In this model eye, the PCLs and the corneal and pupillary centers were matched perfectly in the model eye. However, clinically, the pupillary center sometimes differs from the corneal center, which almost corresponds to the PCL center. In addition, a PCL may become decentered after blinking as mentioned previously. These primary and secondary decentrations may affect the optical performance of the PCLs. Another limitation of this study is that the tested refraction of the model eye was only -1.00 diopter, which is reported the best compromise of depth of focus [17]. Further investigations should be performed at other settings in refraction.

CONCLUSION

In conclusion, a preliminary investigation utilizing a PCL without refractive power might be useful to obtain a full range of vision from far to near if designed optimally. The PCL’s with a1.2-mm or and 1.4-mm clear central zone maintained the best resolution and contrast of the simulated images at for all distances

ACKNOWLEDGEMENTS

2011FY NEDO Innovation Promotion Program. 0822001, Dated August 22, 2011.

REFERENCES

1. Lu Q, Congdon N, He X, Murthy GV, Yang A, He W. Quality of life and near vision impairment due to functional presbyopia among rural Chinese adults. Invest Ophthalmol Vis Sci. 2011; 52: 4118-4123.

2. Lamoureux EL, Fenwick E, Moore K, Klaic M, Borschmann K, Hill K. Impact of the severity of distance and near-vision impairment on depression and vision-specific quality of life in older people living in residential care. Invest Ophthalmol Vis Sci. 2009; 50: 4103-4109.

3. Sheedy JE, Hardy RF. The optics of occupational progressive lenses. Optometry. 2005; 76: 432-441.

4. Fowler C. Recent trends in progressive power lenses. Ophthalmic Physiol Opt. 1998; 18: 234-237.

5. Chapman GJ, Vale A, Buckley J, Scally AJ, Elliott DB. Adaptive gait changes in long-term wearers of contact lens monovision correction. Ophthalmic Physiol Opt. 2010; 30: 281-288.

6. Morgan PB, Efron N. Contact lens correction of presbyopia. Cont Lens Anterior Eye. 2009; 32: 191-192.

7. Uthoff D, Pölzl M, Hepper D, Holland D. A new method of cornea modulation with excimer laser for simultaneous correction of presbyopia and ametropia. Graefes Arch Clin Exp Ophthalmol. 2012; 250: 1649-1661.

8. Reinstein DZ, Carp GI, Archer TJ, Gobbe M. LASIK for presbyopia correction in emmetropic patients using aspheric ablation profiles and a micro-monovision protocol with the Carl Zeiss Meditec MEL 80 and VisuMax. J Refract Surg. 2012; 28: 531-541.

9. Friedrich R. Intraocular lens multifocality combined with the compensation for corneal spherical aberration: a new concept of presbyopia-correcting intraocular lens. Case Rep Ophthalmol. 2012; 3: 375-383.

10. Agresta B, Knorz MC, Kohnen T, Donatti C, Jackson D. Distance and near visual acuity improvement after implantation of multifocal intraocular lenses in cataract patients with presbyopia: a systematic review. J Refract Surg. 2012; 28: 426-435.

11. Pepose JS, Wang D, Altmann GE. Comparison of through-focus image sharpness across five presbyopia-correcting intraocular lenses. Am J Ophthalmol. 2012; 154: 20-28.

12. Ayoubi MG, Leccisotti A, Goodall EA, McGilligan VE, Moore TC. Femtosecond laser in situ keratomileusis versus conductive keratoplasty to obtain monovision in patients with emmetropic presbyopia. J Cataract Refract Surg. 2010; 36: 997-1002.

13. Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg. 2009; 35: 992-997.

14. Torricelli AA, Junior JB, Santhiago MR, Bechara SJ. Surgical management of presbyopia. Clin Ophthalmol. 2012; 6: 1459-1466.

15. Seyeddain O, Hohensinn M, Riha W, Nix G, Rückl T, Grabner G, Dexl AK. Small-aperture corneal inlay for the correction of presbyopia: 3-year follow-up. J Cataract Refract Surg. 2012; 38: 35-45.

16. García-Lázaro S, Ferrer-Blasco T, Radhakrishnan H, Cerviño A, Charman WN, et al. Visual function through 4 contact lens-based pinhole systems for presbyopia. J Cataract Refract Surg. 2012; 38: 858-865.

17. Tabernero J, Artal P. Optical modeling of a corneal inlay in real eyes to increase depth of focus: optimum centration and residual defocus. J Cataract Refract Surg. 2012; 38: 270-277.

18. Gatinel D, El Danasoury A, Rajchles S, Saad A. Recentration of a smallaperture corneal inlay. J Cataract Refract Surg. 2012; 38: 2186-2191.

19. Seyeddain O, Bachernegg A, Riha W, Rückl T, Reitsamer H, Grabner G, Dexl AK. Femtosecond laser-assisted small-aperture corneal inlay implantation for corneal compensation of presbyopia: two-year follow-up. J Cataract Refract Surg. 2013; 39: 234-241.

20. Dexl AK, Seyeddain O, Riha W, Hohensinn M, Rückl T, Reischl V, et al. One-year visual outcomes and patient satisfaction after surgical correction of presbyopia with an intracorneal inlay of a new design. J Cataract Refract Surg. 2012; 38: 262-269.

21. Dexl AK, Seyeddain O, Riha W, Hohensinn M, Rückl T, Hitzl W, et al. Reading performance after implantation of a modified corneal inlay design for the surgical correction of presbyopia: 1-year follow-up. Am J Ophthalmol. 2012; 153: 994-1001.

22. Yılmaz OF, Alagöz N, Pekel G, Azman E, Aksoy EF, Cakır H, et al. Intracorneal inlay to correct presbyopia: Long-term results. J Cataract Refract Surg. 2011; 37: 1275-1281.

23. Tomita M, Kanamori T, Waring GO 4th, Yukawa S, Yamamoto T, Sekiya K, et al. Simultaneous corneal inlay implantation and laser in situ keratomileusis for presbyopia in patients with hyperopia, myopia, or emmetropia: six-month results. J Cataract Refract Surg. 2012; 38: 495-506.

24. Sanders DR, Sanders ML. Near visual acuity for everyday activities with accommodative and monofocal intraocular lenses. J Refract Surg. 2007; 23: 747-751.

25. Chan CC, Boxer Wachler BS. Comparison of the effects of LASIK retreatment techniques on epithelial ingrowth rates. Ophthalmology. 2007; 114: 640-642

26. Cui L, Shen M, Wang MR, Wang J. Micrometer-scale contact lens movements imaged by ultrahigh-resolution optical coherence tomography. Am J Ophthalmol. 2012; 153: 275-283.

Citation

Nishi Y, Negishi K, Watanabe K, Tsubota K (2015) Visual Simulation of Retinal Images with Various Designs of Pinhole Contact Lenses. JSM Ophthalmol 3(1): 1025.

Abstract

Purpose: To determine the optimal design for pinhole type contact lenses (PCLs) without refractive power for obtaining a full range of resolution (clear vision) from far to near using a visual simulation system.

Methods: The PCLs had a 6.0-mm diameter black opaque central zone with varying clear central zones. The total diameter and the base curve of the PCL were 14.0 mm and 8.5 mm, respectively. The visual simulation system consisted of a model eye and a chargecoupled device camera. The different PCLs were placed in front of the model eye and evaluated. Visual simulations were performed using this system at 5, 1, and 0.3 meters through a 3-mm aperture using Landolt visual acuity (VA) charts with different PCLs (clear central zone sizes, 2.0, 1.8, 1.6, 1.4, and 1.2 mm). The contrast levels of the gaps of the Landolt VA charts in the simulated images were analyzed using Photoshop software to determine the optimal PCL design.

Results: The PCL’s with a1.2-mm and 1.4-mm clear central zone maintained the best resolution and contrast of the simulated images for all distances.

Conclusions: Our results suggested that a PCL without refractive power might be useful to obtain a full range of vision from far to near if designed optimally.

Keywords

Presbyopia , Pinhole , Contact lenses , Visual simulation

Received : 22 Nov 2014
Accepted : 13 Dec 2014
Published : 16 Dec 2014
Journals
Annals of Otolaryngology and Rhinology
ISSN : 2379-948X
Launched : 2014
JSM Schizophrenia
Launched : 2016
Journal of Nausea
Launched : 2020
JSM Internal Medicine
Launched : 2016
JSM Hepatitis
Launched : 2016
JSM Oro Facial Surgeries
ISSN : 2578-3211
Launched : 2016
Journal of Human Nutrition and Food Science
ISSN : 2333-6706
Launched : 2013
JSM Regenerative Medicine and Bioengineering
ISSN : 2379-0490
Launched : 2013
JSM Spine
ISSN : 2578-3181
Launched : 2016
Archives of Palliative Care
ISSN : 2573-1165
Launched : 2016
JSM Nutritional Disorders
ISSN : 2578-3203
Launched : 2017
Annals of Neurodegenerative Disorders
ISSN : 2476-2032
Launched : 2016
Journal of Fever
ISSN : 2641-7782
Launched : 2017
JSM Bone Marrow Research
ISSN : 2578-3351
Launched : 2016
JSM Mathematics and Statistics
ISSN : 2578-3173
Launched : 2014
Journal of Autoimmunity and Research
ISSN : 2573-1173
Launched : 2014
JSM Arthritis
ISSN : 2475-9155
Launched : 2016
JSM Head and Neck Cancer-Cases and Reviews
ISSN : 2573-1610
Launched : 2016
JSM General Surgery Cases and Images
ISSN : 2573-1564
Launched : 2016
JSM Anatomy and Physiology
ISSN : 2573-1262
Launched : 2016
JSM Dental Surgery
ISSN : 2573-1548
Launched : 2016
Annals of Emergency Surgery
ISSN : 2573-1017
Launched : 2016
Annals of Mens Health and Wellness
ISSN : 2641-7707
Launched : 2017
Journal of Preventive Medicine and Health Care
ISSN : 2576-0084
Launched : 2018
Journal of Chronic Diseases and Management
ISSN : 2573-1300
Launched : 2016
Annals of Vaccines and Immunization
ISSN : 2378-9379
Launched : 2014
JSM Heart Surgery Cases and Images
ISSN : 2578-3157
Launched : 2016
Annals of Reproductive Medicine and Treatment
ISSN : 2573-1092
Launched : 2016
JSM Brain Science
ISSN : 2573-1289
Launched : 2016
JSM Biomarkers
ISSN : 2578-3815
Launched : 2014
JSM Biology
ISSN : 2475-9392
Launched : 2016
Archives of Stem Cell and Research
ISSN : 2578-3580
Launched : 2014
Annals of Clinical and Medical Microbiology
ISSN : 2578-3629
Launched : 2014
JSM Pediatric Surgery
ISSN : 2578-3149
Launched : 2017
Journal of Memory Disorder and Rehabilitation
ISSN : 2578-319X
Launched : 2016
JSM Tropical Medicine and Research
ISSN : 2578-3165
Launched : 2016
JSM Head and Face Medicine
ISSN : 2578-3793
Launched : 2016
JSM Cardiothoracic Surgery
ISSN : 2573-1297
Launched : 2016
JSM Bone and Joint Diseases
ISSN : 2578-3351
Launched : 2017
JSM Bioavailability and Bioequivalence
ISSN : 2641-7812
Launched : 2017
JSM Atherosclerosis
ISSN : 2573-1270
Launched : 2016
Journal of Genitourinary Disorders
ISSN : 2641-7790
Launched : 2017
Journal of Fractures and Sprains
ISSN : 2578-3831
Launched : 2016
Journal of Autism and Epilepsy
ISSN : 2641-7774
Launched : 2016
Annals of Marine Biology and Research
ISSN : 2573-105X
Launched : 2014
JSM Health Education & Primary Health Care
ISSN : 2578-3777
Launched : 2016
JSM Communication Disorders
ISSN : 2578-3807
Launched : 2016
Annals of Musculoskeletal Disorders
ISSN : 2578-3599
Launched : 2016
Annals of Virology and Research
ISSN : 2573-1122
Launched : 2014
JSM Renal Medicine
ISSN : 2573-1637
Launched : 2016
Journal of Muscle Health
ISSN : 2578-3823
Launched : 2016
JSM Genetics and Genomics
ISSN : 2334-1823
Launched : 2013
JSM Anxiety and Depression
ISSN : 2475-9139
Launched : 2016
Clinical Journal of Heart Diseases
ISSN : 2641-7766
Launched : 2016
Annals of Medicinal Chemistry and Research
ISSN : 2378-9336
Launched : 2014
JSM Pain and Management
ISSN : 2578-3378
Launched : 2016
JSM Women's Health
ISSN : 2578-3696
Launched : 2016
Clinical Research in HIV or AIDS
ISSN : 2374-0094
Launched : 2013
Journal of Endocrinology, Diabetes and Obesity
ISSN : 2333-6692
Launched : 2013
Journal of Substance Abuse and Alcoholism
ISSN : 2373-9363
Launched : 2013
JSM Neurosurgery and Spine
ISSN : 2373-9479
Launched : 2013
Journal of Liver and Clinical Research
ISSN : 2379-0830
Launched : 2014
Journal of Drug Design and Research
ISSN : 2379-089X
Launched : 2014
JSM Clinical Oncology and Research
ISSN : 2373-938X
Launched : 2013
JSM Bioinformatics, Genomics and Proteomics
ISSN : 2576-1102
Launched : 2014
JSM Chemistry
ISSN : 2334-1831
Launched : 2013
Journal of Trauma and Care
ISSN : 2573-1246
Launched : 2014
JSM Surgical Oncology and Research
ISSN : 2578-3688
Launched : 2016
Annals of Food Processing and Preservation
ISSN : 2573-1033
Launched : 2016
Journal of Radiology and Radiation Therapy
ISSN : 2333-7095
Launched : 2013
JSM Physical Medicine and Rehabilitation
ISSN : 2578-3572
Launched : 2016
Annals of Clinical Pathology
ISSN : 2373-9282
Launched : 2013
Annals of Cardiovascular Diseases
ISSN : 2641-7731
Launched : 2016
Journal of Behavior
ISSN : 2576-0076
Launched : 2016
Annals of Clinical and Experimental Metabolism
ISSN : 2572-2492
Launched : 2016
Clinical Research in Infectious Diseases
ISSN : 2379-0636
Launched : 2013
JSM Microbiology
ISSN : 2333-6455
Launched : 2013
Journal of Urology and Research
ISSN : 2379-951X
Launched : 2014
Journal of Family Medicine and Community Health
ISSN : 2379-0547
Launched : 2013
Annals of Pregnancy and Care
ISSN : 2578-336X
Launched : 2017
JSM Cell and Developmental Biology
ISSN : 2379-061X
Launched : 2013
Annals of Aquaculture and Research
ISSN : 2379-0881
Launched : 2014
Clinical Research in Pulmonology
ISSN : 2333-6625
Launched : 2013
Journal of Immunology and Clinical Research
ISSN : 2333-6714
Launched : 2013
Annals of Forensic Research and Analysis
ISSN : 2378-9476
Launched : 2014
JSM Biochemistry and Molecular Biology
ISSN : 2333-7109
Launched : 2013
Annals of Breast Cancer Research
ISSN : 2641-7685
Launched : 2016
Annals of Gerontology and Geriatric Research
ISSN : 2378-9409
Launched : 2014
Journal of Sleep Medicine and Disorders
ISSN : 2379-0822
Launched : 2014
JSM Burns and Trauma
ISSN : 2475-9406
Launched : 2016
Chemical Engineering and Process Techniques
ISSN : 2333-6633
Launched : 2013
Annals of Clinical Cytology and Pathology
ISSN : 2475-9430
Launched : 2014
JSM Allergy and Asthma
ISSN : 2573-1254
Launched : 2016
Journal of Neurological Disorders and Stroke
ISSN : 2334-2307
Launched : 2013
Annals of Sports Medicine and Research
ISSN : 2379-0571
Launched : 2014
JSM Sexual Medicine
ISSN : 2578-3718
Launched : 2016
Annals of Vascular Medicine and Research
ISSN : 2378-9344
Launched : 2014
JSM Biotechnology and Biomedical Engineering
ISSN : 2333-7117
Launched : 2013
Journal of Hematology and Transfusion
ISSN : 2333-6684
Launched : 2013
JSM Environmental Science and Ecology
ISSN : 2333-7141
Launched : 2013
Journal of Cardiology and Clinical Research
ISSN : 2333-6676
Launched : 2013
JSM Nanotechnology and Nanomedicine
ISSN : 2334-1815
Launched : 2013
Journal of Ear, Nose and Throat Disorders
ISSN : 2475-9473
Launched : 2016
Journal of Pharmacology and Clinical Toxicology
ISSN : 2333-7079
Launched : 2013
Annals of Psychiatry and Mental Health
ISSN : 2374-0124
Launched : 2013
Medical Journal of Obstetrics and Gynecology
ISSN : 2333-6439
Launched : 2013
Annals of Pediatrics and Child Health
ISSN : 2373-9312
Launched : 2013
JSM Clinical Pharmaceutics
ISSN : 2379-9498
Launched : 2014
JSM Foot and Ankle
ISSN : 2475-9112
Launched : 2016
JSM Alzheimer's Disease and Related Dementia
ISSN : 2378-9565
Launched : 2014
Journal of Addiction Medicine and Therapy
ISSN : 2333-665X
Launched : 2013
Journal of Veterinary Medicine and Research
ISSN : 2378-931X
Launched : 2013
Annals of Public Health and Research
ISSN : 2378-9328
Launched : 2014
Annals of Orthopedics and Rheumatology
ISSN : 2373-9290
Launched : 2013
Journal of Clinical Nephrology and Research
ISSN : 2379-0652
Launched : 2014
Annals of Community Medicine and Practice
ISSN : 2475-9465
Launched : 2014
Annals of Biometrics and Biostatistics
ISSN : 2374-0116
Launched : 2013
JSM Clinical Case Reports
ISSN : 2373-9819
Launched : 2013
Journal of Cancer Biology and Research
ISSN : 2373-9436
Launched : 2013
Journal of Surgery and Transplantation Science
ISSN : 2379-0911
Launched : 2013
Journal of Dermatology and Clinical Research
ISSN : 2373-9371
Launched : 2013
JSM Gastroenterology and Hepatology
ISSN : 2373-9487
Launched : 2013
Annals of Nursing and Practice
ISSN : 2379-9501
Launched : 2014
JSM Dentistry
ISSN : 2333-7133
Launched : 2013
Author Information X